| Literature DB >> 30008972 |
Edsel Ing1, Angela Zhang1, Evan Michaelov2, Wendy Wang2.
Abstract
BACKGROUND: Dynamic Contour Tonometry (DCT) is touted to be the most accurate tonometer for Intraocular Pressure (IOP) measurement. Non-Contact "air puff" Tonometry (NCT) may be the most commonly used tonometer for screening of IOP. Elevated IOP is important to exclude in patients presenting with headache or vision loss.Entities:
Keywords: Agreement; Bland-Altman plot; Dynamic contour tonometry; Giant cell arteritis; Headache; Intraocular pressure; Non-contact tonometry
Year: 2018 PMID: 30008972 PMCID: PMC6018127 DOI: 10.2174/1874364101812010104
Source DB: PubMed Journal: Open Ophthalmol J ISSN: 1874-3641
Studies comparing Dynamic Contour Tonometry and Non-contact Tonometry.
| Author | Year | n | NCT | Mean Δ IOP | LoA | Mean Age (years) | Population |
|---|---|---|---|---|---|---|---|
| Burvenich | 2005 | 294 | Nidek NT 2000 | 2.7 | ? | ? | Healthy |
| Erdurmus | 2009 | 104 | ? | 0.8 | -0.8 to 3.0 | 61.4 | Glaucoma or Ocular hypertension |
| Ito | 2012 | 74 | Topcon CT-70 | 3.2 | -1.9 to 8.3 | 70.1 | Non- glaucoma |
| Briceno | 2016 | 45 | Visionix VX 120 | 1.6 | -2.9 to 6.1 | 47.2 | Healthy |
| Present study (OD) | 2017 | 106 | Nidek Tonoref II | 3.9 | -2.5 to 10.4 | 72.2 | Headache or |
Legend
N: number of patients.
NCT: non-contact tonometer.
DCT: dynamic contour tonometer.
Mean Δ IOP: mean difference in intraocular pressure from NCT versus DCT.
mm Hg = millimetres of mercury.
LoA: Limits of Agreement between DCT and NCT on Bland-Altman analysis.
OD: right eye.
Mean Intraocular Pressure and Ocular Pulse Amplitude by Tonometry Type.
| - | Right Eye | Left Eye |
|---|---|---|
| Mean NCT (mm Hg) | 15.2 +/- 5.4 | 14.7 +/- 3.6 |
| Mean DCT (mm Hg) | 19.1 +/- 5.9 | 18.2 +/- 4.1 |
| ΔT (mm Hg) | 3.9 +/- 3.2 | 3.5 +/- 3.2 |
| Paired t-test NCT versus DCT | p <.001 | p <0.001 |
| DCT OPA (mm Hg) | 3.0 +/- 1.3 | 3.0+/- 1.3 |
Legend:
NCT = non-contact tonometer.
DCT = dynamic contour tonometer.
ΔT = mean difference in intraocular pressure (DCT – NCT) derived from individual measurements.
OPA = ocular pulse amplitude obtained from DCT.