| Literature DB >> 30519933 |
Edward William Johnston1, Arash Latifoltojar1, Harbir Singh Sidhu1, Navin Ramachandran1, Magdalena Sokolska2, Alan Bainbridge2, Caroline Moore3, Hashim Uddin Ahmed4, Shonit Punwani5.
Abstract
OBJECTIVES: To determine the diagnostic accuracy and interobserver concordance of whole-body (WB)-MRI, vs. 99mTc bone scintigraphy (BS) and 18fluoro-ethyl-choline (18F-choline) PET/CT for the primary staging of intermediate/high-risk prostate cancer.Entities:
Keywords: Choline; Magnetic resonance imaging; Positron emission tomography; Prostate
Mesh:
Year: 2018 PMID: 30519933 PMCID: PMC6510859 DOI: 10.1007/s00330-018-5813-4
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1Patient recruitment flow diagram
Whole-body MRI acquisition parameters.
| Imaging plane | T2-TSE | mDixon (pre- and post-contrast) | DWI (b0, b100, b300, b1000) |
|---|---|---|---|
| Transverse | Coronal | Transverse | |
| TE (ms) | 80 | 1.02/1.8 | 71 |
| TR (ms) | 1228 | 3.0 | 6371 |
| FOV (mm × mm) | 500 × 300 | 502 × 300 | 500 × 306 |
| Voxel size (mm × mm) | 1 × 1 | 2.1 × 2.1 | 4 × 4.2 |
| Number of slices | 40 | 120 | 40 |
| Slice thickness (mm) | 5 | 5 | 5 |
| Acquisition matrix | 500 × 286 | 144 × 238 | 124 × 72 |
| ETL | 91 | 2 | 39 |
| Acceleration factor (SENSE) | 2 | 2 | 2.5 |
| Pixel bandwidth (Hz) | 537 | 1992 | 3369 |
| Scan time (min) | 15.2 | 5.5 × 2 | 47 |
T2-TSE T2-weighted turbo spin echo, mDixon modified Dixon, DWI diffusion-weighted imaging, TE time of echo, TR time of repetition, FOV field of view, ETL echo train length, SENSE sensitivity encoding
Fig. 2Flow diagrams of the statistical methods used in the study. a Flow diagram of intermodality and inter-reader concordance. b Flow diagram for WB-MRI LSR ROC-AUC and intermodality diagnostic accuracy analyses. c Patient-based reference standard
Distribution of lesions on each imaging modality. The first row for each N/M stage represents a comparison between patients who underwent BS, 18F-choline PET/CT, and WB-MRI (n = 33), and the second row represents a comparison between the patients who underwent BS and WB-MRI (n = 56)
| BS | 18F-choline PET/CT | WB-MRI | |
|---|---|---|---|
| N0 | – | 23/33 (69.7%) | 26/33 (78.8%) |
| – | – | 43/56 (76.8%) | |
| N1 | – | 10/33 (30.3%) | 7/33 (21.2%) |
| – | – | 13/56 (23.2%) | |
| M0 | 30/33 (90.9%) | 21/33 (63.6%) | 22/33 (66.7%) |
| 43/56 (76.8%) | – | 34/56 (60.7%) | |
| M1a | – | 6/33 (18.2%) | 3/33 (9.1%) |
| – | – | 6/56 (10.7%) | |
| M1b | 3/33 (9.1%) | 6/33 (18.2%) | 8/33 (24.2%) |
| 13/56 (23.2%) | – | 16/56 (28.6%) | |
| M1c | – | 0/33 (0%) | 0/33 (0%) |
| – | – | 0/56 (0%) |
Interobserver and intermodality concordances
| Local nodes (N1) | Metastatic nodes (M1a) | Metastatic bones (M1b) | |
|---|---|---|---|
| Interobserver concordance for WB-MRI ( | 0.79 | 0.68 | 0.58 |
| Concordance of WB-MRI vs. BS ( | – | – | 0.68 |
| Concordance of WB-MRI vs. 18F-choline PET/CT ( | 0.77 | 0.37 | 0.64 |
WB-MRI whole-body MRI, BS99mTc bone scintigraphy
ROC-AUC for each component of the LSR, performed as part of a patient-based analysis according to the reference standard
| ROC-AUC | N0/N1 ( | M1a ( | M1b ( | Mean ROC-AUC |
|---|---|---|---|---|
| Reader 1 | ||||
| mDixon + DWI | 0.97 (0.91–1.00) | 0.99 (0.96–1.00) | 0.86 (0.72–1.00) | 0.94 |
| + T2-TSE | 0.98 (0.94–1.00) | 0.99 (0.95–1.00) | 0.93 (0.84–1.00) | 0.96 |
| + contrast | 0.98 (0.94–1.00) | 0.97 (0.91–1.00) | 0.90 (0.76–1.00) | 0.95 |
| Reader 2 | ||||
| mDixon + DWI | 0.94 (0.81–1.00) | 0.87 (0.60–1.00) | 0.86 (0.73–1.00) | 0.89 |
| + T2-TSE | 0.94 (0.82–1.00) | 0.87 (0.60–1.00) | 0.94 (0.83–1.00) | 0.91 |
| + contrast | 0.94 (0.82–1.00) | 0.87 (0.60–1.00) | 0.93 (0.82–1.00) | 0.91 |
ROC-AUC receiver operator characteristic area under the curve, LSR locked sequential read, TSE turbo spin echo. Figures in parentheses represent 95% confidence intervals
Performance characteristics of BS, 18F-choline PET/CT, and WB-MRI carried out as a patient-based analysis vs. the reference standard
| N1 ( | M1a ( | M1b ( | |
|---|---|---|---|
| BS | |||
| Sensitivity | – | – | 0.60 (3/5) |
| Specificity | 1.00 (13/13) | ||
| PPV | 1.00 (3/3) | ||
| NPV | 0.87 (13/15) | ||
| 18F-choline PET/CT | |||
| Sensitivity | 1.00 (7/7) | 0.75 (3/4) | 0.80 (4/5) |
| Specificity | 0.82 (9/11) | 0.92 (12/13) | 0.92 (12/13) |
| PPV | 0.77 (7/9) | 0.75 (3/4) | 0.80 (4/5) |
| NPV | 1.00 (9/9) | 0.92 (12/13) | 0.92 (12/13) |
| WB-MRI: reader 1 | |||
| Sensitivity | 1.00 (7/7) | 1.00 (4/4) | 0.80 (4/5) |
| Specificity | 0.91 (10/11) | 0.85 (11/13) | 1.00 (13/13) |
| PPV | 0.88 (7/8) | 0.67 (4/6) | 1.00 (4/4) |
| NPV | 1.00 (10/10) | 1.00 (11/11) | 0.93 (13/14) |
| WB-MRI: reader 2 | |||
| Sensitivity | 1.00 (7/7) | 0.50 (2/4) | 1.00 (5/5) |
| Specificity | 1.00 (11/11) | 1.00 (13/13) | 0.76 (10/13) |
| PPV | 1.00 (7/7) | 1.00 (2/2) | 0.62 (5/8) |
| NPV | 1.00 (11/11) | 0.86 (13/2) | 1.00 (10/10) |
| WB-MRI mean | |||
| Sensitivity | 1.00 | 0.75 | 0.90 |
| Specificity | 0.96 | 0.93 | 0.88 |
| PPV | 0.94 | 0.83 | 0.81 |
| NPV | 1.00 | 0.93 | 0.97 |
WB-MRI whole-body MRI, BS99mTc bone scintigraphy, NPV negative predictive value, PPV positive predictive value. WB-MRI mean is the mean of the two readers
Fig. 3Example of discordant skeletal site in a 64-year-old man with a serum prostate-specific antigen level of 25.2. Top left: Negative 99mTc bone scintigram (BS), Top right: 18F-choline PET/CT showing an area of increased tracer avidity at the right inferior pubic ramus consistent with a metastasis. Bottom left: Axial T2W TSE showing the lesion is of low signal intensity. Bottom right: the lesion has increased diffusion-weighted signal on b = 1000s/mm2. The lesion was considered as an example of a true positive whole-body MRI, true positive 18F-choline PET/CT, and false negative BS
Fig. 4Example of a discordant skeletal site in a 73-year-old man with a serum prostate-specific antigen level of 19.4. Top: Axial T2-weighted turbo spin echo with fused b = 1000s/mm2 showing a metastasis in the left iliac body, which is occult on BS (middle) and 18F-choline PET/CT (bottom). The site was considered to represent a true positive whole-body MRI, false negative 18F-choline PET/CT, and false negative 99mTc bone scintigram