| Literature DB >> 30675577 |
Kevyn Mejia-Hernandez1,2, Angela Chang1,2, Nathan Eardley-Harris2, Ruurd Jaarsma2, Tiffany K Gill1, James M McLean1.
Abstract
HYPOTHESIS ANDEntities:
Keywords: Constant-Murley score; Goniometer; Inclinometer; Shoulder range of motion; Smartphone application; UCLA score
Year: 2018 PMID: 30675577 PMCID: PMC6334873 DOI: 10.1016/j.jses.2017.10.001
Source DB: PubMed Journal: JSES Open Access ISSN: 2468-6026
Figure 1The iPhone applications. (A) GetMyROM. (B) DrGoniometer.
Figure 2Measurements of (A) forward flexion, (B) glenohumeral abduction, (C) external rotation, and (D) internal rotation.
Diagnoses and operations of participants (N = 75)
| Diagnosis | Operation |
|---|---|
| Inflammation (n = 9) | No operation (n = 58) |
SA, subacromial; ACJ, acromioclavicular joint; SCJ, sternoclavicular joint; GHJ, glenohumeral joint; SLAP, superior labral tear from anterior to posterior.
Interobserver reliability for each range of motion (ROM) measurement method
| ROM test | Method | ICC | Lower 95% limit | Upper 95% limit |
|---|---|---|---|---|
| Forward flexion | Clinician | 1.00 | 0.99 | 1.00 |
| Goniometer | 0.99 | 0.99 | 1.00 | |
| GetMyROM | 0.99 | 0.98 | 0.99 | |
| DrGoniometer | 1.00 | 0.99 | 1.00 | |
| Total active abduction | Clinician | 1.00 | 0.99 | 1.00 |
| Goniometer | 0.99 | 0.99 | 1.00 | |
| GetMyROM | 0.99 | 0.99 | 1.00 | |
| DrGoniometer | 0.99 | 0.99 | 1.00 | |
| Active glenohumeral abduction | Clinician | 0.99 | 0.99 | 1.00 |
| Goniometer | 0.98 | 0.97 | 0.99 | |
| GetMyROM | 0.98 | 0.97 | 0.99 | |
| DrGoniometer | 0.98 | 0.97 | 0.99 | |
| Passive glenohumeral abduction | Clinician | 0.99 | 0.99 | 1.00 |
| Goniometer | 0.98 | 0.99 | 0.99 | |
| GetMyROM | 0.97 | 0.95 | 0.99 | |
| DrGoniometer | 0.98 | 0.96 | 0.99 | |
| Active internal rotation | Clinician | 0.99 | 0.99 | 0.99 |
| Goniometer | 0.99 | 0.99 | 0.99 | |
| GetMyROM | 0.98 | 0.97 | 0.99 | |
| DrGoniometer | 0.99 | 0.99 | 1.00 | |
| Passive internal rotation | Clinician | 0.99 | 0.98 | 0.99 |
| Goniometer | 0.99 | 0.99 | 0.99 | |
| GetMyROM | 0.98 | 0.96 | 0.98 | |
| DrGoniometer | 1.00 | 0.99 | 1.00 | |
| Active external rotation | Clinician | 0.99 | 0.98 | 0.99 |
| Goniometer | 0.99 | 0.98 | 0.99 | |
| GetMyROM | 0.99 | 0.98 | 0.99 | |
| DrGoniometer | 1.00 | 0.99 | 1.00 | |
| Passive external rotation | Clinician | 0.99 | 0.99 | 1.00 |
| Goniometer | 1.00 | 0.99 | 1.00 | |
| GetMyROM | 0.99 | 0.98 | 0.99 | |
| DrGoniometer | 0.99 | 0.99 | 1.00 |
ICC, intraclass correlation coefficient.
Bland-Altman plots comparing measurement modalities to the “gold standard” goniometer
| Comparison measurement mean to universal goniometer | Comparative measurement technique | Mean difference | Lower 95% limit | Upper 95% limit |
|---|---|---|---|---|
| Forward flexion | Clinician | 1.69 | −5.17 | 8.55 |
| GetMyROM | −0.76 | −9.64 | 8.11 | |
| DrGoniometer | −0.56 | −9.63 | 8.52 | |
| Total active abduction | Clinician | 0.37 | −6.46 | 7.20 |
| GetMyROM | 0.47 | −7.87 | 8.81 | |
| DrGoniometer | 0.81 | −7.73 | 9.35 | |
| Active glenohumeral abduction | Clinician | −0.18 | −2.90 | 2.54 |
| GetMyROM | −0.19 | −4.71 | 4.32 | |
| DrGoniometer | −0.41 | −6.93 | 6.12 | |
| Passive glenohumeral abduction | Clinician | −0.07 | −2.04 | 1.90 |
| GetMyROM | −0.38 | −4.02 | 3.25 | |
| DrGoniometer | −0.01 | −2.32 | 2.30 | |
| Active internal rotation | Clinician | −0.29 | −5.51 | 4.93 |
| GetMyROM | 0.51 | −7.11 | 8.14 | |
| DrGoniometer | −1.29 | −10.00 | 7.43 | |
| Passive internal rotation | Clinician | 0.00 | −4.66 | 4.66 |
| GetMyROM | 0.55 | −5.04 | 6.13 | |
| DrGoniometer | −1.41 | −9.83 | 7.01 | |
| Active external rotation | Clinician | 0.01 | −5.02 | 5.05 |
| GetMyROM | −0.08 | −8.32 | 8.17 | |
| DrGoniometer | 0.20 | −8.23 | 8.63 | |
| Passive external rotation | Clinician | 0.42 | −5.18 | 6.02 |
| GetMyROM | 0.40 | −7.58 | 8.37 | |
| DrGoniometer | 0.13 | −7.65 | 7.91 |