Literature DB >> 24090661

Modular taper junction corrosion and failure: how to approach a recalled total hip arthroplasty implant.

Robert Pivec1, R Michael Meneghini, William J Hozack, Geoffrey H Westrich, Michael A Mont.   

Abstract

Corrosion at the modular neck-stem taper junction has become an increasingly important topic as several reports have identified this couple as a possible source for early failure with findings similar to failed metal-on-metal hip arthroplasties. Recently, two different modular stem systems from a single manufacturer were voluntarily recalled due to concerns of failure of the modular taper junction. We discuss how to approach the diagnosis and management of patients with these particular stem systems. We further reviewed the literature to evaluate whether this is a manufacturer-specific defect or indicative of a broader trend. Recent studies appear to implicate the basic design of the neck-stem taper junction, rather than a single manufacturer, which is at high risk for fretting and corrosion.
© 2013.

Entities:  

Keywords:  THA; corrosion; modular junction corrosion; modular stem

Mesh:

Year:  2013        PMID: 24090661     DOI: 10.1016/j.arth.2013.08.026

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  14 in total

1.  Revision for taper corrosion at the neck-body junction following total hip arthroplasty: pearls and pitfalls.

Authors:  Mitchell C Weiser; Darwin D Chen
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

2.  CORR Insights®: Long-term results and bone remodeling after THA with a short, metaphyseal-fitting anatomic cementless stem.

Authors:  Michael G Zywiel
Journal:  Clin Orthop Relat Res       Date:  2013-11-27       Impact factor: 4.176

3.  Cementless modular neck stems: are they a safe option in primary total hip arthroplasty?

Authors:  José M Pelayo-de-Tomás; José L Rodrigo-Pérez; Carlos D Novoa-Parra; Alejandro Lizaur-Utrilla; María Morales-Suárez-Varela; José Antonio Blas-Dobón
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-10-29

Review 4.  The biological response to orthopaedic implants for joint replacement: Part I: Metals.

Authors:  Emmanuel Gibon; Derek F Amanatullah; Florence Loi; Jukka Pajarinen; Akira Nabeshima; Zhenyu Yao; Moussa Hamadouche; Stuart B Goodman
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2016-06-21       Impact factor: 3.368

5.  Serologic and radiographic outcome of total hip arthroplasty with CoCr modular neck at mid-term follow-up.

Authors:  M Chillemi; G Placella; A Caraffa; G Cerulli; P Antinolfi
Journal:  Musculoskelet Surg       Date:  2016-09-28

6.  Outcome of revision surgery for adverse local tissue reactions in patients with recalled total hip arthroplasty.

Authors:  Christian Klemt; Sakkadech Limmahakhun; Georges Bounajem; Christopher M Melnic; Michael J Harvey; Young-Min Kwon
Journal:  Arch Orthop Trauma Surg       Date:  2021-04-09       Impact factor: 2.928

Review 7.  Trunnionosis in total hip arthroplasty: a review.

Authors:  Jaydev B Mistry; Morad Chughtai; Randa K Elmallah; Aloise Diedrich; Sidney Le; Melbin Thomas; Michael A Mont
Journal:  J Orthop Traumatol       Date:  2016-02-11

8.  Primary hip replacement stem taper fracture due to corrosion in 3 patients.

Authors:  Michael M Morlock; Dennis Bünte; Harmen Ettema; Cees C Verheyen; Åke Hamberg; Jeremy Gilbert
Journal:  Acta Orthop       Date:  2016-01-11       Impact factor: 3.717

Review 9.  Local Cellular Responses to Titanium Dioxide from Orthopedic Implants.

Authors:  Jie J Yao; Eric A Lewallen; William H Trousdale; Wei Xu; Roman Thaler; Christopher G Salib; Nicolas Reina; Matthew P Abdel; David G Lewallen; Andre J van Wijnen
Journal:  Biores Open Access       Date:  2017-07-01

10.  An Analysis of Foot and Ankle Device Recalls by the Food and Drug Administration.

Authors:  Carl Pellerin; Vinod Panchbhavi; Cory F Janney
Journal:  Cureus       Date:  2018-08-09
View more