Evaluation of the ventilator-user interface of 2 new advanced compact transport ventilators.

François Templier1, Patrick Miroux, François Dolveck, Alexis Descatha, Nathalie-Sybille Goddet, Charles Jeleff, Michel Baer, Marcel Chauvin, Laurent Brochard, Dominique Fletcher.   

Abstract

BACKGROUND: Mechanical ventilation during patient transport frequently utilizes compact portable pneumatic ventilators that have limited ventilator-settings options. New advanced transport ventilators should yield quality improvements, but their user-friendliness needs to be tested.
OBJECTIVE: To evaluate the ventilator-user interface of 2 new transport ventilators.
METHODS: This was a 2-center descriptive study in which the ventilator-user interfaces of the Oxylog 3000 and Elisée 250 were compared by 20 French senior emergency physicians who were initially unfamiliar with these ventilators. Each physician carried out 15 tasks with each ventilator and then assigned each ventilator a satisfaction score.
RESULTS: With the Elisée 250 the task success rate was significantly higher (85.6% vs 66.6% with the Oxylog 3000, p < 0.0001), and the total number of errors was lower (46 vs 113). The main errors were related to inspiratory flow settings with the Oxylog 3000 (31 errors), inspiratory-expiratory ratio settings with the Elisée 250 (11 errors), ventilation mode choice with the Oxylog 3000 (17 errors), trigger sensitivity setting with the Elisée 250 (16 errors) and the Oxylog 3000 (11 errors), and alarm range setting with the Oxylog 3000 (10 errors). The mean satisfaction score was significantly better with the Elisée 250 (81% +/- 7, range 64-92%) than with the Oxylog 3000 (66% +/- 10, range 49-87%) (p < 0.0001).
CONCLUSIONS: The Elisée 250 ventilator-user interface was easier to use than that of the Oxylog 3000. The applicability of these results to other types of users will require further studies, but the types of errors found in our study might help future users.

Entities:  

Mesh:

Year:  2007        PMID: 18028560

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  4 in total

1.  A new global and comprehensive model for ICU ventilator performances evaluation.

Authors:  Nicolas S Marjanovic; Agathe De Simone; Guillaume Jegou; Erwan L'Her
Journal:  Ann Intensive Care       Date:  2017-06-17       Impact factor: 6.925

2.  Usability Study of the User-Interface of Intensive Care Ventilators Based on User Test and Eye-Tracking Signals.

Authors:  Mingyin Jiang; Shenglin Liu; Qingmin Feng; Jiaqi Gao; Qiang Zhang
Journal:  Med Sci Monit       Date:  2018-09-20

3.  Human factors/ergonomics to support the design and testing of rapidly manufactured ventilators in the UK during the COVID-19 pandemic.

Authors:  Sue Hignett; Janette Edmonds; Tracey Herlihey; Laura Pickup; Richard Bye; Emma Crumpton; Mark Sujan; Fran Ives; Daniel P Jenkins; Miranda Newbery; David Embrey; Paul Bowie; Chris Ramsden; Noorzaman Rashid; Alastair Williamson; Anne-Marie Bougeard; Peter MacNaughton
Journal:  Int J Qual Health Care       Date:  2021-01-12       Impact factor: 2.038

4.  The usability of ventilators: a comparative evaluation of use safety and user experience.

Authors:  Plinio P Morita; Peter B Weinstein; Christopher J Flewwelling; Carleene A Bañez; Tabitha A Chiu; Mario Iannuzzi; Aastha H Patel; Ashleigh P Shier; Joseph A Cafazzo
Journal:  Crit Care       Date:  2016-08-20       Impact factor: 9.097

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.