| Literature DB >> 34666177 |
Saurabh Ajit Deshpande1, Ameya Udyavar2.
Abstract
Pacemaker-troubleshooting is an important step in the evaluation of a patient with syncope post-pacemaker-implantation. The basic functions of sensing, pacing and impedance may remain spuriously normal in the case of lead-microfracture or insulation break. We report a case in which the lead dysfunction was diagnosed based on multiple episodes of premature ventricular beats.Entities:
Keywords: Lead dysfunction; Lead revision; Normal lead impedance; Ventricular tachycardia
Year: 2021 PMID: 34666177 PMCID: PMC8811311 DOI: 10.1016/j.ipej.2021.10.003
Source DB: PubMed Journal: Indian Pacing Electrophysiol J ISSN: 0972-6292
Fig. 1Interrogation summary.
Fig. 2Ventricular high rate event.
Differential diagnosis of lead noise.
| Cause | Features | |
|---|---|---|
History of EMI source exposure. Noted in all channels and has high-frequency continuous signal. | ||
Noted on ventricular or Shock EGM High Frequency and low amplitude Can be reproduced with deep inspiration, Valsalva, coughing | ||
Provoked with the isometric exercise of the upper limb or abdominal muscles Noise is absent when the can is removed from the circuit | ||
High threshold and impedance E/o fracture on X-ray Reproduction of noise on manipulation of pocket | ||
High threshold and low impedance E/o loose set screw on X-ray Reproduction of noise on manipulation of pocket | ||
High impedance (episodic) and noise may saturate the amplifier E/o fracture on X-ray | ||
EMI – Electromagnetic interference; E/o – evidence of.
Only in implantable cardioverter-defibrillator (ICD) devices.