The accuracy of a 3D printing surgical guide determined by CBCT and model analysis.

Boyoung Ma1, Taeseok Park2, Inkon Chun2, Kwidug Yun1.   

Abstract

PURPOSE: The aim of this clinical study was to assess the accuracy of the implants placed using a universal digital surgical guide.
MATERIALS AND METHODS: Among 17 patients, 28 posterior implants were included in this study. The digital image of the soft tissue acquired from cast scan and hard tissue from CBCT have been superimposed and planned the location, length, diameter of the implant fixture. Then digital surgical guides were created using 3D printer. Each of angle deviations, coronal, apical, depth deviations of planned and actually placed implants were calculated using CBCT scans and casts. To compare implant positioning errors by CBCT scans and plaster casts, data were analyzed with independent samples t-test.
RESULTS: The results of the implant positioning errors calculated by CBCT and casts were as follows. The means for CBCT analyses were: angle deviation: 4.74 ± 2.06°, coronal deviation: 1.37 ± 0.80 mm, and apical deviation: 1.77 ± 0.86 mm. The means for cast analyses were: angle deviation: 2.43 ± 1.13°, coronal deviation: 0.82 ± 0.44 mm, apical deviation: 1.19 ± 0.46 mm, and depth deviation: 0.03 ± 0.65 mm. There were statistically significant differences between the deviations of CBCT scans and cast.
CONCLUSION: The model analysis showed lower deviation value comparing the CBCT analysis. The angle and length deviation value of the universal digital guide stent were accepted clinically.

Entities:  

Keywords:  Computer-Assisted Radiotherapy Planning; Computer-Assisted Surgery; Stents

Year:  2018        PMID: 30140394      PMCID: PMC6104503          DOI: 10.4047/jap.2018.10.4.279

Source DB:  PubMed          Journal:  J Adv Prosthodont        ISSN: 2005-7806            Impact factor:   1.904


INTRODUCTION

The traditional method of placing an implant is to construct a radiological guiding stents and then converting it to a surgical guiding device after Cone-beam computed tomography (CBCT) were taken.1 However, this traditional surgical guiding stent has complicated and inaccurate lab procedure, and difficult in placing the implant fixture as planned.2 The digital surgical guide was introduced to compensate this. Scanned image of intraoral cavity and CBCT image are used to plan the placement of implant considering bone, mucosa, and tooth.3 Using digital surgical guide, drilling and placing implant at a preset position is possible, which makes less error compare to the traditional method, but only when surgical guide is maintained accurate and stable.4 The accuracy of implant placement with a digital surgical guide is evaluated by the deviation in the planned implant and the placed implant.5 A previous study showed coronal deviation of 1.09 mm, apical deviation of 1.28 mm, and axis angle deviation of 3.9°. The deviation may vary in different studies.6 Most previous studies were conducted with the full edentulous ridges. There are little studies about the accuracy test in the partial edentulous ridges. Furthermore, the accuracy studies were conducted with the digital guide stent only for the only one company. Also, many studies were conducted in the laboratory study, there is no clinical study. In previous studies, the accuracy of implant was assessed by overlapping CBCT before and after surgery. Analysis of implant error is not accurate due to resolution and distortion of CBCT, and error in superimposing two CBCT images. In addition, resolution is decreased due to the metal artifact when there are many metal structures.78 The primary purpose of this study is to assess the implant placement error by using CBCT and plaster cast after placing implant in the posterior tooth with universal digital surgical guide and kit.

MATERIALS AND METHODS

This study was approved by the Institutional Review Board, Chonnam National University Dental Hospital (IRB No. CNUDH-2016-007). To calculate the number of subjects required for this study, in vitro experiment9 was performed using a partial edentulous epoxy model (M. Tech, Seoul, Korea). The number of the subjects is 26, which is calculated using G*power 3.1 program (Heinrich Heine University, Düsseldorf, Germany). 28 implants were selected in this study considering 10% failure rate. The following criteria were used to recruit 28 implants placement (Table 1).
Table 1

Patient selection criteria for inclusion and exclusion

Inclusion criteriaExclusion criteria
① Missing premolar or molar for more than 3 months① Acute periodontitis
② Wish to replace the missing tooth with an implant② Heavy smoking
③ Willing to sign for informed consent③ Physical or psychological disorders prohibiting implant treatment
④ Sufficient bone for implant placement④ Previous therapeutic radiation of the head-neck region
⑤ Sufficient attached mucosa present⑤ Younger than 18 years
⑥ Mouth opening ≥ 40 mm
⑦ Good general health
Before placing an implant, patient's preliminary impression was taken and the diagnostic model was fabricated using hard plaster. 3D model scanner (Freedom HD, Degree of Freedom, Seoul, Korea) was used to scan the diagnostic model and the information of patient's intraoral soft tissue surface was saved as Surface Tesselation Language (STL) file. Patient's hard tissue information was obtained by taking CBCT (Alphard-3030, ASAHI Rogentgen, Kyoto, Japan) and saved as Digital imaging and communications in medicine (DICOM) file. After superimposing the STL and DICOM files on the remaining natural teeth using In2guide (Cybermed, Seoul, Korea) software, the surgical guide was fabricated using 3D printer considering the diameter, length, and position of implant (Fig. 1). 4 weeks after placing the implant with digital surgical guide, CBCT was taken to evaluate the implant accuracy (Fig. 2). The term ‘planned implant’ is used to describe the preset position of implant before the surgery, and the ‘inserted implant’ is the actual placement of fixture after the surgery. The coronal deviation (in mm), apical deviation (in mm), and angle deviation (in °) of planned implant and inserted implant were measured to evaluate the accuracy (Fig. 3).
Fig. 1

Pre-surgical procedure. (A) Diagnostic cast, (B) STL data scanned by 3D model scanner, (C) Pre-surgical CBCT image, (D) Planning on position and angulation of the implant, (E) Surgical guide fabricated by stereolithography.

Fig. 2

Surgical procedures of flapless implant fixture installation by using surgical guide. (A) Surgical template applied over the edentulous area and adjacent teeth, (B) Removal of soft tissue with punch, (C) Drilling through the implant template with a drill, (D) Postoperative view of installed implant.

Fig. 3

Overlapping observation between planned and inserted implants for measurement of deviations. (A) Cross-sectional view of the overlapping, (B) Panoramic view of the overlapping.

Eight weeks after the implant placement, polyvinylsiloxane (Honigum, DMG, Hamburg, Germany) was used to take impression to make hard plaster model and be scanned; this was called inserted model (Fig. 4). Lab analog inside the plaster was removed after scanning. A lab cylinder screw, a lab cylinder body, and a lab analog were connected to the metal sleeve of the surgical guide used for implant placement, and the lab analog was fixed with a hard plaster to regenerate the planned model, the model of planned position of implant before the surgery. The STL file obtained by scanning pre-operative and postoperative models was superimposed and analyzed with a 3D analysis program (Geomagic control X, 3D Systems, Morrisville, NC, USA). Depth deviation is obtained by subtracting the z-axis coordinate of the post-operative model from the pre-operative model. If the value is positive, the implant fixture is located more apically and if the value is negative, the implant is located more on coronal (Fig. 5).
Fig. 4

Inserted model. (A) The cast of inserted model, (B) STL data of inserted model scanned by 3D model scanner, (C) Component used for fabrication of planned model, (D) The cast of planned model, (E) STL data of planned model scanned by 3D model scanner.

Fig. 5

(A) Illustrations of the deviations between planned and inserted implant on the method with cast, (B) The accuracy analysis between planned and inserted implant by 3D analysis program.

All statistical analysis is performed by SPSS Ver 23.0 (SPSS Inc., Chicago, IL, USA). Angle deviation, coronal deviation, and apical deviation measured by CBCT and cast model are statistically analyzed by Independent t-test after performing normality test using Kolmogorov-Smirnov test. All results were tested for significance at the level of P < .05. Pearson correlation analysis on angle, coronal and apical deviation was also used in searching for the correlativity of the two methods.

RESULTS

The accuracy measured with CBCT and cast model is as follows (Table 2). The means for CBCT analyses were: angle deviation: 4.74 ± 2.06°, coronal deviation: 1.37 ± 0.80 mm, and apical deviation: 1.77 ± 0.86 mm. The means for cast analyses were: angle deviation: 2.43 ± 1.13°, coronal deviation: 0.82 ± 0.44 mm, apical deviation: 1.19 ± 0.46 mm, and depth deviation: 0.03 ± 0.65 mm. The angular, coronal, and apical deviations were significantly smaller in measurement using cast model than those measured using the CBCT method (P = .01). Also, apical deviation was bigger than coronal deviation (Fig. 6). Angular, coronal and apical deviation in CBCT and cast showed positive correlation and there were significant difference among them (P = .001, .024) (Fig. 7).
Table 2

The angle, coronal, apical and depth deviation between CBCT and cast

MethodMean ± SDMaxMinP value
Angle deviation (°)CBCT4.74 ± 2.06a8.860.000.001
Cast2.43 ± 1.13b5.701.15
Coronal deviation (mm)CBCT1.37 ± 0.80a3.760.181.000
Cast0.82 ± 0.44b1.850.13
Apical deviation (mm)CBCT1.77 ± 0.86a3.760.450.003
Cast1.19 ± 0.46b2.510.37
Depth deviation (mm)Cast−0.03 ± 0.651.58−2.12

Different superscript letters indicate significant differences.

Fig. 6

The angle and length deviation between CBCT and cast. Results show mean ± SD angle deviation. * indicates statistical differences (P < .05).

Fig. 7

Scatter plot for the Pearson correlation between CBCT analysis and cast analysis. (A) The angle deviation (P = .001, R = 0.635), (B) The coronal deviation (P = .054, R = 0.368), (C) The apical deviation (P = .024, R = 0.426).

DISCUSSION

While digital surgical guide makes it possible to drill and place an implant in a preset position, the procedure has to include accurate position and accurate analysis before the surgery. Effort has been made to improve the accuracy of digital surgical guide as well as maintenance of surgical guide and errors occurring during the manufacturing process. During implant procedure, 1 angle deviation makes 0.34 mm length deviation in the 10-mm fixture apical area. 5° angle deviation makes 1.7 mm length deviation. If the space between implant and tooth root were set to 1.5 mm during implant planning, 5° angle error will impair the tooth root. Thus the angle deviation should be no more than 3° to implant installed safely without the tooth damaged.10 If the important anatomical structure such as inferior alveolar nerve is close by, acceptable surgical guide's maximum angle deviation is less than 3° and maximum vertical error is less than −1.5 mm.10 Less loosening of implant and passive fit is possible when the angle between the hex of fixture and hexagonal freedom of abutment is less than 5° and the distance is 150 µm.1112 This study shows that angular, coronal, and apical deviation are accurate enough to avoid the damage of major anatomical structure during the procedure, but less accurate to connect directly to pre-manufactured hexagonal implant prosthesis. Therefore, it is necessary to manufacture prosthesis by taking an impression after the implant placement, or to use non-hexagonl implant fixture.1 In this study, deviation measured by CBCT is similar to that of other studies but angle deviation is somewhat higher. The reason for this is that the previous studies used surgical kit and implant fixture of the subsidiary company that makes surgical guide, while universal surgical guide kit is used in this study. Also, compared the previous studies, the more rearmost molars are included in this study. Reference marker was not used when taking CBCT, so higher error occurred during overlapping preoperative and postoperative CBCT. The angle, coronal, and apical deviation were statistically significantly smaller than those of CBCT in this study (P < .05). Not only there is no error occurred while superimposing pre-operative and postoperative CBCTs, but also can additional radiation be decreased by using cast model to analyze accuracy. However, errors can occur when taking impression or making plaster cast model. There was more error in overlapping CBCTs because of the absence of marker in CBCT analysis. Error can be decreased by using only one cast model to reproduce preoperative and postoperative cast model. Further studies in comparing these two methods are required by adding marker to improve accuracy of superimposition. In cast model analysis, depth deviation can be measured by setting axis of the implant as z-axis and obtaining difference in the z-axis. Average of depth deviation is −0.03 ± 0.65 mm, which is more on the coronal side than planned position. This result is similar to that of the previous study, especially when those who have less experience with digital guided surgery tend to have less reliability in the accuracy of the surgical guide.13 This study shows that the deviation obtained by the plaster cast is significantly smaller, which can be useful in evaluation of implant placement accuracy. In addition, the angular deviation may become larger toward the farthest tooth in implant placement.

CONCLUSION

The present study results showed significantly smaller deviation values using cast model analysis than those measured using the CBCT superimposition method. The angle and length deviation value of the universal digital guide stent were clinically acceptable.
  17 in total

1.  Passive fit and accuracy of three dental implant impression techniques.

Authors:  Firas A Al Quran; Bashar A Rashdan; AbdelRahman A Abu Zomar; Saul Weiner
Journal:  Quintessence Int       Date:  2012-02       Impact factor: 1.677

2.  Virtually planned and template-guided implant surgery: an experimental model matching approach.

Authors:  Ai Komiyama; Andreas Pettersson; Margareta Hultin; Karin Näsström; Björn Klinge
Journal:  Clin Oral Implants Res       Date:  2010-09-27       Impact factor: 5.977

3.  Accuracy of implant placement using precision surgical guides with varying occlusogingival heights: an in vitro study.

Authors:  Chanseop Park; Ariel J Raigrodski; Jacob Rosen; Charles Spiekerman; Robert M London
Journal:  J Prosthet Dent       Date:  2009-06       Impact factor: 3.426

4.  Accuracy of computer-aided oral implant surgery: a clinical and radiographic study.

Authors:  Francesco Valente; Guido Schiroli; Andrea Sbrenna
Journal:  Int J Oral Maxillofac Implants       Date:  2009 Mar-Apr       Impact factor: 2.804

Review 5.  Accuracy of computer-aided implant placement.

Authors:  N Van Assche; M Vercruyssen; W Coucke; W Teughels; R Jacobs; M Quirynen
Journal:  Clin Oral Implants Res       Date:  2012-10       Impact factor: 5.977

Review 6.  Guided surgery with tooth-supported templates for single missing teeth: A critical review.

Authors:  Alessandro Pozzi; Giovanni Polizzi; Peter K Moy
Journal:  Eur J Oral Implantol       Date:  2016       Impact factor: 3.123

7.  Computed tomography-based evaluation of template (NobelGuide™)-guided implant positions: a prospective radiological study.

Authors:  Christoph Vasak; Georg Watzak; André Gahleitner; Georg Strbac; Michael Schemper; Werner Zechner
Journal:  Clin Oral Implants Res       Date:  2011-01-18       Impact factor: 5.977

8.  Factors influencing transfer accuracy of cone beam CT-derived template-based implant placement.

Authors:  Alexandra Behneke; Matthias Burwinkel; Nikolaus Behneke
Journal:  Clin Oral Implants Res       Date:  2011-10-24       Impact factor: 5.977

9.  Reliability of implant placement after virtual planning of implant positions using cone beam CT data and surgical (guide) templates.

Authors:  Hans-Joachim Nickenig; Stephan Eitner
Journal:  J Craniomaxillofac Surg       Date:  2007-06-18       Impact factor: 2.078

10.  Clinical accuracy of 3 different types of computed tomography-derived stereolithographic surgical guides in implant placement.

Authors:  Oguz Ozan; Ilser Turkyilmaz; Ahmet Ersan Ersoy; Edwin A McGlumphy; Stephen F Rosenstiel
Journal:  J Oral Maxillofac Surg       Date:  2009-02       Impact factor: 1.895

View more
  7 in total

1.  Accuracy of computer-assisted surgery in immediate implant placement: An experimental study.

Authors:  Nicole Báez-Marrero; José Luis Rafel; Yalil Augusto Rodríguez-Cárdenas; Aron Aliaga-Del Castillo; Heraldo Luis Dias-Da Silveira; Luis Ernesto Arriola-Guillén
Journal:  J Indian Soc Periodontol       Date:  2021-09-27

2.  Patient-specific desktop 3D-printed guides for pelvic tumour resection surgery: a precision study on cadavers.

Authors:  Mónica García-Sevilla; Lydia Mediavilla-Santos; María Teresa Ruiz-Alba; Rubén Pérez-Mañanes; José Antonio Calvo-Haro; Javier Pascau
Journal:  Int J Comput Assist Radiol Surg       Date:  2021-02-22       Impact factor: 2.924

3.  Patient Eligibility for Standardized Treatment of the Edentulous Mandible: A Retrospective CBCT-Based Assessment of Mandibular Morphology.

Authors:  Walid Aouini; France Lambert; Luc Vrielinck; Bart Vandenberghe
Journal:  J Clin Med       Date:  2019-05-07       Impact factor: 4.241

4.  Accuracy of Guided Implant Surgery in the Edentulous Jaw Using Desktop 3D-Printed Mucosal Supported Guides.

Authors:  Rani D'haese; Tom Vrombaut; Geert Hommez; Hugo De Bruyn; Stefan Vandeweghe
Journal:  J Clin Med       Date:  2021-01-20       Impact factor: 4.241

Review 5.  The Modern and Digital Transformation of Oral Health Care: A Mini Review.

Authors:  Muhammad Syafiq Alauddin; Ahmad Syukran Baharuddin; Mohd Ifwat Mohd Ghazali
Journal:  Healthcare (Basel)       Date:  2021-01-25

6.  Digitally prefabricated versus conventionally fabricated implant-supported full-arch provisional prosthesis: a retrospective cohort study.

Authors:  Chaoqun Chen; Haiyan Lai; Huiyong Zhu; Xinhua Gu
Journal:  BMC Oral Health       Date:  2022-08-09       Impact factor: 3.747

Review 7.  Cone Beam Computed Tomography in Oral and Maxillofacial Surgery: An Evidence-Based Review.

Authors:  Robert Weiss; Andrew Read-Fuller
Journal:  Dent J (Basel)       Date:  2019-05-02
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.