Literature DB >> 10834151

Turbulence characteristics downstream of bileaflet aortic valve prostheses.

J S Liu1, P C Lu, S H Chu.   

Abstract

This study was focused on a series of in vitro tests on the turbulent flow characteristics of three bileaflet aortic valves: St. Jude Medical (SJM), CarboMedics (CM), and Edwards Tekna (modified Duromedics, DM). The flow fields of the valves were measured in a pulsatile flow model with a laser-Doppler anemometer (LDA) at the aortic sinus area downstream of the valves. The heart rate was set at 70 beats per minute, the cardiac output was maintained at 5 liters per minute, and the aortic pressure wave forms were kept within the physiological range. Cycle-resolved analysis was applied to obtain turbulence data, including mean velocity, Reynolds stresses, autocorrelation coefficients, energy spectral density functions, and turbulence scales. The Reynolds shear stresses of all three valves induced only minor damage to red blood cells, but directly damaged the platelets, increasing the possibility of thrombosis. The smallest turbulence length scale, which offers a more reliable estimate of the effects of turbulence on blood cell damage, was three times the size of red blood cells and five times the size of platelets. This suggests that there is more direct interaction with the blood cells, thus causing more damage.

Entities:  

Mesh:

Year:  2000        PMID: 10834151     DOI: 10.1115/1.429643

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  19 in total

1.  Numerical comparison of the closing dynamics of a new trileaflet and a bileaflet mechanical aortic heart valve.

Authors:  Chi-Pei Li; Po-Chien Lu
Journal:  J Artif Organs       Date:  2012-06-13       Impact factor: 1.731

2.  Measurements of steady flow through a bileaflet mechanical heart valve using stereoscopic PIV.

Authors:  Chris Hutchison; Pierre Sullivan; C Ross Ethier
Journal:  Med Biol Eng Comput       Date:  2010-11-16       Impact factor: 2.602

3.  The effect of implantation orientation of a bileaflet mechanical heart valve on kinematics and hemodynamics in an anatomic aorta.

Authors:  Iman Borazjani; Fotis Sotiropoulos
Journal:  J Biomech Eng       Date:  2010-11       Impact factor: 2.097

4.  In vitro hemodynamic assessment of a novel polymeric transcatheter aortic valve.

Authors:  Megan Heitkemper; Hoda Hatoum; Lakshmi Prasad Dasi
Journal:  J Mech Behav Biomed Mater       Date:  2019-06-19

5.  Particle image velocimetry study of pulsatile flow in bi-leaflet mechanical heart valves with image compensation method.

Authors:  Yubing Shi; Tony Joon Hock Yeo; Yong Zhao; Ned H C Hwang
Journal:  J Biol Phys       Date:  2007-03-28       Impact factor: 1.365

Review 6.  Review of numerical methods for simulation of mechanical heart valves and the potential for blood clotting.

Authors:  Mohamad Shukri Zakaria; Farzad Ismail; Masaaki Tamagawa; Ahmad Fazli Abdul Aziz; Surjatin Wiriadidjaja; Adi Azrif Basri; Kamarul Arifin Ahmad
Journal:  Med Biol Eng Comput       Date:  2017-07-26       Impact factor: 2.602

7.  Determination of Reynolds Shear Stress Level for Hemolysis.

Authors:  Choon-Sik Jhun; Megan A Stauffer; John D Reibson; Eric E Yeager; Raymond K Newswanger; Joshua O Taylor; Keefe B Manning; William J Weiss; Gerson Rosenberg
Journal:  ASAIO J       Date:  2018 Jan/Feb       Impact factor: 2.872

8.  Device Thrombogenicity Emulator (DTE)--design optimization methodology for cardiovascular devices: a study in two bileaflet MHV designs.

Authors:  Michalis Xenos; Gaurav Girdhar; Yared Alemu; Jolyon Jesty; Marvin Slepian; Shmuel Einav; Danny Bluestein
Journal:  J Biomech       Date:  2010-05-21       Impact factor: 2.712

9.  The effect of turbulent viscous shear stress on red blood cell hemolysis.

Authors:  Jen-Hong Yen; Sheng-Fu Chen; Ming-Kai Chern; Po-Chien Lu
Journal:  J Artif Organs       Date:  2014-03-12       Impact factor: 1.731

10.  Stented valve dynamic behavior induced by polyester fiber leaflet material in transcatheter aortic valve devices.

Authors:  Hoda Hatoum; Frederick Heim; Lakshmi Prasad Dasi
Journal:  J Mech Behav Biomed Mater       Date:  2018-06-28
View more