PURPOSE: To investigate the anatomical and biomechanical feasibility of the unilateral C1 double screw [pedicle screw (PS) + lateral mass screw (LMS)] and ipsilateral C2 PS combined with contralateral C2 laminar screw (LS)-rod fixation for atlantoaxial instability by comparison with traditional posterior fixation methods.
METHODS: Fifteen sets of complete dry bony specimens of atlas were used for morphometric analysis. The working length, width and thickness of the C1 PSs and LMSs were manually measured. Ten fresh-frozen cervical spines (C0-C7) were used to complete the range of motion (ROM) testing in their intact condition, under destabilization and after stabilization by the following procedures: unilateral C1-C2 PS rod fixation (Group A), bilateral C1-C2 PS rod fixation (Group B), and unilateral C1 double screw and ipsilateral C2 PS combined with contralateral C2 LS rod fixation (Group C).
RESULTS: The working thickness of the C1 PS was ≤ 3.5 mm in only one (1/15 = 6.7%) specimen. The other parameters were > 3.5 mm in all specimens. In the ROM test, all fixation groups showed significantly reduced flexibility in all directions compared with both the intact and destabilization groups. Further, Groups B and C showed better stability in all directions than Group A. However, no significant differences were observed between Groups B and C.
CONCLUSION: The C1 unilateral lateral mass could mostly contain two screws(PS + LMS) with diameters ≤ 3.5 mm. The novel technique of unilateral C1 double screw and ipsilateral C2 PS combined with contralateral C2 LS rod fixation provided better stability than unilateral PS rod fixation and similar as bilateral PS rod fixation. Therefore, it is a feasible salvage method that provides a new insight into atlantoaxial instability. These slides can be retrieved under Electronic Supplementary Material.

Entities:  

Keywords:  Atlantoaxial instability; Biomechanics; C1 double screw; Morphometric analysis; Unilateral fixation

Mesh:

Year:  2018        PMID: 30539243     DOI: 10.1007/s00586-018-5853-0

Source DB:  PubMed          Journal:  Eur Spine J        ISSN: 0940-6719            Impact factor:   3.134


  24 in total

1.  Posterior atlantoaxial fixation: biomechanical in vitro comparison of six different techniques.

Authors:  Marcus Richter; René Schmidt; Lutz Claes; Wolfhart Puhl; Hans-Joachim Wilke
Journal:  Spine (Phila Pa 1976)       Date:  2002-08-15       Impact factor: 3.468

2.  Posterior C2 fixation using bilateral, crossing C2 laminar screws: case series and technical note.

Authors:  Neill M Wright
Journal:  J Spinal Disord Tech       Date:  2004-04

3.  Posterior atlantoaxial transpedicular screw and plate fixation. Technical note.

Authors:  Jyi-Feng Chen; Chieh-Tsai Wu; Sai-Cheung Lee; Shih-Tseng Lee
Journal:  J Neurosurg Spine       Date:  2005-03

4.  The feasibility of inserting atlas lateral mass screws via the posterior arch.

Authors:  Michael J Lee; Ezequiel Cassinelli; K Daniel Riew
Journal:  Spine (Phila Pa 1976)       Date:  2006-11-15       Impact factor: 3.468

5.  Biomechanical comparison of unicortical versus bicortical C1 lateral mass screw fixation.

Authors:  Jason C Eck; Matt P Walker; Bradford L Currier; Qingshan Chen; Michael J Yaszemski; Kai-Nan An
Journal:  J Spinal Disord Tech       Date:  2007-10

6.  C1 anatomy and dimensions relative to lateral mass screw placement.

Authors:  David M Christensen; Robert K Eastlack; James J Lynch; Michael J Yaszemski; Bradford L Currier
Journal:  Spine (Phila Pa 1976)       Date:  2007-04-15       Impact factor: 3.468

7.  Morphometric evaluation of screw fixation in atlas via posterior arch and lateral mass.

Authors:  Mingsheng Tan; Huimin Wang; Yunting Wang; Guangbo Zhang; Ping Yi; Zirong Li; Hongyu Wei; Feng Yang
Journal:  Spine (Phila Pa 1976)       Date:  2003-05-01       Impact factor: 3.468

8.  Anatomic considerations for the pedicle screw placement in the first cervical vertebra.

Authors:  Xiang-Yang Ma; Qing-Shui Yin; Zeng-Hui Wu; Hong Xia; Jing-Fa Liu; Shi-Zhen Zhong
Journal:  Spine (Phila Pa 1976)       Date:  2005-07-01       Impact factor: 3.468

9.  C1-C2 pedicle screw fixation with rigid cantilever beam construct: case report and technical note.

Authors:  Daniel K Resnick; Edward C Benzel
Journal:  Neurosurgery       Date:  2002-02       Impact factor: 4.654

10.  Posterior C1-C2 fusion with polyaxial screw and rod fixation.

Authors:  J Harms; R P Melcher
Journal:  Spine (Phila Pa 1976)       Date:  2001-11-15       Impact factor: 3.468

View more
  2 in total

1.  A biomechanical comparison of crossed and parallel rod configurations in atlantoaxial internal fixation.

Authors:  Feng Qiu; Xiao-Bao Zou; Xi-Lin Xu; Wei-Cheng Jiang; Geng-Chao Liu; Xiang-Yang Ma
Journal:  Eur Spine J       Date:  2020-11-12       Impact factor: 3.134

2.  Translaminar screw of C1 for the reinforcement of subaxial cervical spine reconstruction.

Authors:  Luis E Carelli; Alderico Girão; Juan P Cabrera
Journal:  J Craniovertebr Junction Spine       Date:  2022-06-13
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.