Artificial acrylic finger nails may alter pulse oximetry measurement.

Jochen Hinkelbein1, Heike Koehler, Harald V Genzwuerker, Fritz Fiedler.   

Abstract

INTRODUCTION: Pulse oximetry is the most common technique to monitor oxygen saturation (SpO(2)) during intensive care therapy. However, intermittent co-oximetry is still the "gold standard" (SaO(2)). Besides acrylic nails, numerous other factors have been reported to interfere with pulse oximetry. Data of measurements with artificial finger nails are not sufficiently published.
MATERIALS AND METHODS: A prospective clinical-experimental trial in mechanically ventilated and critically ill patients of an ICU was performed. Patients were randomly assigned to either group S (S: Siemens pulse oximeter) or group P (P: Philips pulse oximeter) prior to the measurements. SpO(2) was determined in each patient three times alternately in standard ((N)SpO(2)) and sideways position at the natural nail ((N90)SpO(2)). For the reference measurements oxygen saturation was measured by means of a haemoximeter (co-oximetry). Thereafter, SpO(2) was obtained at the acrylic finger nail in the same way ((A)SpO(2) and (A90)SpO(2)). Bias was calculated as DeltaS=(N)SpO(2)-SaO(2) and DeltaS=(A)SpO(2)-SaO(2). Accuracy (mean difference) and precision (standard deviation) were used to determine the measurement discrepancy. P<0.05 was considered significant.
RESULTS: Accuracy and precision without acrylic nails applied were comparable to SaO(2) in both groups (n.s.). With acrylic nails applied a bias of DeltaS=-1.1+/-3.14% for group S (P=0.00522) and a bias of DeltaS=+0.8+/-3.04% for group P was calculated (n.s.).
CONCLUSION: Acrylic finger nails may impair the measurement of oxygen saturation depending on the pulse oximeter used and may cause significant inaccuracy. Hence, removal of artificial acrylic finger nails may be helpful to assure an accurate and precise measurement with pulse oximetry.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17353079     DOI: 10.1016/j.resuscitation.2006.11.018

Source DB:  PubMed          Journal:  Resuscitation        ISSN: 0300-9572            Impact factor:   5.262


  7 in total

1.  The effects of gel-based manicure on pulse oximetry.

Authors:  Jia Lin Jacklyn Yek; Hairil Rizal Abdullah; June Pheck Suan Goh; Yew Weng Chan
Journal:  Singapore Med J       Date:  2019-03-11       Impact factor: 1.858

2.  Pulse oximetry error in a patient with a Santa Ana haemoglobinopathy.

Authors:  Alistair Robertson; Amin Rahemtulla
Journal:  BMJ Case Rep       Date:  2016-09-06

3.  Impact of ethnicity on the accuracy of measurements of oxygen saturations: A retrospective observational cohort study.

Authors:  Mansoor N Bangash; James Hodson; Felicity Evison; Jaimin M Patel; Andrew McD Johnston; Suzy Gallier; Elizabeth Sapey; Dhruv Parekh
Journal:  EClinicalMedicine       Date:  2022-05-06

4.  Screening for obstructive sleep apnea among hospital outpatients.

Authors:  Michel Hug; Katrin Uehli; Stig Solbach; Stefanie Brighenti-Zogg; Selina Dürr; Sabrina Maier; Jörg Daniel Leuppi; David Miedinger
Journal:  PLoS One       Date:  2018-05-30       Impact factor: 3.240

5.  A multicentre prospective observational study comparing arterial blood gas values to those obtained by pulse oximeters used in adult patients attending Australian and New Zealand hospitals.

Authors:  Janine Pilcher; Laura Ploen; Steve McKinstry; George Bardsley; Jimmy Chien; Lesley Howard; Sharon Lee; Lutz Beckert; Maureen Swanney; Mark Weatherall; Richard Beasley
Journal:  BMC Pulm Med       Date:  2020-01-09       Impact factor: 3.317

6.  The feasibility, safety, and efficacy of upper limb garment-integrated blood flow restriction training in healthy adults.

Authors:  Bhavit Dhokia; Elspeth Olivia Mabin; Warren Jeremy Bradley; Bradley Stephen Neal
Journal:  Pilot Feasibility Stud       Date:  2022-02-08

7.  The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: a systematic review and meta-analysis.

Authors:  Chunhu Shi; Mark Goodall; Jo Dumville; James Hill; Gill Norman; Oliver Hamer; Andrew Clegg; Caroline Leigh Watkins; George Georgiou; Alexander Hodkinson; Catherine Elizabeth Lightbody; Paul Dark; Nicky Cullum
Journal:  BMC Med       Date:  2022-08-16       Impact factor: 11.150

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.