Mingyin Jiang1,2, Shenglin Liu1,2, Jiaqi Gao1,2, Qingmin Feng1,2, Qiang Zhang1,2.
Abstract
BACKGROUND Poor ergonomic design of ventilators can result in human errors. In this study, we evaluated the ergonomics of ventilators through respiratory therapists' performance, workload, and user experience. MATERIAL AND METHODS Sixteen respiratory therapists were recruited to this usability study of 3 ventilators. Participants had to perform 7 tasks on each ventilator. Respiratory therapists' performance was measured by task errors of all tasks for each participant. Workload was measured by objective measurement (blink rate and duration) and by subjective measurement (NASA-TLX). User experience was assessed by the USE Questionnaire. RESULTS For task errors, significant differences were found among ventilators (p<0.05) and the Evital 4 received higher task errors when compared to the Servo I (p<0.05). For blink rate, significant differences were found in tasks of starting the ventilator, ventilator monitoring values recognition, ventilator setting parameters modification, alarm parameter recognition, and resetting among ventilators (p<0.05). Furthermore, blink duration was also found to be significant differently in tasks of starting the ventilator, mode and setting parameters recognition, ventilator monitoring values recognition, ventilator mode modification, and alarm parameter recognition and resetting, as well as in the average of all tasks (p<0.05). For perceived workload, the Evital 4 received higher NASA-TLX scores among ventilators. For user experience, the Servo I received the highest scores on the USE Questionnaire among the ventilators. CONCLUSIONS The study provides a comprehensive evaluation method of user interface based on respiratory therapists' performance, workload, and user experience. In addition, this study suggests that the ergonomic design of the Evital 4 is poor. Finally, we found that eye motion (blink rate and duration) may be useful to assess the ergonomics of a user interface.Entities:
Mesh:
Year: 2018 PMID: 30552313 PMCID: PMC6319161 DOI: 10.12659/MSM.911853
Source DB: PubMed Journal: Med Sci Monit ISSN: 1234-1010
Blink rate for each task in ventilators and mean differences (MD=Mr1–Mr2) with the results of post hoc multiple comparisons of blink rate.
| Task | Mean blink rate for each task among ventilators | p | Post hoc tests with Bonferroni correction | p | ||||
|---|---|---|---|---|---|---|---|---|
|
| ||||||||
| Evital 4 | Servo I | Boaray 5000D | ||||||
|
|
| |||||||
| Mean±SD | Mean±SD | Mean±SD | Mr1 | Mr2 | MD (Mr1–Mr2) | |||
| Task 1: start the ventilator (blinks/second) | 0.201±0.557 | 0.115±0.593 | 0.122±0.067 | p<.01 | Servo I | Evital 4 | −0.086 | p<.01 |
|
| ||||||||
| Boaray 5000D | Evital 4 | −0.079 | 0.02 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | −0.007 | 1.00 | |||||
|
| ||||||||
| Task 2: mode and setting parameters recognition (blinks/second) | 0.164±0.078 | 0.201±0.542 | 0.184±0.035 | 0.82 | ||||
|
| ||||||||
| Task 3: ventilator monitoring values recognition (blinks/second) | 0.103±0.074 | 0.186±0.041 | 0.156±0.044 | 0.01 | Servo I | Evital 4 | 0.083 | p<.01 |
|
| ||||||||
| Boaray 5000D | Evital 4 | 0.053 | 0.34 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | 0.030 | 0.34 | |||||
|
| ||||||||
| Task 4: ventilator setting parameters modification (blinks/second) | 0.120±0.052 | 0.198±0.038 | 0.168±0.030 | 0.01 | Servo I | Evital 4 | 0.078 | 0.01 |
|
| ||||||||
| Boaray 5000D | Evital 4 | 0.048 | 0.23 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | 0.030 | 0.65 | |||||
|
| ||||||||
| Task 5: ventilator mode modification (blinks/second) | 0.105±0.047 | 0.146±0.036 | 0.136±0.049 | 0.08 | ||||
|
| ||||||||
| Task 6: alarm parameter recognition and resetting (blinks/second) | 0.108±0.051 | 0.151±0.067 | 0.179±0.056 | 0.02 | Servo I | Evital 4 | 0.043 | 0.16 |
|
| ||||||||
| Boaray 5000D | Evital 4 | 0.071 | 0.01 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | −0.028 | 1.00 | |||||
|
| ||||||||
| Task 7: respond to alarm (blinks/second) | 0.128±0.079 | 0.148±0.085 | 0.131±0.038 | 0.36 | ||||
|
| ||||||||
| Average blink rate of all tasks (second) | 0.132±0.029 | 0.163±0.016 | 0.153±0.015 | 0.06 | ||||
Positive MD values representing Mr1 lower workload than Mr2.
Blink duration for each task in ventilators and mean differences (MD=Mr1–Mr2) with the results of post hoc multiple comparisons of blink duration.
| Task | Mean blink rate for each task among ventilators | p | Post hoc tests with Bonferroni correction | p | ||||
|---|---|---|---|---|---|---|---|---|
|
| ||||||||
| Evital 4 | Servo I | Boaray 5000D | ||||||
|
|
| |||||||
| Mean±SD | Mean±SD | Mean±SD | Mr1 | Mr2 | MD (Mr1–Mr2) | |||
| Task1: start the ventilator (second) | 0.117±0.170 | 0.101±0.019 | 0.101±0.009 | p<.01 | Servo I | Evital 4 | −0.016 | 0.01 |
|
| ||||||||
| Boaray 5000D | Evital 4 | −0.016 | 0.02 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | 0.000 | 1.00 | |||||
|
| ||||||||
| Task2: mode and setting parameters recognition (second) | 0.097±0.014 | 0.123±0.026 | 0.111±0.026 | 0.02 | Servo I | Evital 4 | 0.026 | 0.02 |
|
| ||||||||
| Boaray 5000D | Evital 4 | 0.014 | 0.40 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | 0.012 | 0.65 | |||||
|
| ||||||||
| Task3: ventilator monitoring values recognition (second) | 0.091±0.010 | 0.123±0.010 | 0.111±0.020 | p<.01 | Servo I | Evital 4 | 0.032 | p<.01 |
|
| ||||||||
| Boaray 5000D | Evital 4 | 0.020 | 0.01 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | 0.012 | 1.00 | |||||
|
| ||||||||
| Task4: ventilator setting parameters modification (second) | 0.104±0.018 | 0.123±0.034 | 0.114±0.027 | 0.45 | ||||
|
| ||||||||
| Task5: ventilator mode modification (second) | 0.093±0.014 | 0.118±0.021 | 0.112±0.018 | 0.01 | Servo I | Evital 4 | 0.025 | 0.01 |
|
| ||||||||
| Boaray 5000D | Evital 4 | 0.019 | 0.28 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | 0.006 | 0.75 | |||||
|
| ||||||||
| Task6: alarm parameter recognition and resetting (second) | 0.094±0.013 | 0.109±0.019 | 0.118±0.022 | 0.03 | Servo I | Evital 4 | 0.015 | 0.23 |
|
| ||||||||
| Boaray 5000D | Evital 4 | 0.024 | 0.04 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | −0.009 | 1.00 | |||||
|
| ||||||||
| Task7: respond to alarm (second) | 0.097±0.023 | 0.112±0.025 | 0.104±0.023 | 0.29 | ||||
|
| ||||||||
| Average blink duration of all tasks (second) | 0.099±0.009 | 0.116±0.013 | 0.111±0.009 | p<.01 | Servo I | Evital 4 | 0.017 | 0.01 |
|
| ||||||||
| Boaray 5000D | Evital 4 | 0.005 | 0.01 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | 0.012 | 1.00 | |||||
Positive MD values representing Mr1 lower workload than Mr2.
NASA-TLX workload scores for each ventilator and mean differences (MD=Mr1–Mr2) with the results of post hoc multiple comparisons of NASA-TLX workload scores.
| NASA-TLX workload | NASA-TLX workload scores | p | Post hoc tests with Bonferroni correction | p | ||||
|---|---|---|---|---|---|---|---|---|
|
| ||||||||
| Evital 4 | Servo I | Boaray 5000D | ||||||
|
|
| |||||||
| Mean±SD | Mean±SD | Mean±SD | Mr1 | Mr2 | MD (Mr1–Mr2) | |||
| TLX (task load index) | 40.042±12.304 | 23.750±7.628 | 26.083±5.791 | p<.01 | Servo I | Evital 4 | −16.292 | p<.01 |
|
| ||||||||
| Boaray 5000D | Evital 4 | −13.959 | 0.02 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | −2.333 | 1.00 | |||||
|
| ||||||||
| Mental demand | 10.375±4.286 | 4.625±5.513 | 8.333±4.286 | p<.01 | Servo I | Evital 4 | −5.750 | p<.01 |
|
| ||||||||
| Boaray 5000D | Evital 4 | −2.042 | 0.28 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | −3.708 | 0.47 | |||||
|
| ||||||||
| Physical demand | 2.168±3.783 | 0.833±1.450 | 0.958±1.088 | 0.57 | ||||
|
| ||||||||
| Temporal demand | 8.083±5.158 | 6.250±4.021 | 6.083±3.593 | 0.50 | ||||
|
| ||||||||
| Performance | 8.958±5.817 | 3.792±3.300 | 4.667±2.271 | 0.01 | Servo I | Evital 4 | −5.166 | 0.02 |
|
| ||||||||
| Boaray 5000D | Evital 4 | −4.291 | 0.05 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | −0.875 | 1.00 | |||||
|
| ||||||||
| Effort | 7.625±5.497 | 5.333±3.578 | 3.333±3.070 | 0.08 | ||||
|
| ||||||||
| Frustration | 2.833±3.197 | 2.917±2.920 | 2.708±2.412 | 0.77 | ||||
Positive MD values representing Mr1 lower workload than Mr2.
The USE Questionnaire scores for each ventilator and mean differences (MD=Mr1–Mr2) with the results of post hoc multiple comparisons of the USE Questionnaire scores.
| User experience | The USE Questionnaire scores | P | Post hoc tests with Bonferroni correction | P | ||||
|---|---|---|---|---|---|---|---|---|
|
| ||||||||
| Evital 4 | Servo I | Boaray 5000D | ||||||
|
|
| |||||||
| Mean±SD | Mean±SD | Mean±SD | Mr1 | Mr2 | MD (Mr1–Mr2) | |||
| The overall average scores | 4.894±0.981 | 5.684±0.900 | 5.328±0.747 | p<.01 | Servo I | Evital 4 | 0.790 | p<.01 |
|
| ||||||||
| Boaray 5000D | Evital 4 | 0.434 | 0.34 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | 0.356 | 0.34 | |||||
|
| ||||||||
| Usefulness | 4.885±0.961 | 5.878±1.032 | 5.360±0.716 | p<.01 | Servo I | Evital 4 | 0.993 | p<.01 |
|
| ||||||||
| Boaray 5000D | Evital 4 | 0.475 | 0.56 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | 0.518 | 0.10 | |||||
|
| ||||||||
| Ease of use | 4.976±1.028 | 5.688±0.811 | 5.221±0.795 | 0.06 | ||||
|
| ||||||||
| Ease of learning | 4.828±0.835 | 5.563±0.824 | 5.188±0.854 | 0.07 | ||||
|
| ||||||||
| Satisfaction | 4.812±1.155 | 5.527±1.035 | 5.536±0.764 | 0.03 | Servo I | Evital 4 | 0.715 | 0.04 |
|
| ||||||||
| Boaray 5000D | Evital 4 | 0.724 | 0.65 | |||||
|
| ||||||||
| Servo I | Boaray 5000D | −0.009 | 0.65 | |||||
Positive MD values representing Mr1 lower workload than Mr2.
Randomization table for device testing.
| Participant number | Ventilator type | ||
|---|---|---|---|
| 1 | Boaray 5000D | Servo I | Evital 4 |
| 2 | Boaray 5000D | Evital 4 | Servo I |
| 3 | Servo I | Boaray 5000D | Evital 4 |
| 4 | Servo I | Evital 4 | Boaray 5000D |
| 5 | Evital 4 | Servo I | Boaray 5000D |
| 6 | Evital 4 | Boaray 5000D | Servo I |
| 7 | Servo I | Evital 4 | Boaray 5000D |
| 8 | Boaray 5000D | Servo I | Evital 4 |
| 9 | Servo I | Evital 4 | Boaray 5000D |
| 10 | Evital 4 | Servo I | Boaray 5000D |
| 11 | Servo I | Evital 4 | Boaray 5000D |
| 12 | Evital 4 | Servo I | Boaray 5000D |
| 13 | Evital 4 | Boaray 5000D | Servo I |
| 14 | Servo I | Boaray 5000D | Evital 4 |
| 15 | Evital 4 | Servo I | Boaray 5000D |
| 16 | Boaray 5000D | Servo I | Evital 4 |
This table details the randomization table for device testing by the respiratory therapists.