Comparison of the Omron HEM-637 wrist monitor to the auscultation method with the wrist position sensor on or disabled.

Steven A Yarows1.   

Abstract

BACKGROUND: To determine whether the Omron HEM637 wrist model with the wrist positioning sensor turned on (Son) is more accurate relative to upper arm auscultation by trained professionals than when the sensor was manually turned off (Soff).
METHODS: Forty-four subjects, at least 30 years old, had repeated, sequential dual-observer upper arm auscultatory measurements (5 to 6 each) and oscillometric Omron HEM637 wrist measurements (4 each). Nineteen subjects were assigned to the wrist sensor On group and 25 were assigned to the wrist sensor Off group. A total of 425 auscultatory and 164 wrist measurements were analyzed.
RESULTS: The Omron HEM-637 measured the blood pressure (BP) with equal accuracy to the observers using the auscultatory technique (difference -1.37 +/- 8.51/3.47 +/- 8.07 mm Hg, P =.71/.14). The wrist sensor did improve the accuracy of the measurements compared to the subjects that had the sensor deactivated. The sensor On group (Son) measured the systolic BP (0.82 +/- 9.83 mm Hg) and diastolic BP (-0.72 +/- 9.07 mm Hg) statistically the same as by auscultation (P =.86 for systolic BP and P =.83 for diastolic BP). The sensor Off (Soff) group measured the SBP (-3.03 +/- 7.12 mm Hg) and diastolic BP (-5.56 +/- 6.68 mm Hg) statistically different than auscultation (P =.46 for systolic BP and P =.02 for diastolic BP). The higher (negative) measurement for both the systolic BP and diastolic BP suggests that the average position of the wrist was 1.75 inches (4.4 cm) below the heart level in this group with the sensor off.
CONCLUSIONS: This study demonstrated that the Omron HEM-637 monitor with a wrist sensor more accurately measured BP compared to the same model with the sensor turned off.

Entities:  

Mesh:

Year:  2004        PMID: 14700513     DOI: 10.1016/j.amjhyper.2003.09.003

Source DB:  PubMed          Journal:  Am J Hypertens        ISSN: 0895-7061            Impact factor:   2.689


  6 in total

1.  Validation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010.

Authors:  Hakuo Takahashi; Masamichi Yoshika; Toyohiko Yokoi
Journal:  Vasc Health Risk Manag       Date:  2013-05-28

2.  Validation of a wrist-type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-9601T.

Authors:  Mitsuo Kuwabara; Kanako Harada; Yukiko Hishiki; Takayoshi Ohkubo; Kazuomi Kario; Yutaka Imai
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-05-24       Impact factor: 3.738

3.  Validation of two automatic devices, Omron HEM-6232T and HEM-6181, for self-measurement of blood pressure at the wrist according to the ANSI/AAMI/ISO 81060-2:2013 protocol and the European Society of Hypertension International Protocol revision 2010.

Authors:  Kanako Saito; Yukiko Hishiki; Hakuo Takahashi
Journal:  Vasc Health Risk Manag       Date:  2019-03-06

4.  Reproducibility of wrist home blood pressure measurement with position sensor and automatic data storage.

Authors:  Sakir Uen; Rolf Fimmers; Miriam Brieger; Georg Nickenig; Thomas Mengden
Journal:  BMC Cardiovasc Disord       Date:  2009-05-27       Impact factor: 2.298

5.  Validation of two watch-type wearable blood pressure monitors according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-6410T-ZM and HEM-6410T-ZL.

Authors:  Mitsuo Kuwabara; Kanako Harada; Yukiko Hishiki; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-02-25       Impact factor: 3.738

6.  Validation of a wrist-type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-9600T.

Authors:  Mitsuo Kuwabara; Kanako Harada; Yukiko Hishiki; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-01-04       Impact factor: 3.738

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.