Evaluation of the user interface simplicity in the modern generation of mechanical ventilators.

Yoshihiro Uzawa1, Yoshitsugu Yamada, Masayuki Suzukawa.   

Abstract

OBJECTIVE: We designed this study to evaluate the simplicity of the user interface in modern-generation mechanical ventilators. We hypothesized that different designs in the user interface could result in different rates of operational failures.
SETTING: A laboratory in a tertiary teaching hospital.
DESIGN: Crossover design.
SUBJECTS: Twenty-one medical resident physicians who did not possess operating experience with any of the selected ventilators.
METHODS: Four modern mechanical ventilators were selected: Dräger Evita XL, Maquet Servo-i, Newport e500, and Puritan Bennett 840. Each subject was requested to perform 8 tasks on each ventilator. Two objective variables (the number of successfully completed tasks without operational failures and the operational time) and the overall subjective rating of the ease of use, measured with a 100-mm visual analog scale were recorded.
RESULTS: The total percentage of operational failures made for all subjects, for all tasks, was 23%. There were significant differences in the rates of operational failures and operational time among the 4 ventilators. Subjects made more operational failures in setting up the ventilators and in making ventilator-setting changes than in reacting to alarms. The subjective feeling of the ease of use was also significantly different among the ventilators.
CONCLUSION: The design of the user interface is relevant to the occurrence of operational failures. Our data indicate that ventilator designers could optimize the user-interface design to reduce the operational failures; therefore, basic user interface should be standardized among the clinically used mechanical ventilators.

Mesh:

Year:  2008        PMID: 18291049

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  4 in total

1.  Evaluation of the user-friendliness of seven new generation intensive care ventilators.

Authors:  Laurence Vignaux; Didier Tassaux; Philippe Jolliet
Journal:  Intensive Care Med       Date:  2009-07-29       Impact factor: 17.440

2.  ICU mechanical ventilators, technological advances vs. user friendliness: the right picture is worth a thousand numbers.

Authors:  Jean Christophe Marie Richard; Robert M Kacmarek
Journal:  Intensive Care Med       Date:  2009-07-29       Impact factor: 17.440

3.  A new global and comprehensive model for ICU ventilator performances evaluation.

Authors:  Nicolas S Marjanovic; Agathe De Simone; Guillaume Jegou; Erwan L'Her
Journal:  Ann Intensive Care       Date:  2017-06-17       Impact factor: 6.925

4.  Usability Study of the User-Interface of Intensive Care Ventilators Based on User Test and Eye-Tracking Signals.

Authors:  Mingyin Jiang; Shenglin Liu; Qingmin Feng; Jiaqi Gao; Qiang Zhang
Journal:  Med Sci Monit       Date:  2018-09-20
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.