Validation of a wrist-type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-9600T.

Mitsuo Kuwabara1,2, Kanako Harada2, Yukiko Hishiki2, Kazuomi Kario1.   

Abstract

The purpose of the present study was to evaluate the performance of the Omron HEM-9600T, an automatic wrist-type device for self BP measurement, in the sitting position with the wrist at heart level and supine position according to the ANSI/AAMI/ISO81060-2:2013 guidelines. In the supine position, we evaluated the device under 3 different conditions: using the supine with sideways palm position, the supine with upwards palm position, and the supine with downwards palm position. After 106 subjects were screened and 21 subjects were excluded, the same 85 subjects (38 men [44.7%] and 47 women [55.3%]) were included in the analyses for each position. The average age of the subjects was 54.5 ± 12.2 years (mean ± SD). The mean wrist circumference was 17.0 ± 2.4 cm. The wrist size distribution fulfilled the requirements of the guidelines. The mean differences between reference BPs and HEM-9600T readings were 1.0 ± 6.7/1.4 ± 5.7 mm Hg, 6.6 ± 7.2/5.5 ± 6.0 mm Hg, 4.8 ± 7.2/4.9 ± 5.8 mm Hg, and 2.1 ± 7.2/2.8 ± 6.8 mm Hg for SBP/DBP in the sitting position, supine with sideways palm position, supine with upwards palm position, and supine with downwards palm position, respectively. In conclusion, the Omron HEM-9600T in the sitting position fulfilled the validation criteria of the ANSI/AAMI/ISO81060-2:2013 guidelines. On the other hand, the accuracies of HEM-9600T in the supine position differed depending on the positioning of the palm, with only the downwards palm-position measurement fulfilling both validation criteria of the ANSI/AAMI/ISO81060-2:2013 guidelines. ©2019 Wiley Periodicals, Inc.

Entities:  

Keywords:  ANSI/AAMI/ISO81060-2:2013; blood pressure; home nocturnal blood pressure; self-measurement; validation; wrist-type blood pressure monitor

Mesh:

Year:  2019        PMID: 30609129      PMCID: PMC8030361          DOI: 10.1111/jch.13464

Source DB:  PubMed          Journal:  J Clin Hypertens (Greenwich)        ISSN: 1524-6175            Impact factor:   3.738


  20 in total

1.  Comparison of the Omron HEM-637 wrist monitor to the auscultation method with the wrist position sensor on or disabled.

Authors:  Steven A Yarows
Journal:  Am J Hypertens       Date:  2004-01       Impact factor: 2.689

2.  Prognostic value of ambulatory and home blood pressures compared with office blood pressure in the general population: follow-up results from the Pressioni Arteriose Monitorate e Loro Associazioni (PAMELA) study.

Authors:  Roberto Sega; Rita Facchetti; Michele Bombelli; Giancarlo Cesana; Giovanni Corrao; Guido Grassi; Giuseppe Mancia
Journal:  Circulation       Date:  2005-04-04       Impact factor: 29.690

3.  Nighttime home blood pressure and the risk of hypertensive target organ damage.

Authors:  Joji Ishikawa; Satoshi Hoshide; Kazuo Eguchi; Shizukiyo Ishikawa; Kazuyuki Shimada; Kazuomi Kario
Journal:  Hypertension       Date:  2012-08-14       Impact factor: 10.190

4.  Home blood pressure measurement has a stronger predictive power for mortality than does screening blood pressure measurement: a population-based observation in Ohasama, Japan.

Authors:  T Ohkubo; Y Imai; I Tsuji; K Nagai; J Kato; N Kikuchi; A Nishiyama; A Aihara; M Sekino; M Kikuya; S Ito; H Satoh; S Hisamichi
Journal:  J Hypertens       Date:  1998-07       Impact factor: 4.844

5.  The association between home vs. ambulatory night-time blood pressure and end-organ damage in the general population.

Authors:  Annika S Lindroos; Jouni K Johansson; Pauli J Puukka; Ilkka Kantola; Veikko Salomaa; Eeva P Juhanoja; Sam S E Sivén; Pekka Jousilahti; Antti M Jula; Teemu J Niiranen
Journal:  J Hypertens       Date:  2016-09       Impact factor: 4.844

6.  Comparative Effects of an Angiotensin II Receptor Blocker (ARB)/Diuretic vs. ARB/Calcium-Channel Blocker Combination on Uncontrolled Nocturnal Hypertension Evaluated by Information and Communication Technology-Based Nocturnal Home Blood Pressure Monitoring - The NOCTURNE Study.

Authors:  Kazuomi Kario; Naoko Tomitani; Hiroshi Kanegae; Hajime Ishii; Kazuaki Uchiyama; Kayo Yamagiwa; Toshihiko Shiraiwa; Tomohiro Katsuya; Tetsuro Yoshida; Kiyomi Kanda; Shinji Hasegawa; Satoshi Hoshide
Journal:  Circ J       Date:  2017-03-17       Impact factor: 2.993

7.  Effect of dosing time of angiotensin II receptor blockade titrated by self-measured blood pressure recordings on cardiorenal protection in hypertensives: the Japan Morning Surge-Target Organ Protection (J-TOP) study.

Authors:  Kazuomi Kario; Satoshi Hoshide; Motohiro Shimizu; Yuichiro Yano; Kazuo Eguchi; Joji Ishikawa; Shizukiyo Ishikawa; Kazuyuki Shimada
Journal:  J Hypertens       Date:  2010-07       Impact factor: 4.844

8.  Prognostic accuracy of day versus night ambulatory blood pressure: a cohort study.

Authors:  José Boggia; Yan Li; Lutgarde Thijs; Tine W Hansen; Masahiro Kikuya; Kristina Björklund-Bodegård; Tom Richart; Takayoshi Ohkubo; Tatiana Kuznetsova; Christian Torp-Pedersen; Lars Lind; Hans Ibsen; Yutaka Imai; Jiguang Wang; Edgardo Sandoya; Eoin O'Brien; Jan A Staessen
Journal:  Lancet       Date:  2007-10-06       Impact factor: 79.321

9.  Validation of a wrist-type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-9600T.

Authors:  Mitsuo Kuwabara; Kanako Harada; Yukiko Hishiki; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-01-04       Impact factor: 3.738

10.  Sleep Blood Pressure Self-Measured at Home as a Novel Determinant of Organ Damage: Japan Morning Surge Home Blood Pressure (J-HOP) Study.

Authors:  Kazuomi Kario; Satoshi Hoshide; Hajime Haimoto; Kayo Yamagiwa; Kiyoshi Uchiba; Shoichiro Nagasaka; Yuichiro Yano; Kazuo Eguchi; Yoshio Matsui; Motohiro Shimizu; Joji Ishikawa; Shizukiyo Ishikawa
Journal:  J Clin Hypertens (Greenwich)       Date:  2015-02-16       Impact factor: 3.738

View more
  12 in total

Review 1.  Diversity of and initiatives for hypertension management in Asia-Why we need the HOPE Asia Network.

Authors:  Kazuomi Kario; Yook-Chin Chia; Apichard Sukonthasarn; Yuda Turana; Jinho Shin; Chen-Huan Chen; Peera Buranakitjaroen; Jennifer Nailes; Satoshi Hoshide; Saulat Siddique; Jorge Sison; Arieska Ann Soenarta; Guru Prasad Sogunuru; Jam Chin Tay; Boon Wee Teo; Yu-Qing Zhang; Sungha Park; Huynh Van Minh; Naoko Tomitani; Tomoyuki Kabutoya; Narsingh Verma; Tzung-Dau Wang; Ji-Guang Wang
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-11-26       Impact factor: 3.738

Review 2.  COVID-19 and hypertension-evidence and practical management: Guidance from the HOPE Asia Network.

Authors:  Kazuomi Kario; Yuji Morisawa; Apichard Sukonthasarn; Yuda Turana; Yook-Chin Chia; Sungha Park; Tzung-Dau Wang; Chen-Huan Chen; Jam Chin Tay; Yan Li; Ji-Guang Wang
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-07-09       Impact factor: 3.738

3.  Validation of a wrist-type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-9601T.

Authors:  Mitsuo Kuwabara; Kanako Harada; Yukiko Hishiki; Takayoshi Ohkubo; Kazuomi Kario; Yutaka Imai
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-05-24       Impact factor: 3.738

4.  Nighttime Blood Pressure Measured by Home Blood Pressure Monitoring as an Independent Predictor of Cardiovascular Events in General Practice.

Authors:  Kazuomi Kario; Hiroshi Kanegae; Naoko Tomitani; Yukie Okawara; Takeshi Fujiwara; Yuichiro Yano; Satoshi Hoshide
Journal:  Hypertension       Date:  2019-06       Impact factor: 10.190

Review 5.  Key Points of the 2019 Japanese Society of Hypertension Guidelines for the Management of Hypertension.

Authors:  Kazuomi Kario
Journal:  Korean Circ J       Date:  2019-12       Impact factor: 3.243

Review 6.  Clinical significance of nocturnal home blood pressure monitoring and nocturnal hypertension in Asia.

Authors:  Takeshi Fujiwara; Satoshi Hoshide; Naoko Tomitani; Hao-Min Cheng; Arieska Ann Soenarta; Yuda Turana; Chen-Huan Chen; Huynh Van Minh; Guru Prasad Sogunuru; Jam Chin Tay; Tzung-Dau Wang; Yook-Chin Chia; Narsingh Verma; Yan Li; Ji-Guang Wang; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2021-02-16       Impact factor: 3.738

7.  Validation of two watch-type wearable blood pressure monitors according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-6410T-ZM and HEM-6410T-ZL.

Authors:  Mitsuo Kuwabara; Kanako Harada; Yukiko Hishiki; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-02-25       Impact factor: 3.738

Review 8.  Asian management of hypertension: Current status, home blood pressure, and specific concerns in Japan.

Authors:  Tomoyuki Kabutoya; Satoshi Hoshide; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-10-17       Impact factor: 3.738

9.  Validation of a wrist-type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-9600T.

Authors:  Mitsuo Kuwabara; Kanako Harada; Yukiko Hishiki; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-01-04       Impact factor: 3.738

10.  Management of Hypertension in the Digital Era: Small Wearable Monitoring Devices for Remote Blood Pressure Monitoring.

Authors:  Kazuomi Kario
Journal:  Hypertension       Date:  2020-08-03       Impact factor: 10.190

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.