| Literature DB >> 30519571 |
Sunghee Joo1, Kyeongjin Lee2, Changho Song1.
Abstract
BACKGROUND: The monitoring and rehabilitation of pulmonary function can be immensely important for long-term performance of daily life activities in stroke patients. In recent times, smartphone game-based assessment (SGA) has been gaining in popularity as an alternative to laboratory assessments. Hence, the aims of this study were (1) to quantify the reliability and validity of SGA for pulmonary function and (2) to assess the validity of SGA in comparison to spirometry. MATERIALS/Entities:
Mesh:
Year: 2018 PMID: 30519571 PMCID: PMC6241341 DOI: 10.1155/2018/2439312
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1Two instrument systems: (a) spirometry (b) smartphone game application.
The mean and standard deviations of parameters of spirometry and SG.
| Parameters | Mean ± SD | |
|---|---|---|
| Spirometer | FVC (L) | 3.26 ± 1.07 |
| FVC (%) | 86.91 ± 25.92 | |
| FEV1 (L) | 2.47 ± 0.66 | |
| FEV1 (%) | 79.30 ± 18.36 | |
|
| ||
| GA | Real-time (s) | 8.5 ± 0.69 |
FVC=forced vital capacity; FEV1=forced expiratory volume in 1 second; GA=game-based assessment; SG=smartphone game.
Mean ± SD and reliability measured with smartphone-game application.
| Mean ± SD | ICC (2,1) 95%CI | 95% LOA | SEM | SEM (%) | MDC | MDC (%) | CV (%) | |
|---|---|---|---|---|---|---|---|---|
| Intra-test | 8.4 ± 0.92 | 0.84 | 0.30~2.33 | 0.37 | 4.41 | 1.02 | 12.22 | 24.84 |
| (0.65-0.93) | ||||||||
|
| ||||||||
| Inter-test | 8.9 ± 0.18 | 0.96 | 1.66~3.80 | 0.3 | 0.41 | 0.10 | 1.16 | 15.04 |
| (0.91-0.98) | ||||||||
CV=coefficients of variation of method error; ICC=intracorrelation coefficient; LOA=limits of agreement; MDC=minimum detectable change; SEM=standard error of measurement.
Concurrent validity measured with spirometry and smartphone-game application.
| Parameters | ICC (2,1) 95%CI | 95% LOA | CV ( %) |
|---|---|---|---|
| FVC - GA | 0.91 (0.83-0.95) | 75.35~142.22 | 19.34 |
|
| |||
| FVC % - GA | 0.92 (0.83-0.96) | 74.22~140.03 | 19.03 |
|
| |||
| FEV1 – GA | 0.93 (0.86-0.96) | 67.12~126.72 | 17.23 |
|
| |||
| FEV1 % - GA | 0.94 (0.89-0.97) | 61.67~116.37 | 15.82 |
CV=coefficients of variation of method error; FVC=forced vital capacity; FEV1=forced expiratory volume in 1 second; GA=game-based assessment; SG=smartphone game; ICC=intracorrelation coefficient; LOA=limits of agreement.
Figure 2Bland-Altman Plot.