Sintayehu Ambachew1, Kasaw Adane2, Abebaw Worede3, Tadele Melak1, Daniel Asmelash1, Shewaneh Damtie3, Habtamu Wondifraw Baynes1, Molla Abebe1, Belete Biadgo1.
Abstract
BACKGROUND: Laboratory services have been described as the major processes contributing to safe patient care in the modern healthcare sector. However, occurrences of errors in the overall testing processes impair the clinical decision-making process. Such errors are supposed to be high in resource-poor countries, like Ethiopia. The objective of this study was to assess errors in the total testing process in the Clinical Chemistry laboratory of the University of Gondar Hospital, Northwest Ethiopia.Entities:
Keywords: Analytical errors; clinical laboratory; post-analytical errors; pre-analytical errors; quality
Mesh:
Year: 2018 PMID: 29983521 PMCID: PMC6016342 DOI: 10.4314/ejhs.v28i2.15
Source DB: PubMed Journal: Ethiop J Health Sci ISSN: 1029-1857
Quality indicators in the pre-analytical, analytical and post-analytical phases
| Quality indicators of pre-analytical phase | |
| Number of requests with errors concerning patient name/total number of requests | |
| Number of requests with errors concerning hospital number/total number of requests | |
| Number of requests with errors concerning patient sex/total number of requests | |
| Number of requests with errors concerning patient age/total number of requests | |
| Number of requests with errors concerning physician name/total number of requests | |
| Number of requests with errors concerning patient clinical data/total number of requests | |
| Number of requests with errors concerning patient location/total number of requests | |
| Number of requests with errors concerning date of sampling/total number of requests | |
| Number of requests with errors concerning test ordered/total number of requests | |
| Number of requests with errors concerning time of sampling/total number of requests | |
| Number of samples haemolysed/total number of samples | |
| Number of samples lipemic /total number of samples | |
| Number of insufficient volume of samples/ total number of samples | |
| Number of samples mislabeled /total number of samples | |
| Number of samples with no request/total number of samples | |
| Number of IQC values that exceed the selected target /total quality control run | |
| Number of EQC exceeding the target range in EQAS-PT schemes / total number of quality control runin EQA | |
| Number of reports delivered outside the specified time/total number of reports | |
| Number of critical values not communicated/total number of reports | |
| Number of data transcription errors/total number of reports |
EQAS: External Quality Assessment schemes, EQC: external quality control, IQC: Internal Quality Control, IPD: Inpatient Department, OPD: Out Patient Department, PT: Proficiency Testing, UOG: University of Gondar
Frequency of missed data on routinely submitted test request forms in Clinical Chemistry laboratory at UOG Hospital from February to March 2016, Northwest Ethiopia
| Data type | OPD N(%) n=2287 | IPD N(%) n=972 | Total N(%) n=3259 | Sig. (2-tailed) |
| Patient name | 0(0.0) | 0(0.0) | 0(0.0) | |
| Hospital number | 1(0.04) | 3(0.3) | 4(0.1) | 0.048 |
| Patient sex | 13(0.6) | 5(0.5) | 18(0.6) | 0.849 |
| Patient age | 14(0.6) | 5(0.5) | 19(0.6) | 0.737 |
| Physician name | 1006(44.0) | 256(26.3) | 1262(38.7) | 0.000 |
| Patient clinical data | 2269(99.2) | 957(98.5) | 3226(99.0) | 0.049 |
| Patient location | 32(1.4) | 28(2.9) | 60(1.8) | 0.000 |
| Date of sampling | 1072(46.9) | 235(24.2) | 1307(40.1) | 0.000 |
| Test ordered | 1(0.04) | 0(0.0) | 1(0.03) | 0.515 |
| Time of sampling | 2281(99.7) | 956(98.4) | 3237(99.3) | 0.000 |
| Over all incomplete request form | 2286(99.9) | 970(99.8) | 3256(99.9) | 0.160 |
Type and frequency of reason for sample rejected in Clinical Chemistry laboratory at UOG Hospital from February to March 2016, Northwest Ethiopia.
| Parameter | OPD | IPD | Total |
| Haemolysis | 19(0.6) | 22(0.7) | 41(1.3) |
| Lipemic sample | 4(0.1) | 0(0.0) | 4(0.1) |
| Insufficient volume of sample | 0(0.0) | 2(0.1) | 2(0.1) |
| Mislabeled | 29(0.9) | 6(0.2) | 35(1.1) |
| Sample with no request/request with no | 28(0.9) | 11(0.3) | 39(1.2) |
| Test not ordered/inappropriate test | 0(0.0) | 2(0.1) | 2(0.1) |
| Total | 80(2.5) | 43(1.3) | 123(3.8) |
IPD: Inpatient Department, OPD: Out Patient Department, UOG: University of Gondar
Error frequency of analytical phase in Clinical Chemistry laboratory at UOG Hospital from February to March 2016, Northwest Ethiopia.
| IQC N (%) | EQC N (%) | Total N (%) | |
| Pass | 481(85.6) | 17(48.6) | 498(83.4) |
| Fail | 81(14.4) | 18(51.4) | 99(16.6) |
| Total | 562(100.0) | 35(100.0) | 597(100.0) |
IQC: Internal Quality Control, EQC: External Quality Control
Error frequency of post-analytical phase in Clinical Chemistry laboratory at UOG Hospital from February to March 2016, Northwest Ethiopia.
| Parameter | (n = 3136) |
| Excessive TAT | 270 (8.6) |
| Not informed critical result | 15 (0.48) |
| Data transcription errors | 6 (0.2) |
| Total | 291 (9.3) |
TAT: turnaround time, UOG: University of Gondar
Distribution of total testing process error frequency in the Clinical Chemistry laboratory at UOG Hospital from February to March 2016, Northwest Ethiopia.
| Phase | Error frequency (%) | Error frequency (%) |
| A | B | |
| Pre-analytical*6518 | 3379(51.8) | 3379(89.6) |
| Analytical*597 | 99(16.6) | 99(2.6) |
| Post-analytical*3136 | 291(9.3) | 291(7.7) |
| Total*10251 | 3769(36.8) | 3769(100) |
The asterisk (*) indicates denominator for each phase in the ‘A’ column, A: Error frequency in the pre-analytical, analytical, and post-analytical phases, B: Overall percent of errors in the three analytical phases