Satoshi Tanabe1, Osamu Umetsu2, Toshikazu Sasage2, Satoru Utsunomiya3, Ryota Kuwabara2, Toshiki Kuribayashi2, Hiromasa Takatou2, Gen Kawaguchi2, Hidefumi Aoyama4.
Abstract
BACKGROUND & AIMS: A new real-time tracking radiotherapy (RTRT) system, the SyncTraX FX4 (Shimadzu, Kyoto, Japan), consisting of four X-ray tubes and four ceiling-mounted flat panel detectors (FPDs) combined with a linear accelerator, was installed at Uonuma Kikan Hospital (Niigata, Japan) for the first time worldwide. In addition to RTRT, the SyncTraX FX4 system enables bony structure-based patient verification. Here we provide the first report of this system's clinical commissioning for intracranial stereotactic radiotherapy (SRT). MATERIALS &Entities:
Keywords: commissioning; image-guided radiotherapy (IGRT); stereotactic radiotherapy
Mesh:
Year: 2018 PMID: 30273444 PMCID: PMC6236847 DOI: 10.1002/acm2.12467
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1(a) An overview of SyncTraX FX4 system with the TrueBeam. (b) Four predetermined combinations of x‐ray tubes and flat panel detectors (presets).
Figure 2Verification image of SyncTraX FX4‐based bone matching using a head phantom.
Figure 3(a) Image evaluation phantom. (b) Image of the image evaluation phantom taken under the following conditions: 80 kV, 250 mA, 10 ms.
Figure 4Image of the image evaluation phantom. (a) Scaling. (b) Spatial resolution. (c) Contrast. (d) Uniformity.
Figure 5Measurement setup to confirm the degree of coordinate coincidence of the isocenter for each coordinate system. The center of a tungsten ball was aligned with the radiation isocenter of the TrueBeam.
Figure 6(a) Head phantom used to evaluate the localization accuracy between CBCT and the SyncTraX FX4. (b) Head phantom with a 5‐mm‐dia. gold marker used for the end‐to‐end test. (c) A treatment planning CT image for the end‐to‐end test. The isocenter in the treatment planning was aligned with the center of the gold marker.
Figure 7Experimental setup for the measurement of air kerma in (a) Tube #1 and (b) Tube #3. DID and SDD represent the detector‐to‐imager distance and source‐to‐detector distance respectively.
Evaluation of each item for image quality for each FPD (flat panel detector)
| Scaling distance (mm) | Spatial resolution | Contrast CNR | Uniformity CV (%) | |
|---|---|---|---|---|
| FPD #1 | 50.09 | Identifiable | 110.9 | 0.23 |
| FPD #2 | 50.23 | Identifiable | 104.9 | 0.10 |
| FPD #3 | 50.23 | Identifiable | 101.9 | 0.16 |
| FPD #4 | 50.22 | Identifiable | 102.4 | 0.11 |
CNR: contrast‐to‐noise ratio; CV: coefficient of variation.
Discrepancy of the coordinates between the image/radiation center of TrueBeam and the center of SyncTraX FX4
| Image center of TrueBeam and image center of SyncTraX FX4 | Radiation center of TrueBeam and image center of SyncTraX FX4 | |||||
|---|---|---|---|---|---|---|
| AP | SI | LR | AP | SI | LR | |
| Preset #1 | −0.11 | 0.03 | −0.06 | 0.40 | 0.32 | 0.04 |
| Preset #2 | 0.05 | 0.08 | −0.10 | 0.39 | 0.26 | 0.01 |
| Preset #3 | 0.00 | 0.00 | −0.10 | 0.38 | 0.33 | −0.05 |
| Preset #4 | −0.03 | 0.07 | −0.04 | 0.37 | 0.22 | 0.06 |
All data are in mm. AP: anterior–posterior; SI: superior–inferior; LR: left–right.
Mean value, standard deviation, and RMS between the CBCT and SyncTraX FX4
| AP mm | SI mm | LR mm | Rotation deg | Pitch deg | Roll deg | |
|---|---|---|---|---|---|---|
| Mean | ||||||
| Preset #1 | 0.32 | −0.25 | 0.050 | −0.25 | −0.17 | 0.074 |
| Preset #2 | 0.29 | −0.16 | 0.12 | 0.033 | −0.18 | 0.090 |
| Preset #3 | 0.29 | −0.19 | 0.076 | −0.11 | −0.14 | 0.072 |
| Preset #4 | 0.30 | −0.21 | 0.11 | −0.16 | −0.15 | 0.063 |
| SD | ||||||
| Preset #1 | 0.067 | 0.14 | 0.13 | 0.11 | 0.10 | 0.060 |
| Preset #2 | 0.080 | 0.12 | 0.10 | 0.059 | 0.047 | 0.053 |
| Preset #3 | 0.084 | 0.13 | 0.11 | 0.066 | 0.064 | 0.058 |
| Preset #4 | 0.083 | 0.12 | 0.12 | 0.076 | 0.083 | 0.071 |
| RMS | ||||||
| Preset #1 | 0.065 | 0.13 | 0.12 | 0.10 | 0.10 | 0.059 |
| Preset #2 | 0.078 | 0.12 | 0.10 | 0.057 | 0.046 | 0.052 |
| Preset #3 | 0.082 | 0.12 | 0.10 | 0.064 | 0.063 | 0.056 |
| Preset #4 | 0.081 | 0.12 | 0.11 | 0.074 | 0.081 | 0.069 |
CBCT: cone‐beam computed tomography; RMS: root‐mean‐square; SD: standard deviation.
Figure 8(a)–(f) Two‐dimensional correlation plots. (g)–(l) Bland‐Altman error analysis between CBCT and preset #3 in the SyncTraX FX4 system.
Figure 9Relationship between the tube current and air kerma in (a) Preset #3 and (b) Preset #4.
Figure 10The images from the end‐to‐end test. The numbers followed by “G” and “C” are the gantry and couch angles respectively.
Median value of the offset between the center of the marker and the image center for each preset in the end‐to‐end test
| Median, mm (range) | |
|---|---|
| Preset #1 | 0.35 (0.14–0.57) |
| Preset #2 | 0.31 (0.27–0.49) |
| Preset #3 | 0.31 (0.14–0.49) |
| Preset #4 | 0.31 (0.14–0.57) |