| Literature DB >> 30338198 |
Carl Pellerin1, Vinod Panchbhavi2, Cory F Janney3.
Abstract
Introduction Orthopaedic devices represent 12% of all medical device recalls. Products are approved through pre-market approval (PMA) or the 510(k) premarket notification process. No previous evaluation was found in the literature evaluating foot and ankle device recalls. The field of foot and ankle subspecialty has seen a rapid growth in innovation related to implants in recent years. Methods The Food and Drug Administration (FDA) Device Recall database was evaluated for all foot and ankle devices from 2007 through 2017 for the manufacturer, process of approval, type of implant, recall class, dates of initiation and termination of the recall, manufacturer determined reason, quantity affected, and distribution within the United States or internationally. Results A total of 161 products from 33 companies were identified with 158 (98.1%) approved through the 510(k) process. The most common reason for device recall was due to the device breaking intraoperatively or postoperatively. The average length of the recall was 487.5 days. Conclusions Device recall is not an uncommon event with the majority of products approved through the less demanding 510(k) process.Entities:
Keywords: ankle; device; fda; foot; recall
Year: 2018 PMID: 30338198 PMCID: PMC6177061 DOI: 10.7759/cureus.3123
Source DB: PubMed Journal: Cureus ISSN: 2168-8184
Manufacturer-determined reasoning for recall of the devices from January 1, 2007 through December 31, 2017.
| Recall reason | Number of products recalled |
| Product breakage | 39 |
| Sterility issue | 37 |
| Labeling design | 30 |
| Device design | 27 |
| Manufacturing defect | 22 |
| Software design | 3 |
| Packaging issue | 2 |
| Employee error | 1 |
| Total products | 161 |
Figure 1Manufacturer-determined reasoning for all products recalled.
Figure 2FDA approval process utilized by the foot and ankle devices recalled during the selected timeframe.
FDA - Food and Drug Administration.