| Literature DB >> 30309216 |
Sung Joon Kim1, Jeong Won Lee1, Min Kyu Kang2, Jae-Chul Kim2, Jeong Eun Lee2, Shin-Hyung Park2, Mi Young Kim3, Seoung-Jun Lee3, Soo-Ho Moon3, Byoung-Soo Ko1.
Abstract
PURPOSE: A hybrid-dynamic conformal arc therapy (HDCAT) technique consisting of a single half-rotated dynamic conformal arc beam and static field-in-field beams in two directions was designed and evaluated in terms of dosimetric benefits for radiotherapy of lung cancer.Entities:
Keywords: Arc therapy; Lung neoplasms; Radiotherapy planning; Three-dimensional conformal radiotherapy
Year: 2018 PMID: 30309216 PMCID: PMC6226139 DOI: 10.3857/roj.2018.00171
Source DB: PubMed Journal: Radiat Oncol J ISSN: 2234-1900
Fig. 1.Hybrid-dynamic conformal arc radiotherapy plan, consisting of a single dynamic conformal arc (dark green) and static conformal fields (gray).
Comparison of the target dose-volume parameters for HDCAT and 3D-CRT
| HDCAT | 3D-CRT | p-value | |
|---|---|---|---|
| CI | 0.74 ± 0.06 | 0.62 ± 0.06 | <0.001a) |
| HI | 1.10 ± 0.02 | 1.10 ± 0.02 | 0.616[ |
Values are presented as mean ± standard deviation.
HDCAT, hybrid dynamic conformal arc radiotherapy; 3D-CRT, three-dimensional conformal radiotherapy; CI, conformity index; HI, homogeneity index.
Paired t-test.
Fig. 2.Dose distributions of hybrid-dynamic conformal arc radiotherapy plan (left) and three-dimensional conformal radiotherapy plan (right). Internal target volume in red and planning target volume in cyan.
Fig. 3.Average dose-volume histograms of hybrid-dynamic conformal arc radiotherapy (HDCAT, solid line) and three-dimensional conformal radiotherapy (3D-CRT, dashed line) plans.
Comparison of doses to organs at risk
| HDCAT | 3D-CRT | p-value | |
|---|---|---|---|
| Total lung | |||
| Mean dose (Gy) | 12.0 ± 3.6 | 12.2 ± 3.7 | 0.102[ |
| V5 (%) | 50.8 ± 14.7 | 41.6 ± 13.0 | <0.001[ |
| V10 (%) | 32.5 ± 11.9 | 32.3 ± 10.6 | 0.819[ |
| V13 (%) | 26.9 ± 10.2 | 28.5 ± 10.0 | 0.097[ |
| V20 (%) | 21.4 ± 8.2 | 24.5 ± 8.8 | <0.001[ |
| V30 (%) | 14.2 ± 6.1 | 15.1 ± 6.4 | 0.022[ |
| V40 (%) | 8.8 ± 3.9 | 10.3 ± 4.5 | <0.001[ |
| V50 (%) | 5.7 ± 2.7 | 7.1 ± 3.2 | <0.001[ |
| Heart | |||
| V40 (%) | 5.2 ± 3.9 | 7.6 ± 5.5 | <0.001[ |
| V50 (%) | 1.8 ± 1.6 | 3.1 ± 2.8 | 0.001[ |
| Spinal cord | |||
| Dmax (Gy) | 34.8 ± 9.4 | 42.5 ± 7.8 | <0.001[ |
Values are presented as mean ± standard deviation.
HDCAT, hybrid dynamic conformal arc radiotherapy; 3D-CRT, three-dimensional conformal radiotherapy; Dmax, maximum dose.
Paired t-test.
Wilcoxon signed-rank test.