Literature DB >> 30756324

A Miniaturized Fourier Transform Electrostatic Linear Ion Trap Mass Spectrometer: Mass Range and Resolution.

Joshua T Johnson1, Kenneth W Lee1, Jay S Bhanot1, Scott A McLuckey2.   

Abstract

Mass resolution (M/ΔMFWHM) was increased by reducing the axial length of a Fourier transform electrostatic linear ion trap (FT-ELIT) mass spectrometer. The increase in mass resolution corresponds directly to increased axial ion frequencies in the FT-ELIT. Increased mass resolution was demonstrated for equivalent transient lengths in a 5.25″ versus 2.625″ ELIT using the isotopes of [bradykinin+2H]2+ and [insulin+5H]5+ as test ions. Both bradykinin and insulin show mass resolution increases of ~ 90% allowing baseline resolution of the [insulin+5H]5+ isotopes after only 300 ms of data acquisition. Relative changes in mass/charge range were explored using mirror switching to trap ions injected axially into the ELIT. When trapping ions using mirror switching, the mass/charge range in a FT-ELIT mass spectrometer for a given switch time is determined by the time required for fast ions to enter and exit the trap after one reflection versus the time it takes for slow ions to enter the trap. By reducing the length of the FT-ELIT mass spectrometer while maintaining a constant distance from the point from which ions are initially accelerated to the entrance mirror, only the low m/z limit is affected for a given mirror switching time. For the two ELIT lengths examined here, the effective mass/charge range at any given switch time is reduced from m/zlow-8.9*m/zlow for the 5.25″ ELIT to m/zlow-5.2*m/zlow for the 2.625″ ELIT. Graphical Abstract.

Entities:  

Keywords:  Electrostatic linear ion trap; Fourier transform mass spectrometry; Miniaturization

Year:  2019        PMID: 30756324     DOI: 10.1007/s13361-018-02126-x

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  15 in total

1.  Fourier transform time-of-flight mass spectrometry in an electrostatic ion beam trap

Authors: 
Journal:  Anal Chem       Date:  2000-09-01       Impact factor: 6.986

2.  Fourier transform ion cyclotron resonance cell with dynamic harmonization of the electric field in the whole volume by shaping of the excitation and detection electrode assembly.

Authors:  Ivan A Boldin; Eugene N Nikolaev
Journal:  Rapid Commun Mass Spectrom       Date:  2011-01-15       Impact factor: 2.419

3.  Induction charge detector with multiple sensing stages.

Authors:  Manuel Gamero-Castaño
Journal:  Rev Sci Instrum       Date:  2007-04       Impact factor: 1.523

4.  Artifacts in Fourier transform mass spectrometry.

Authors:  Raman Mathur; Peter B O'Connor
Journal:  Rapid Commun Mass Spectrom       Date:  2009-02       Impact factor: 2.419

5.  Absorption mode Fourier transform electrostatic linear ion trap mass spectrometry.

Authors:  Ryan T Hilger; Phillip J Wyss; Robert E Santini; Scott A McLuckey
Journal:  Anal Chem       Date:  2013-08-15       Impact factor: 6.986

6.  A multi-stage image charge detector made from printed circuit boards.

Authors:  Brandon L Barney; R Terik Daly; Daniel E Austin
Journal:  Rev Sci Instrum       Date:  2013-11       Impact factor: 1.523

Review 7.  Fourier transform ion cyclotron resonance mass spectrometry: a primer.

Authors:  A G Marshall; C L Hendrickson; G S Jackson
Journal:  Mass Spectrom Rev       Date:  1998 Jan-Feb       Impact factor: 10.946

8.  Utility of Higher Harmonics in Electrospray Ionization Fourier Transform Electrostatic Linear Ion Trap Mass Spectrometry.

Authors:  Eric T Dziekonski; Joshua T Johnson; Scott A McLuckey
Journal:  Anal Chem       Date:  2017-03-30       Impact factor: 6.986

9.  Fourier-Transform MS and Closed-Path Multireflection Time-of-Flight MS Using an Electrostatic Linear Ion Trap.

Authors:  Eric T Dziekonski; Joshua T Johnson; Kenneth W Lee; Scott A McLuckey
Journal:  Anal Chem       Date:  2017-10-03       Impact factor: 6.986

10.  21 Tesla FT-ICR Mass Spectrometer for Ultrahigh-Resolution Analysis of Complex Organic Mixtures.

Authors:  Donald F Smith; David C Podgorski; Ryan P Rodgers; Greg T Blakney; Christopher L Hendrickson
Journal:  Anal Chem       Date:  2018-01-05       Impact factor: 6.986

View more