| Literature DB >> 30846895 |
Toshiro Suzuki1, Ryouhei Hattori2, Tomonori Minagawa1, Takeshi Uehara3, Teruyuki Ogawa1, Osamu Ishizuka1.
Abstract
BACKGROUND: There have been no investigations of intestinal injury induced by surgical sealing devices, especially focusing heat conduction from the back of active blades during laparoscopic surgery.Entities:
Keywords: Animal model; Bowel injury; Intra-operative heat injury; Laparoscopic surgery; Sealing device
Mesh:
Year: 2019 PMID: 30846895 PMCID: PMC6400247 DOI: 10.4293/JSLS.2018.00100
Source DB: PubMed Journal: JSLS ISSN: 1086-8089 Impact factor: 2.172
Histological Changes in the Pig Intestine Touched to the Back of the AB During Activation Were Compared for Each Device
| Type of sealing devices | Depth of Heat Injury (+: a Sample with Thermal Denaturation) | ||||||
|---|---|---|---|---|---|---|---|
| Serous Membrane | Longitudinal Muscle | Circular Muscle | Deep Submucosal Layer | Shallow Submucosal Layer | Muscularis Mucosae | Mucosal Layer | |
| ENSEAL TRIO® | ++++ | ++++ | ++++ | ++++ | +++ | +++ | +++ |
| ENSEAL Round® | +++ | +++ | +++ | +++ | +++ | +++ | +++ |
| ENSEAL G2 Articulating® | +++++ | ++++ | − | ++++ | + | + | + |
| LigaSure V® | +++++ | ++ | − | ++ | − | − | − |
| LigaSure Blunt Tip® | ++ | ++ | − | ++ | ++ | ++ | − |
| LigaSure Maryland Jaw® | +++++ | ++++ | + | ++++ | ++ | − | + |
| BiClamp® | ++++ | ++++ | − | ++++ | ++++ | ++++ | − |
| HARMONIC ACE®+ | +++++ | +++++ | ++ | ++++ | + | + | + |
| THUNDERBEAT® | +++++ | +++++ | + | +++ | ++ | ++ | ++ |
Devices Used in this Study and their Features
| Product Name | Manufacturer | Energy System | Function | Release Year |
|---|---|---|---|---|
| ENSEAL TRIO® | ETHICON (Blue Ash, Ohio, USA) | Bipolar | Sealing and cut coagulation | 2010 |
| ENSEAL Round® | ETHICON | Bipolar | Sealing and cut coagulation | 2010 |
| ENSEAL G2 Articulating® | ETHICON | Bipolar | Sealing and cut coagulation | 2015 |
| LigaSure V® | Medtronic, PLC (Dublin, Ireland) | Bipolar | Sealing and cut coagulation | 2004 |
| LigaSure Blunt® | Medtronic plc | Bipolar | Sealing and cut coagulation | 2014 |
| LigaSure Maryland® | Medtronic, PLC | Bipolar | Sealing and cut coagulation | 2015 |
| BiClamp® | ERBE (Tübingen, Germany) | Bipolar | Coagulation | 2009 |
| HARMONIC ACE®+ | ETHICON | Ultrasonic | Sealing and cut | 2015 |
| THUNDERBEAT® | OLYMPUS (Tokyo, Japan) | Ultrasonic Bipolar | Sealing and cut coagulation | 2013 |