| Literature DB >> 30348126 |
Eva Larsson1, Anna Molnar2, Gerd Holmström2.
Abstract
BACKGROUND: The aim was, first, to collect normative data of the optic nerve head and the peripapillary retinal nerve fibre layer (RNFL) thickness assessed with Cirrus SD-OCT, in healthy children in a population-based study; second, using these data, to examine repeatability, reproducibility and the interocular difference.Entities:
Keywords: Children; Interocular difference; Normal values; Optical coherence tomography (OCT); Repeatability; Reproducibility
Mesh:
Year: 2018 PMID: 30348126 PMCID: PMC6196413 DOI: 10.1186/s12886-018-0940-x
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Mean values (SD) and ranges of optic nerve head and peripapillary retinal nerve fibre layer (RNFL) assessed with Cirrus OCT, with intra- and inter-examiner variation
| First examiner ( | Intra-examiner variation ( | Inter-examiner variation ( | ||||
|---|---|---|---|---|---|---|
| Mean (SD) | Range | CV mean (SD) | ICC (CI) | CV mean (SD) | ICC (CI) | |
| Average RFNL thickness (μm) | 99.2 (8.8) | 75.3–122.5 | 1.6% (1.2) | 0.943 (0.920–0.961) | 1.5% (1.4) | 0.940 (0.914–0.959) |
| Inferior | 130.8 (15.1) | 91.3–168.5 | 2.9% (2.1) | 0.916 (0.882–0.942) | 2.8% (2.4) | 0.889 (0.841–0.923) |
| Superior | 123.2 (14.8) | 66.0–167.0 | 3.0% (2.6) | 0.848 (0.791–0.893) | 2.8% (4.1) | 0.851 (0.790–0.896) |
| Nasal | 74.4 (11.5) | 47.0–102.3 | 3.3% (2.4) | 0.925 (0.895–0.948) | 3.1% (2.6) | 0.925 (0.892–0.948) |
| Temporal | 67.8 (8.2) | 47.3–84.0 | 3.9% (11) | 0.894 (0.853–0.926) | 3.2% (4.8) | 0.767 (0.677–0.835) |
| Disc area (mm2) | 1.89 (0.37) | 1.09–3.02 | 4.2% (5.4) | 0.919 (0.887–0.944) | 2.9% (4.1) | 0.949 (0.926–0.965) |
| Rim area (mm2) | 1.52 (0.26) | 1.10–2.29 | 3.8% (5.0) | 0.902 (0.864–0.932) | 2.8% (3.6) | 0.929 (0.897–0.951) |
| Average C/D ratio | 0.40 (0.15) | 0.07–0.66 | 4.5% (8.4) | 0.975 (0.964–0.983) | 3.2% (4.8) | 0.990 (0.985–0.993) |
| Vertical C/D ratio | 0.39 (0.14) | 0.06–0.63 | 7.5% (12) | 0.942 (0.918–0.960) | 4.6% (6.6) | 0.977 (0.966–0.984) |
| Cup volume (mm3) | 0.10 (0.10) | 0.0–0.42 | 11.0% (23) | 0.987 (0.982–0.991) | 8.1% (12.4) | 0.993 (0.990–0.995) |
C/D ratio cup disc ratio, CI confidence interval, CV coefficient of variance, ICC intraclass correlation, n number of eyes
Fig. 1Bland–Altman plot of repeatability (intra-observer variability) in assessments of average RNFL thickness. The plot shows the differences between examinations 1 and 2
Fig. 2Bland–Altman plot of reproducibility (inter-observer variability) in assessments of average RNFL thickness. The plot shows the differences between examiners 1 and 2
Measures of symmetry between right and left in 53 pair of eyes assessed with Cirrus OCT
| Right eye | Left eye | Correlation RE-LE | Interocular difference RE-LE | ||
|---|---|---|---|---|---|
| Mean (SD) | Mean (SD) | Mean difference | 95% limits of agreement | ||
| Average RFNL thickness (μm) | 98.7 (8.4) | 98.8 (8.9) | 0.881** | − 0.126 | −8.402–8.150 |
| Inferior | 130.1 (15.3) | 129.7 (13.6) | 0.706** | 0.519 | − 21.45–22.49 |
| Superior | 120.1 (14.1) | 125.3 (14.7) | 0.584** | − 5.226 | −31.05–20.6 |
| Nasal | 75.8 (11.5) | 72.9 (11.2) | 0.734** | 2.909 | −13.39–19.21 |
| Temporal | 68.8 (8.7) | 66.5 (7.6) | 0.684** | 2.230 | −11.36–15.82 |
| Disc area (mm2) | 1.85 (0.39) | 1.90 (0.32) | 0.780** | − 0.044 | −0.529–0.441 |
| Rim area (mm2) | 1.49 (0.26) | 1.52 (0.26) | 0.810** | − 0.035 | −0.344–0.274 |
| Average C/D ratio | 0.39 (0.16) | 0.41 (0.14) | 0.989** | − 0.014 | −0.210–0.181 |
| Vertical C/D ratio | 0.38 (0.16) | 0.40 (0.13) | 0.713** | − 0.022 | −0.242–0.197 |
| Cup volume (mm3) | 0.10 (0.10) | 0.09 (0.08) | 0.832** | 0.005 | −0.106–0.117 |
**p < 0.01 RE right eye, LE left eye, SD standard deviation