| Literature DB >> 30510796 |
Yu-Chuan Wang1, Bo Huang1, Kang Li2, Peng-Kang He2, Er-Dong Chen2, Yu-Long Xia2, Jie Jiang2, Qin-Hui Sheng2, Jing Zhou2, Yan-Sheng Ding2.
Abstract
OBJECTIVES: To identify optimal predefined criteria (OPC) for filters of the VisiTag™ module in the CARTO 3 system during pulmonary vein isolation (PVI).Entities:
Year: 2018 PMID: 30510796 PMCID: PMC6231390 DOI: 10.1155/2018/8960941
Source DB: PubMed Journal: Cardiol Res Pract ISSN: 2090-0597 Impact factor: 1.866
Mean values of CF, FTI, and ID in each ablation segment after the initial wide-area circumferential ablation around ipsilateral PVs without FOT setting.
| Segments | CF (g) | FTI (g·s) | ID (Ω) |
|---|---|---|---|
| Left superior | 11.88 ± 5.90 | 512.49 ± 172.83 | 13.16 ± 6.16 |
| Left anterior superior | 8.91 ± 3.73 | 418.03 ± 148.63 | 12.15 ± 5.67 |
| Left anterior | 10.39 ± 4.18 | 427.16 ± 163.99 | 14.51 ± 6.39 |
| Left anterior inferior | 9.06 ± 3.49 | 427.26 ± 153.31 | 12.55 ± 6.19 |
| Left inferior | 11.38 ± 4.44 | 441.65 ± 152.46 | 14.02 ± 5.33 |
| Left posterior inferior | 14.04 ± 5.68 | 495.35 ± 174.51 | 13.13 ± 4.84 |
| Left posterior | 11.76 ± 5.15 | 445.90 ± 143.90 | 11.24 ± 4.89 |
| Left posterior superior | 11.83 ± 5.33 | 477.17 ± 152.47 | 12.13 ± 5.58 |
| Right superior | 14.29 ± 4.68 | 552.94 ± 140.85 | 15.46 ± 6.21 |
| Right anterior superior | 14.02 ± 4.63 | 575.31 ± 149.48 | 17.67 ± 5.71 |
| Right anterior | 17.89 ± 5.65 | 615.66 ± 153.57 | 15.85 ± 450 |
| Right anterior inferior | 11.56 ± 4.96 | 514.95 ± 213.81 | 14.22 ± 8.50 |
| Right inferior | 11.81 ± 5.51 | 482.34 ± 181.52 | 14.91 ± 6.85 |
| Right posterior inferior | 11.92 ± 5.83 | 511.77 ± 203.11 | 11.50 ± 4.78 |
| Right posterior | 12.96 ± 5.09 | 517.56 ± 153.14 | 10.78 ± 4.48 |
| Right posterior superior | 13.77 ± 6.92 | 545.55 ± 239.42 | 12.00 ± 5.81 |
CF: contact force; FTI: force time integral; ID: impedance drop.
Mean values of CF, FTI, and ID in each ablation segment after the initial wide-area circumferential ablation around ipsilateral PVs with FOT setting being minimal force 5 g with 50% stability time.
| Segments | CF (g) | FTI (g·s) | ID (Ω) |
|---|---|---|---|
| Left superior | 12.28 ± 5.58 | 537.32 ± 148.71 | 13.58 ± 5.61 |
| Left anterior superior | 10.06 ± 3.08 | 489.00 ± 141.33 | 12.89 ± 5.16 |
| Left anterior | 11.19 ± 3.57 | 491.10 ± 137.84 | 15.41 ± 5.51 |
| Left anterior inferior | 9.68 ± 3.01 | 467.32 ± 144.73 | 13.00 ± 5.66 |
| Left inferior | 11.65 ± 4.17 | 456.23 ± 137.10 | 14.21 ± 5.04 |
| Left posterior inferior | 14.04 ± 5.68 | 495.35 ± 174.51 | 13.13 ± 4.84 |
| Left posterior | 11.76 ± 5.15 | 445.90 ± 143.90 | 11.24 ± 4.89 |
| Left posterior superior | 11.83 ± 5.33 | 477.17 ± 152.47 | 12.13 ± 5.58 |
| Right superior | 14.53 ± 4.22 | 557.19 ± 130.73 | 15.61 ± 5.96 |
| Right anterior superior | 14.02 ± 4.63 | 575.31 ± 149.48 | 17.67 ± 5.71 |
| Right anterior | 17.89 ± 5.65 | 615.66 ± 153.57 | 15.85 ± 450 |
| Right anterior inferior | 11.90 ± 4.65 | 536.34 ± 197.24 | 14.46 ± 8.27 |
| Right inferior | 12.36 ± 5.03 | 514.32 ± 157.16 | 15.36 ± 6.26 |
| Right posterior inferior | 12.96 ± 4.79 | 559.48 ± 167.37 | 12.19 ± 4.26 |
| Right posterior | 12.96 ± 5.09 | 517.56 ± 153.14 | 10.78 ± 4.48 |
| Right posterior superior | 14.09 ± 6.63 | 565.96 ± 229.08 | 12.27 ± 5.53 |
CF: contact force; FTI: force time integral; ID: impedance drop.
Figure 1An example of different parameter settings of the VisiTag™ module for pulmonary vein isolation: a 3 mm maximal moving range of the catheter and 20 s minimal stability time of the ablation catheter were constant. (a) FOT setting being zero; (b) FOT setting being minimal force 5 g with 50% stability time; (c) FOT setting being minimal force 10 g with 50% stability time. Different color lines indicated different segments of ablation line. White arrow indicated a TG at the situation of FOT setting being a minimal force of 5 g with 50% stability time, which did not display when FOT setting is zero. FOT: force over time; LEAL: left encircling ablation line; REAL: right encircling ablation line.