Thomas Kaufmann1, Ramon P Clement1, Bart Hiemstra1,2, Jaap Jan Vos1, Thomas W L Scheeren1, Frederik Keus2, Iwan C C van der Horst2.
Abstract
BACKGROUND: Cardiac output measurements may inform diagnosis and provide guidance of therapeutic interventions in patients with hemodynamic instability. The FloTrac™ algorithm uses uncalibrated arterial pressure waveform analysis to estimate cardiac output. Recently, a new version of the algorithm has been developed. The aim was to assess the agreement between FloTrac™ and routinely performed cardiac output measurements obtained by critical care ultrasonography in patients with circulatory shock.Entities:
Keywords: Cardiac output; Critical care ultrasonography; Critically ill; Intensive care; Monitoring; Pulse contour analysis; Shock
Year: 2019 PMID: 31011425 PMCID: PMC6460822 DOI: 10.1186/s40560-019-0373-5
Source DB: PubMed Journal: J Intensive Care ISSN: 2052-0492
Fig. 1Patient flowchart. Abbreviations: CCUS, critical care ultrasonography
Patient characteristics
| Patient characteristics ( | ||
| Age (years) | 65 (9) | |
| Male gender, | 14 (82%) | |
| Body mass index (kg m−2) | 25.7 (4.7) | |
| Clinical characteristics on study inclusion | ||
| Heart rate (bpm) | 95 (26) | |
| Systolic arterial pressure (mmHg) | 102 (15) | |
| Diastolic arterial pressure (mmHg) | 55 (6) | |
| Mean arterial pressure (mmHg) | 69 (7) | |
| Norepinephrine therapy, | 16 (94%) | |
| Norepinephrine dose (μg kg−1 min−1) | 0.13 (0.05, 0.38) | |
| Mechanical ventilation, | 12 (71%) | |
| Positive end-expiratory pressure (cm H2O) | 8 (6, 9) | |
| AVPU score, | Alert | 2 (12%) |
| Verbal | 4 (24%) | |
| Passive | 1 (6%) | |
| Unresponsive | 10 (58%) | |
| Mottling score, | None | 3 (19%) |
| Modest | 2 (13%) | |
| Mild | 7 (43%) | |
| Moderate | 3 (19%) | |
| Severe | 1 (6%) | |
| Urine output (mL kg−1 h−1) | 0.49 (0.26, 0.66) | |
| Lactate (mmol L−1) | 1.7 (1.4, 3.4) | |
| APACHE IV—score (points) | 92 (32) | |
| SAPS II— (points) | 56 (17) | |
Data are presented as the mean and standard deviation, median and interquartile ranges or as absolute frequencies with percentages as appropriate
Abbreviations: AVPU alert, verbal, pain, unresponsive; APACHE Acute Physiology And Chronic Health Evaluation; SAPS Simplified Acute Physiology Score
Fig. 2Scatter plot of cardiac output measured by FloTrac™ and CCUS. Abbreviations: COAP, cardiac output measured using fourth-generation FloTrac™ algorithm; COCCUS, cardiac output measured by critical care ultrasonography
Fig. 3Bland-Altman plot for repeated measurements showing the comparison between COAP and COCCUS. The mean bias between COAP and COCCUS and the upper and lower limits of agreement (LOA) are presented. Abbreviations: COAP, cardiac output measured using fourth-generation FloTrac™ algorithm; COCCUS, cardiac output measured by critical care ultrasonography
Fig. 4Four-quadrant plot showing the trend of COAP versus COCCUS. Exclusion zone of 0.5 L min−1 (white area). Abbreviations: COAP, cardiac output measured using fourth-generation FloTrac™ algorithm; COCCUS, cardiac output measured by critical care ultrasonography