Significant Delay in the Detection of Desaturation between Finger Transmittance and Earlobe Reflectance Oximetry Probes during Fiberoptic Bronchoscopy: Analysis of 104 Cases.

Barak Pertzov1,2,3, Elitsur Brachfeld4,5, Avraham Unterman6,5, Evgeni Gershman6,5, Nader Abdel-Rahman6,5, Dror Rosengarten6,5, Mordechai Reuven Kramer6,5.   

Abstract

PURPOSE: There is clinical significance to a delay in response time for detecting desaturation by pulse oximetry. Our aim in this study was to compare the response time of the reflectance and transmittance saturation probes during fiberoptic bronchoscopy (FOB) under monitored anesthesia care.
METHODS: A prospective study included 104 patients scheduled for FOB. Patients were monitored with transmittance (finger) and reflectance (ear) oximetry probes. The response time was evaluated during desaturation and resaturation. We also acquired blood tests for arterial oxygen saturation to assess the agreement with the oximetry probes.
RESULTS: Ninety patients had a desaturation episode during FOB and were included in the final analysis. Mean time difference between the reflectance ear probe (reference probe) and transmittance finger probe for the detection of desaturation (SpO2 = 90%) was + 36 s (CI 27.0-45.0, P < 0.001). The time difference between probes at end of desaturation episode (SpO2 = 95%) was + 31 s (CI 19.0-43.0; P < 0.001). A significant difference in response time was evident throughout the episode in all saturation values. The reflectance ear probe showed better agreement with arterial blood gases. The bias (and precision) for the earlobe and finger oximeters were of 0.24 (1.04) and 2.31 (3.37), respectively.
CONCLUSION: The data displayed by a centrally located reflectance probe are more accurate and allows for earlier identification, treatment, and resolution of desaturation events. In light of these data and the added value of the reflectance probe ability to measure transcutaneous PCO2, we recommend monitoring bronchoscopy by a reflectance oximetry probe.

Entities:  

Keywords:  Bronchoscopy; Delay; Oxygen saturation; Reflectance; Transmittance

Mesh:

Substances:

Year:  2018        PMID: 30474708     DOI: 10.1007/s00408-018-0180-0

Source DB:  PubMed          Journal:  Lung        ISSN: 0341-2040            Impact factor:   2.584


  26 in total

1.  Propofol safety in bronchoscopy: prospective randomized trial using transcutaneous carbon dioxide tension monitoring.

Authors:  Uri Carmi; Mordechai R Kramer; Dmitry Zemtzov; Dror Rosengarten; Oren Fruchter
Journal:  Respiration       Date:  2011-09-16       Impact factor: 3.580

2.  Forehead SpO2 monitoring compared to finger SpO2 recording in emergency transport.

Authors:  M Nuhr; K Hoerauf; A Joldzo; N Frickey; R Barker; L Gorove; T Puskas; A Kober
Journal:  Anaesthesia       Date:  2004-04       Impact factor: 6.955

Review 3.  Pulse oximetry.

Authors:  K K Tremper; S J Barker
Journal:  Anesthesiology       Date:  1989-01       Impact factor: 7.892

4.  Changes in oxygen saturation and transcutaneous carbon dioxide and oxygen levels in patients undergoing fibreoptic bronchoscopy.

Authors:  E N Evans; K Ganeshalingam; P Ebden
Journal:  Respir Med       Date:  1998-05       Impact factor: 3.415

5.  Errors in 14 pulse oximeters during profound hypoxia.

Authors:  J W Severinghaus; K H Naifeh; S O Koh
Journal:  J Clin Monit       Date:  1989-04

6.  A comparison of transmittance and reflectance pulse oximetry during vascular surgery.

Authors:  David B Wax; Philip Rubin; Steven Neustein
Journal:  Anesth Analg       Date:  2009-12       Impact factor: 5.108

7.  [Are transcutaneous blood gas measurement monitors (SenTec Digital Monitor system) useful for the respiratory management of chronic respiratory failure?].

Authors:  Rie Yazawa; Masahiro Gamo; Atsuko Kokawa; Yoshifumi Hirose
Journal:  Masui       Date:  2008-04

8.  Comparison of forehead and digit oximetry in surgical/trauma patients at risk for decreased peripheral perfusion.

Authors:  Lynn Schallom; Carrie Sona; Maryellen McSweeney; John Mazuski
Journal:  Heart Lung       Date:  2007 May-Jun       Impact factor: 2.210

Review 9.  Pulse oximetry for perioperative monitoring.

Authors:  Tom Pedersen; Amanda Nicholson; Karen Hovhannisyan; Ann Merete Møller; Andrew F Smith; Sharon R Lewis
Journal:  Cochrane Database Syst Rev       Date:  2014-03-17

10.  Complications of bronchoscopy: A concise synopsis.

Authors:  David L Stahl; Kathleen M Richard; Thomas J Papadimos
Journal:  Int J Crit Illn Inj Sci       Date:  2015 Jul-Sep
View more
  3 in total

1.  Assessment of Systemic and Cerebral Oxygen Saturation during Diagnostic Bronchoscopy: A Prospective, Randomized Study.

Authors:  Attila Vaskó; Sándor Kovács; Béla Fülesdi; Csilla Molnár
Journal:  Emerg Med Int       Date:  2020-12-09       Impact factor: 1.112

2.  Comparison of Ring Pulse Oximetry Using Reflective Photoplethysmography and PSG in the Detection of OSA in Chinese Adults: A Pilot Study.

Authors:  Rui Zhao; Jianbo Xue; Xueli Zhang; Maohuan Peng; Jing Li; Bing Zhou; Long Zhao; Thomas Penzel; Meir Kryger; Xiao Song Dong; Zhancheng Gao; Fang Han
Journal:  Nat Sci Sleep       Date:  2022-08-18

3.  Investigation of Photoplethysmography Behind the Ear for Pulse Oximetry in Hypoxic Conditions with a Novel Device (SPYDR).

Authors:  Brian Bradke; Bradford Everman
Journal:  Biosensors (Basel)       Date:  2020-04-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.