Literature DB >> 30854571

The effects of gel-based manicure on pulse oximetry.

Jia Lin Jacklyn Yek1, Hairil Rizal Abdullah1, June Pheck Suan Goh1, Yew Weng Chan1.   

Abstract

INTRODUCTION: Pulse oximetry is the standard monitoring technique of functional oxygen saturation (SpO2). As the use of fingernail polish has been described to alter SpO2 readings, its removal is commonly recommended prior to measurement. Gel-based manicures have gained popularity in recent years due to their attractiveness and longevity. However, the removal of gel nail polish requires a specialised procedure. Valuable time and resources can be saved if removal can be avoided. To our knowledge, there are no available studies on the effect of gel-based manicures on pulse oximetry readings. Hence, we evaluated the effect with two oximeters, using different technology and wavelength combinations.
METHODS: 17 healthy female adult volunteers were recruited for this single-blind randomised controlled trial. Subjects with hypothermia, hypotension, poor plethysmographic waveform and nail pathology were excluded. Colours tested were: black, purple, navy blue, green, light blue, white, yellow, orange, pink and red. Pulse oximetry was measured at 15- and 30-second intervals using two different pulse oximeters, the Philips M1191BL and Masimo SET®. Means were compared using paired t-tests.
RESULTS: Using the Masimo oximeter, light blue (ΔM = 0.97% ± 0.96%; p = 0.001) and orange (ΔM = 0.76 ± 1.17%; p = 0.016) gel nail polish resulted in a statistically significant increase from baseline SpO2 readings. With the Philips oximeter, the limits of agreement ranged from 2% for pink to 17% for black, indicating imprecision.
CONCLUSION: Gel-based manicures can result in overestimations of actual readings, delaying detection of hypoxaemia. Gel nail polish should be routinely removed or an alternative monitoring technique sought. Copyright: © Singapore Medical Association.

Entities:  

Keywords:  SpO2; gel manicure; nail polish; oxygen saturation; pulse oximetry

Mesh:

Substances:

Year:  2019        PMID: 30854571      PMCID: PMC6717771          DOI: 10.11622/smedj.2019031

Source DB:  PubMed          Journal:  Singapore Med J        ISSN: 0037-5675            Impact factor:   1.858


  12 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

2.  CONSORT 2010 statement: updated guidelines for reporting parallel group randomised trials.

Authors:  Kenneth F Schulz; Douglas G Altman; David Moher
Journal:  Int J Surg       Date:  2011-10-13       Impact factor: 6.071

Review 3.  Potential errors in pulse oximetry. III: Effects of interferences, dyes, dyshaemoglobins and other pigments.

Authors:  A C Ralston; R K Webb; W B Runciman
Journal:  Anaesthesia       Date:  1991-04       Impact factor: 6.955

4.  Missing data imputation: focusing on single imputation.

Authors:  Zhongheng Zhang
Journal:  Ann Transl Med       Date:  2016-01

5.  Does fingernail polish affect pulse oximeter readings?

Authors:  Ann M Rodden; Leigh Spicer; Vanessa A Diaz; Terrence E Steyer
Journal:  Intensive Crit Care Nurs       Date:  2006-10-24       Impact factor: 3.072

6.  Nail polish does not significantly affect pulse oximetry measurements in mildly hypoxic subjects.

Authors:  Loren G Yamamoto; Julienne A Yamamoto; Joelle B Yamamoto; Brennan E Yamamoto; Patricia P Yamamoto
Journal:  Respir Care       Date:  2008-11       Impact factor: 2.258

Review 7.  Cosmetically Induced Disorders of the Nail with Update on Contemporary Nail Manicures.

Authors:  Evan A Rieder; Antonella Tosti
Journal:  J Clin Aesthet Dermatol       Date:  2016-04-01

8.  Pulse oximeter failure thresholds in hypotension and vasoconstriction.

Authors:  J W Severinghaus; M J Spellman
Journal:  Anesthesiology       Date:  1990-09       Impact factor: 7.892

9.  Artificial acrylic finger nails may alter pulse oximetry measurement.

Authors:  Jochen Hinkelbein; Heike Koehler; Harald V Genzwuerker; Fritz Fiedler
Journal:  Resuscitation       Date:  2007-03-13       Impact factor: 5.262

10.  Effect of nail polish on oxygen saturation determined by pulse oximetry in critically ill patients.

Authors:  Jochen Hinkelbein; Harald V Genzwuerker; Reiner Sogl; Fritz Fiedler
Journal:  Resuscitation       Date:  2006-11-13       Impact factor: 5.262

View more