Stented valve dynamic behavior induced by polyester fiber leaflet material in transcatheter aortic valve devices.

Hoda Hatoum1, Frederick Heim2, Lakshmi Prasad Dasi3.   

Abstract

OBJECTIVE: This study aims at assessing the global dynamic behavior, elastic deformability, closing energy and turbulence of rigid versus deformable stented (RS vs DS) valve systems with deformable and rigid textile materials (DT vs RT) through studying the stent-valve interaction compared to a bioprosthetic material in transcatheter aortic valves (TAV).
METHODS: Three 19 mm stented textile TAV designs (RS-DT, RS-RT and DS-RT) with different stent and leaflet properties were tested and compared with a control bioprosthetic TAV (RS-DB) in a left heart simulator flow loop under physiological pressure and flow. Particle Image Velocimetry and high speed imaging were performed. Pressure gradients (PG), leakage fractions (LF), Pinwheeling indices (PI), closing energy (E) and Reynolds shear stresses (RSS) were calculated.
RESULTS: (a) PGs and LFs were 11.86 ± 0.51 mmHg, 11.70 ± 0.34%; 8.84 ± 0.40 mmHg, 29.80 ± 0.76%; 11.59 ± 0.12 mmHg, 14.23 ± 1.64%; and 7.05 ± 0.09 mmHg, 12.08 ± 0.45% % for RS-DB, RS-DT, RS-RT and DS-RT respectively. (b) PIs were 15.79 ± 2.34%, 4.36 ± 0.84%, 2.47 ± 0.51% and 2.03 ± 0.33% for RS-DB, RS-DT, RS-RT and DS-RT respectively. (c) E is lowest for DS-RT (0.0010 ± 0.0002 J) followed by RS-RT (0.0017 ± 0.0002 J), RS-DB (0.0023 ± 0.0004 J) and highest with RS-DT (0.0036 ± 0.0007 J). (d) At peak systole lowest RSS was obtained with RS-DT (87.82 ± 0.58 Pa) and highest with DS-RT (122.98 ± 1.87 Pa).
CONCLUSION: PGs, LFs, PIs and E were improved with DS-RT compared to other textile TAVs and RS-DB. Despite achieving more RSS than the rest of TAVs, DS-RT still falls within the same range of RSS produced by the other 2 valves and control exceeding the threshold for platelet activation.
Copyright © 2018 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Bioprosthetic valves; Deformability; Textile valves; Transcatheter aortic valves; Turbulence

Mesh:

Substances:

Year:  2018        PMID: 29986298      PMCID: PMC6089245          DOI: 10.1016/j.jmbbm.2018.06.038

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  28 in total

1.  Turbulence characteristics downstream of bileaflet aortic valve prostheses.

Authors:  J S Liu; P C Lu; S H Chu
Journal:  J Biomech Eng       Date:  2000-04       Impact factor: 2.097

2.  Estimation of shear stress-related blood damage in heart valve prostheses--in vitro comparison of 25 aortic valves.

Authors:  M Giersiepen; L J Wurzinger; R Opitz; H Reul
Journal:  Int J Artif Organs       Date:  1990-05       Impact factor: 1.595

3.  Fiber heart valve prosthesis: influence of the fabric construction parameters on the valve fatigue performances.

Authors:  Antoine Vaesken; Frederic Heim; Nabil Chakfe
Journal:  J Mech Behav Biomed Mater       Date:  2014-08-27

4.  Implantation Depth and Rotational Orientation Effect on Valve-in-Valve Hemodynamics and Sinus Flow.

Authors:  Hoda Hatoum; Jennifer Dollery; Scott M Lilly; Juan A Crestanello; Lakshmi Prasad Dasi
Journal:  Ann Thorac Surg       Date:  2018-05-26       Impact factor: 4.330

Review 5.  A relationship between Reynolds stresses and viscous dissipation: implications to red cell damage.

Authors:  S A Jones
Journal:  Ann Biomed Eng       Date:  1995 Jan-Feb       Impact factor: 3.934

6.  Shear-induced activation of platelets.

Authors:  J M Ramstack; L Zuckerman; L F Mockros
Journal:  J Biomech       Date:  1979       Impact factor: 2.712

Review 7.  Flow in prosthetic heart valves: state-of-the-art and future directions.

Authors:  Ajit P Yoganathan; K B Chandran; Fotis Sotiropoulos
Journal:  Ann Biomed Eng       Date:  2005-12       Impact factor: 3.934

8.  Effect of severe bioprosthetic valve tissue ingrowth and inflow calcification on valve-in-valve performance.

Authors:  Hoda Hatoum; Jennifer Dollery; Scott M Lilly; Juan A Crestanello; Lakshmi Prasad Dasi
Journal:  J Biomech       Date:  2018-05-04       Impact factor: 2.712

9.  Simulation of long-term fatigue damage in bioprosthetic heart valves: effects of leaflet and stent elastic properties.

Authors:  Caitlin Martin; Wei Sun
Journal:  Biomech Model Mechanobiol       Date:  2013-10-04

10.  Theory to predict shear stress on cells in turbulent blood flow.

Authors:  Khandakar Niaz Morshed; David Bark; Marcio Forleo; Lakshmi Prasad Dasi
Journal:  PLoS One       Date:  2014-08-29       Impact factor: 3.240

View more
  7 in total

1.  In vitro hemodynamic assessment of a novel polymeric transcatheter aortic valve.

Authors:  Megan Heitkemper; Hoda Hatoum; Lakshmi Prasad Dasi
Journal:  J Mech Behav Biomed Mater       Date:  2019-06-19

2.  Spatiotemporal Complexity of the Aortic Sinus Vortex as a Function of Leaflet Calcification.

Authors:  Hoda Hatoum; Lakshmi Prasad Dasi
Journal:  Ann Biomed Eng       Date:  2019-02-01       Impact factor: 3.934

3.  Impact of BASILICA on Sinus and Neo-Sinus Hemodynamics after Valve-in-Valve with and without Coronary Flow.

Authors:  Hoda Hatoum; Pablo Maureira; Scott Lilly; Lakshmi Prasad Dasi
Journal:  Cardiovasc Revasc Med       Date:  2019-06-29

4.  Fetal Transcatheter Trileaflet Heart Valve Hemodynamics: Implications of Scaling on Valve Mechanics and Turbulence.

Authors:  Hoda Hatoum; Shelley Gooden; Megan Heitkemper; Kevin M Blum; Jason Zakko; Martin Bocks; Tai Yi; Yen-Lin Wu; Yadong Wang; Christopher K Breuer; Lakshmi Prasad Dasi
Journal:  Ann Biomed Eng       Date:  2020-02-12       Impact factor: 3.934

5.  Impact of Leaflet Laceration on Transcatheter Aortic Valve-in-Valve Washout: BASILICA to Solve Neosinus and Sinus Stasis.

Authors:  Hoda Hatoum; Pablo Maureira; Scott Lilly; Lakshmi Prasad Dasi
Journal:  JACC Cardiovasc Interv       Date:  2019-07-08       Impact factor: 11.195

6.  On the Significance of Systolic Flow Waveform on Aortic Valve Energy Loss.

Authors:  Hoda Hatoum; Brandon L Moore; Lakshmi Prasad Dasi
Journal:  Ann Biomed Eng       Date:  2018-07-20       Impact factor: 3.934

7.  A turbulence in vitro assessment of On-X and St Jude Medical prostheses.

Authors:  Hoda Hatoum; Pablo Maureira; Lakshmi Prasad Dasi
Journal:  J Thorac Cardiovasc Surg       Date:  2019-02-21       Impact factor: 5.209

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.