Evaluation of the Fourth-Generation FloTrac/Vigileo System in Comparison With the Intermittent Bolus Thermodilution Method in Patients Undergoing Cardiac Surgery.

Yusuke Kusaka1, Fumihiro Ohchi2, Toshiaki Minami2.   

Abstract

OBJECTIVES: The aim of this study was to evaluate the accuracy, precision, and trending ability of the fourth-generation FloTrac/Vigileo system (version 4.00; Edwards Lifesciences, Irvine, CA) by comparing cardiac output derived from FloTrac/Vigileo system (COAP) with that measured by a pulmonary artery catheter (COTD), and to determine the effects of hemodynamic variables on the bias between COTD and COAP.
DESIGN: A prospective study.
SETTING: University hospital. PARTICIPANTS: Thirty patients undergoing elective cardiac surgery using cardiopulmonary bypass.
INTERVENTIONS: Including hemodynamic variables, COTD and COAP were measured simultaneously at the following 10 time points: after the induction of anesthesia, at the start of operation, after sternotomy, before and after the administration of heparin, before and after the administration of protamine, at the start of sternal closure, at the end of operation, and on arrival to intensive care unit.
MEASUREMENTS AND MAIN RESULTS: In total, 280 pairs of datasets were obtained. Bland-Altman analysis showed a bias of -0.41 L/min, a precision of 0.72 L/min, and limits of agreement of -1.85 and 1.03 L/min, with a percentage error of 37.1%. The concordance rate determined by 4-quadrant plot analysis and the polar concordance rate were 76% and 79%, respectively. The linear mixed-effect model revealed that the bias was influenced strongly by the difference in pulse pressure between the radial and femoral artery (p < 0.001), and the systemic vascular resistance index (p < 0.001).
CONCLUSION: The fourth-generation FloTrac/Vigileo system still lacks accuracy and trending ability in cardiac surgery, and the discrepancy in cardiac output measurement depends on the peripheral vascular tone. Further improvement of this system is needed.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  FloTrac/Vigileo system; cardiac output; pulmonary artery catheter; pulse pressure; systemic vascular resistance index

Mesh:

Year:  2018        PMID: 30077561     DOI: 10.1053/j.jvca.2018.06.017

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  9 in total

1.  Using extra systoles and the micro-fluid challenge to predict fluid responsiveness during cardiac surgery.

Authors:  Simon T Vistisen; Jonas M Berg; Mattheus F Boekel; Marco Modestini; Remco Bergman; Jayant S Jainandunsing; Massimo A Mariani; Thomas W L Scheeren
Journal:  J Clin Monit Comput       Date:  2018-11-09       Impact factor: 2.502

Review 2.  Accuracy of non-invasive and minimally invasive hemodynamic monitoring: where do we stand?

Authors:  Issa Pour-Ghaz; Theodore Manolukas; Nathalie Foray; Joel Raja; Aranyak Rawal; Uzoma N Ibebuogu; Rami N Khouzam
Journal:  Ann Transl Med       Date:  2019-09

3.  Bioreactance and fourth-generation pulse contour methods in monitoring cardiac index during off-pump coronary artery bypass surgery.

Authors:  Laura Anneli Ylikauma; Pasi Petteri Ohtonen; Tiina Maria Erkinaro; Merja Annika Vakkala; Janne Henrik Liisanantti; Jari Uolevi Satta; Tatu Sakari Juvonen; Timo Ilari Kaakinen
Journal:  J Clin Monit Comput       Date:  2021-05-26       Impact factor: 1.977

4.  Hemodynamic Monitoring Options in COVID-19.

Authors:  Olga Rozental; Richard Thalappillil; Robert S White; Shreyajit R Kumar; Christopher W Tam
Journal:  J Cardiothorac Vasc Anesth       Date:  2020-05-06       Impact factor: 2.628

5.  Disagreement in cardiac output measurements between fourth-generation FloTrac and critical care ultrasonography in patients with circulatory shock: a prospective observational study.

Authors:  Thomas Kaufmann; Ramon P Clement; Bart Hiemstra; Jaap Jan Vos; Thomas W L Scheeren; Frederik Keus; Iwan C C van der Horst
Journal:  J Intensive Care       Date:  2019-04-11

6.  The value of a superior vena cava collapsibility index measured with a miniaturized transoesophageal monoplane continuous echocardiography probe to predict fluid responsiveness compared to stroke volume variations in open major vascular surgery: a prospective cohort study.

Authors:  Şerban-Ion Bubenek-Turconi; Adham Hendy; Sorin Băilă; Anca Drăgan; Ovidiu Chioncel; Liana Văleanu; Bianca Moroșanu; Vlad-Anton Iliescu
Journal:  J Clin Monit Comput       Date:  2019-07-05       Impact factor: 2.502

7.  Efficacy of early goal-directed therapy using FloTrac/EV1000 to improve postoperative outcomes in patients undergoing off-pump coronary artery bypass surgery: a randomized controlled trial.

Authors:  Sirirat Tribuddharat; Thepakorn Sathitkarnmanee; Kriangsak Ngamsaengsirisup; Sanpicha Sornpirom
Journal:  J Cardiothorac Surg       Date:  2022-08-21       Impact factor: 1.522

8.  Comparison of accuracy of two uncalibrated pulse contour cardiac output monitors in off-pump coronary artery bypass surgery patients using pulmonary artery catheter-thermodilution as a reference.

Authors:  Ramakrishna Mukkamala; Benjamin A Kohl; Aman Mahajan
Journal:  BMC Anesthesiol       Date:  2021-07-10       Impact factor: 2.217

9.  Efficacy of Intraoperative Hemodynamic Optimization Using FloTrac/EV1000 Platform for Early Goal-Directed Therapy to Improve Postoperative Outcomes in Patients Undergoing Coronary Artery Bypass Graft with Cardiopulmonary Bypass: A Randomized Controlled Trial.

Authors:  Sirirat Tribuddharat; Thepakorn Sathitkarnmanee; Kriangsak Ngamsangsirisup; Krisana Nongnuang
Journal:  Med Devices (Auckl)       Date:  2021-06-28
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.