| Literature DB >> 30581857 |
Nilofar Massoudi1, Mohammad Fathi1, Navid Nooraei2, Alireza Salehi1.
Abstract
OBJECTIVES: The aim of the present study was to compare two supraglottic airway (SGA) devices (i.e., the i-gel® © Intersurgical Ltd and air-Q® (Reusable) Cookgas company) in terms of the insertion time, amount of leak during ventilation with maximum positive pressure, and postoperative complications in patients referring to Modarres Hospital in Tehran.Entities:
Mesh:
Year: 2018 PMID: 30581857 PMCID: PMC6276498 DOI: 10.1155/2018/5202957
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Characteristics of the i-gel and air-Q groups.
|
|
|
| |
|---|---|---|---|
|
| 37.30 ±15.8 | 33.73 ± 11.4 | 0.549 |
|
| 14 (46.7%) | 13 (43.3%) | 0.795b |
|
| 21.40 ± 1.6 | 22.23 ± 1.9 | 0.069 |
|
| 2.21 ± 0.6 | 2.21 ± 0.6 | 1.00 |
|
| 1.63 ± 0.5 | 1.63 ± 0.5 | 1.00 |
aBy the Mann–Whitney test; bby Pearson's chi-square.
Figure 1Mean insertion time (in seconds) and mean number of attempts for insertion in the i-gel and air-Q groups.
Figure 2Mean oropharyngeal leak pressure (cmH2O) in the i-gel and air-Q groups.
Comparison of hemodynamic characteristics between the i-gel and air-Q groups.
|
|
|
|
| |
|---|---|---|---|---|
|
| First | 34.50 ±3.3 | 37.63 ± 4.8 | 0.009 |
| Second | 30.57 ± 2.6 | 35.43 ± 3.5 | < 0.001 | |
| Third | 28.90 ± 2.7 | 36.43 ± 4.2 | < 0.001 | |
| Forth | 28.80 ± 2.8 | 36.13 ± 3.8 | < 0.001 | |
| Friedman test results (Chi2, p value) | (47.02, < 0.001) | (4.81, 0.187) | . | |
|
| ||||
|
| First | 60.30 ± 6.7 | 62.03 ± 7.2 | 0.513 |
| Second | 51.67 ± 5.0 | 65.17 ± 5.9 | < 0.001 | |
| Third | 60.97 ± 5.1 | 66.80 ± 3.4 | < 0.001 | |
| Forth | 55.27 ± 7.7 | 69.03 ± 6.8 | < 0.001 | |
| Friedman test results (Chi2, p value) | (43.08, < 0.001) | (14.29, 0.003) | . | |
|
| ||||
|
| First | 85.03 ± 14.0 | 86.67 ± 13.7 | 0.587 |
| Second | 72.47 ± 9.3 | 86.50 ± 11.2 | < 0.001 | |
| Third | 75.57 ± 6.1 | 87.70 ± 8.1 | < 0.001 | |
| Forth | 74.23 ± 7.5 | 85.80 ± 7.2 | < 0.001 | |
| Friedman test results (Chi2, p value) | (21.44, < 0.001) | (2.59, 0.459) | . | |
aBy the Mann–Whitney test.
Figure 3Trends of hemodynamic parameters in the i-gel and air-Q groups.
Comparison of adverse events following the insertion of supraglottic airway devices in the i-gel and air-Q groups.
|
|
|
| |
|---|---|---|---|
|
| 3 (10%) | 6 (20%) | 0.472 |
|
| 0 (0%) | 6 (20%) | 0.024 |
|
| 3 (10%) | 6 (20%) | 0.472 |
aBy Fisher's exact test.