Seong Hwan Kim1, Yong-Beom Park1, Min-Ku Song1, Jung-Won Lim1, Han-Jun Lee1.
Abstract
PURPOSE: To evaluate the reliability and validity of the femorotibial mechanical axis angle from radiographs in the weight bearing (WB) and supine positions compared with navigation-measured values.Entities:
Keywords: Arthroplasty; Computer assisted surgery; Knee; Radiograph; Reliability; Validity
Year: 2018 PMID: 30466253 PMCID: PMC6254869 DOI: 10.5792/ksrr.18.028
Source DB: PubMed Journal: Knee Surg Relat Res ISSN: 2234-0726
Fig. 1Preoperative whole leg radiographs (WLRs) in the supine position (A) and weight bearing position (B). Postoperative WLRs in the supine position (C) and weight bearing position (D).
Fig. 2Navigation values of the mechanical axis during operation (A) and after prosthesis implantation (B).
Analysis of Radiographic and Navigation Measurements of the Lower Limb Alignment
| Variable | Mean±SD | 95% CI | p-value |
|---|---|---|---|
| Preop value | |||
| Preop supine radiograph (A) | 7.7±6.9 | 6.5 to 8.9 | B (p=0.000) |
| Preop WB radiograph (B) | 10.9±5.2 | 9.5 to 12.3 | A (p=0.000) |
| Initial navigation (C) | 10.0±6.8 | 8.4 to 11.6 | A (p=0.000) |
| Discrepancy between supine radiograph and navigation (D) | 3.3±2.6 | 2.7 to 3.9 | E (p=0.0001) |
| Discrepancy between WB radiograph and navigation (E) | 1.9±1.3 | 1.6 to 2.2 | D (p=0.0001) |
| Postop value | |||
| Postop supine radiograph | 0.4±2.6 | −0.2 to 1.0 | 0.098 |
| Postop WB radiograph | 0.5±2.9 | −0.2 to 1.2 | |
| Postop navigation | 0.1±1.1 | −0.2 to 0.4 | |
| Discrepancy between supine radiograph and navigation | 0.6±1.8 | 0.2 to 1.0 | 0.46 |
| Discrepancy between WB radiograph and navigation | 0.8±1.8 | 0.4 to 1.2 | |
| Outlier over 3° (case) | |||
| Postop supine radiograph | 9 | 0.279 | |
| Postop WB radiograph | 11 | ||
| Postop navigation | 5 | ||
SD: standard deviation, CI: confidence interval, Preop: preoperative, WB: weight bearing, Postop: postoperative.
p<0.05 was shown between pre- and postoperative values.
Analysis of the Radiograph-to-Navigation Correlation for Lower Limb Alignment Measurement
| Variable | Mean±SD (°) | Femorotibial mechanical axis on radiograph: navigation | ||
|---|---|---|---|---|
|
| ||||
| One-way ANOVA (p-value) | PCC (p-value) | ICC (95% CI) | ||
| Initial navigation value | 10.0±6.8 | - | ||
| Preop supine radiograph | 7.7±6.9 | 0.000 | 0.852 (0.000) | 0.743 (0.513–0.845) |
| Preop WB radiograph | 10.9±5.2 | 0.079 | 0.876 (0.000) | 0.818 (0.712–0.898) |
| Postop navigation value | 0.1±1.1 | |||
| Postop supine radiograph | 0.4±2.6 | 0.098 | 0.251 (0.04) | 0.093 (0.034–0.135) |
| Postop WB radiograph | 0.5±2.9 | 0.283 (0.02) | 0.182 (0.054–0.246) | |
SD: standard deviation, ANOVA: analysis of variance, PCC: Pearson correlation coefficient, ICC: intraclass correlation coefficient, CI: confidence interval, Preop: preoperative, WB: weight bearing, Postop: postoperative.
Standard Error of Measurement (SEM), 95% confidence interval (CI) for SEM for Precision, and 95% Limits of Agreement (LOA)
| Variable | Mean±SD | Overall radiograph: navigation value | ||
|---|---|---|---|---|
|
| ||||
| SEM | 95% CI for SEM | 95% LOA | ||
| Preop supine radiograph (°) | 7.7±6.9 | 2.6 | 2.6 to 12.8 | −4.7 to 9.3 |
| Preop WB radiograph (°) | 10.9±5.2 | 1.8 | 7.3 to 14.5 | −7.6 to 5.7 |
| Postop supine radiograph (°) | 0.4±2.6 | 2.3 | −4.1 to 4.9 | −4.6 to 5.7 |
| Postop WB radiograph (°) | 0.5±2.9 | 2.5 | −4.3 to 5.3 | −5.9 to 5.1 |
SD: standard deviation, Preop: preoperative, WB: weight bearing, Postop: postoperative.
Fig. 3Bland-Altman plots for agreement of navigation and preoperative measurements. (A) Plot for preoperative supine radiographic and navigation values. (B) Plot for preoperative standing radiographic and navigation values. The Bland-Altman plot showed less discrepancy between weight bearing whole leg radiographs (WLRs) and navigation values, but the range of 95% limits of agreement was not acceptable for both the supine-positioned WLRs and weight bearing WLRs. SD: standard deviation.
Fig. 4Bland-Altman plots for agreement between navigation and postoperative measurements. (A) Plot for postoperative supine radiographic values and navigation values. (B) Plot for postoperative standing radiographic values and navigation values. The Bland-Altman plot showed similar discrepancy between the supine/standing radiographic values and navigation values, and the range of 95% limits of agreement was not acceptable. Moreover, the radiographic values and navigation values showed the proportional error of measurements. SD: standard deviation.