IOP measured by dynamic contour tonometry correlates with IOP measured by Goldmann applanation tonometry and non-contact tonometry in Japanese individuals.

Ken Ito1, Akihiko Tawara, Toshiaki Kubota, Yukinori Harada.   

Abstract

PURPOSE: The purpose of this study was threefold. We sought to compare the intraocular pressure (IOP) measured by dynamic contour tonometry (DCT) with that measured by Goldmann applanation tonometry (GAT) and noncontact tonometry (NCT). We also examined the influence of central corneal thickness (CCT) and corneal curvature radius (CCR) on the IOP measurements. Last, we investigated the factors that could affect the ocular pulse amplitude (OPA) measurements.
METHODS: Seventy-four patients with no history of intraocular surgery were enrolled in this study. We measured IOP by DCT, GAT, and NCT, and the CCT, CCR, and axial length (AL) in the right eye of each patient. We also measured OPA by DCT. We subsequently analyzed the correlation of IOP measurements between GAT and DCT and between NCT and DCT. We also examined the influence of CCT, CCR, and AL on IOP readings by the 3 tonometers. In addition, we investigated the factors that could affect the OPA measurements.
RESULTS: The mean IOP measured by DCT was 2.8 mm Hg higher than that by GAT and 3.2 mm Hg higher than that by NCT. This difference was greater with thinner CCT in the lower IOP group than in the higher IOP group. IOP measurements by both GAT and NCT significantly correlated with CCT; however, IOP measurement by DCT did not correlate with CCT. No significant correlations were shown between the IOP measured by each of the 3 tonometers and either CCR or AL. OPA measurements positively correlated with age, IOP measurement by DCT, and pulse pressure.
CONCLUSIONS: IOP measured by DCT correlates with IOP measured by GAT or NCT with a roughly 3.0 mm Hg higher value, and these differences were greater in the patients with a thinner CCT. IOP measurements by both GAT and NCT significantly correlated with CCT; however, IOP measurement by DCT did not correlate with CCT. Our findings also indicate that OPA measured using DCT shows a positive correlation with patient age, IOP measurement by DCT, and pulse pressure.

Entities:  

Mesh:

Year:  2012        PMID: 21173706     DOI: 10.1097/IJG.0b013e31820275b4

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  8 in total

1.  Diurnal variations of central corneal thickness and intraocular pressure in dogs from 8:00 am to 8:00 pm.

Authors:  Eva Martín-Suárez; Carolina Molleda; Rodrigo Tardón; Alba Galán; José Gallardo; José Molleda
Journal:  Can Vet J       Date:  2014-04       Impact factor: 1.008

2.  Impact of corneal parameters on intraocular pressure measurements in different tonometry methods.

Authors:  Aleksandra Zakrzewska; Marta P Wiącek; Anna Machalińska
Journal:  Int J Ophthalmol       Date:  2019-12-18       Impact factor: 1.779

3.  Effect of 0.01% atropine eyedrops on intraocular pressure in schoolchildren: a randomized clinical trial.

Authors:  Javaria Bukhari; Shi-Fei Wei; Shi-Ming Li; Wen-Zai An; Jia-Ling Du; Xin-Tong Liang; Jia-He Gan; Jia-Xin Tian; Wei-Ling Bai; Zhi-Ning Cai; Lei Yin; Ning-Li Wang
Journal:  Int J Ophthalmol       Date:  2022-09-18       Impact factor: 1.645

4.  Analysis of differences in intraocular pressure evaluation performed with contact and non-contact devices.

Authors:  Michele Lanza; Michele Rinaldi; Ugo Antonello Gironi Carnevale; Silvio di Staso; Mario Bifani Sconocchia; Ciro Costagliola
Journal:  BMC Ophthalmol       Date:  2018-09-03       Impact factor: 2.209

5.  Comparison of Non-contact Tonometry and Goldmann Applanation Tonometry Measurements in Non-pathologic High Myopia.

Authors:  Peiyuan Wang; Yunhe Song; Fengbin Lin; Zhenyu Wang; Xinbo Gao; Weijing Cheng; Meiling Chen; Yuying Peng; Yuhong Liu; Xiulan Zhang; Shida Chen
Journal:  Front Med (Lausanne)       Date:  2022-03-03

6.  Correlation between ocular perfusion pressure and ocular pulse amplitude in glaucoma, ocular hypertension, and normal eyes.

Authors:  Bruno P Figueiredo; Sebastião Cronemberger; Fabio N Kanadani
Journal:  Clin Ophthalmol       Date:  2013-08-09

7.  Comparison of Dynamic Contour Tonometry and Non-contact Tonometry in Older Patients Presenting with Headache or Vision Loss.

Authors:  Edsel Ing; Angela Zhang; Evan Michaelov; Wendy Wang
Journal:  Open Ophthalmol J       Date:  2018-06-22

8.  Intraocular pressure according to different types of tonometry (non-contact and Goldmann applanation) in patients with different degrees of bilateral tearing.

Authors:  Bo Ram Seol; Tae Gu Kang; Bonhyeok Gu
Journal:  PLoS One       Date:  2019-09-16       Impact factor: 3.240

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.