| Literature DB >> 33488088 |
Mulugeta Teshome1, Abebaw Worede2, Daniel Asmelash2.
Abstract
BACKGROUND: Currently, the use of clinical laboratory tests is growing at a promising rate and about 80% of the clinical decisions made are based on the laboratory test results. Therefore, it is a major task to achieve quality service. This study was conducted to assess the magnitude of errors in the total testing process of Clinical Chemistry Laboratory and to evaluate analytical quality control using sigma metrics.Entities:
Keywords: Clinical Chemistry Laboratory; Ethiopia; analytical errors; post-analytical errors; pre-analytical errors; sigma metrics
Year: 2021 PMID: 33488088 PMCID: PMC7815085 DOI: 10.2147/JMDH.S286679
Source DB: PubMed Journal: J Multidiscip Healthc ISSN: 1178-2390
Frequency of Missed Information on Laboratory Request Forms
| S. No. | Variables | Missed Information |
|---|---|---|
| %(Error/Total) | ||
| 1 | MRN | 0(0/4719) |
| 2 | Patient age | 26.9(1271/4719) |
| 3 | Gender of patient | 24.1(1138/4719) |
| 4 | Signature of the physician | 77.5(3658/4719) |
| 5 | Clinical history of the patient | 100(4719/4719) |
| 6 | Location of the patient (Clinic/ward) | 66.2(3122/4719) |
| 7 | Date of ordered | 57.5(2715/4719) |
| 8 | Test ordered | 0(0/4719) |
| 9 | Appropriate and authorized requests | 1.7(80/4719) |
| Total | 39.3(16,703/42,471) | |
Abbreviation: MRN, medical record number.
Analysis of Errors in Pre-Analytical Quality Indicators
| S. No. | Variables | Pre-Analytical Errors | |
|---|---|---|---|
| %(Error/Total) | % Total Errors | ||
| 1 | Hemolyzed specimen | 2.4(111/4719) | 35.4 |
| 2 | Icteric specimen | 2.1(97/4719) | 30.9 |
| 3 | Mislabeled specimen | 1(45/4719) | 14.3 |
| 4 | Lipemic specimen | 0.5(25/4719) | 8 |
| 5 | Inadequate specimen | 0.3(14/4719) | 4.5 |
| 6 | Clotted sample | 0.3(13/4719) | 4.1 |
| 7 | Specimen collected at wrong time | 0.2(8/4719) | 2.5 |
| 8 | Test tube broken in the centrifuge | 0.02(1/4719) | 0.3 |
| 9 | Specimen collected with wrong test tube | 0(0/4719) | 0 |
| 10 | Specimen without its request | 0(0/4719) | 0 |
| 11 | Specimen collected for unavailable test | 0(0/4719) | 0 |
| Overall rejection rate of specimens | 4.7(223/4719) | ||
| Total pre-analytical errors from all pre analytical QIs (Including laboratory test requests form and specimens) | 18(17,017/94,380) | ||
Abbreviation: QIs, quality indicators.
Analysis of Errors in Analytical Quality Indicators
| S. No. | Variables | Analytical Error |
|---|---|---|
| %(Error/Total) | ||
| 1 | Non-linear results released without retesting | 100(38/38) |
| 2 | Questionable results contradicted | 100(28/28) |
| 3 | EQC failed | 23.1(15/65) |
| 4 | Preventive maintenance not performed | 16.7(9/54) |
| 5 | Equipment malfunction | 9.1(4/44) |
| 6 | Reference range not available | 6.3(1/16) |
| 7 | Laboratory fail to establish linearity range | 6.3(1/16) |
| 8 | IQC result failed | 6.2(81/1317) |
| 9 | Failure to perform daily IQC | 5.3(73/1382) |
| 10 | Working reagents prepared incorrectly | 4.5(4/88) |
| 11 | Electric interruption | 3(141/4719) |
| 12 | Reagent stock out during analysis | 1.3(62/4719) |
| 13 | Calibration failed | 0(0/13) |
| 14 | Methods not updated upon new reagent | 0(0/16) |
| 15 | Reagents used after their expiration date | 0(0/691) |
| Total Errors from all analytical QIs | 3.5(457/13,206) | |
Abbreviations: EQC, external quality control; IQC, internal quality control; QIs, quality indicators.
Analysis of Errors in Post-Analytical Quality Indicators
| S. No. | Variables | Post-Analytical Error |
|---|---|---|
| %(Error/Total) | ||
| 1 | Requests released without result verification | 99.5(4694/4719) |
| 2 | Critical values not communicated immediately | 13.3(6/45) |
| 3 | Unrecorded test results | 1.3(62/4719) |
| 4 | Results released out of TAT | 1.1(53/4719) |
| 5 | All tests not performed as requested | 0.2(8/4719) |
| 6 | Result printouts attached to the wrong request | 0.1(4/4719) |
| 7 | Left over samples not retained as the policy | 0(0/4719) |
| 8 | Laboratory requests with results lost | 0(0/4719) |
| 9 | Request with incorrect unit of reporting | 0(0/4719) |
| Total Errors from all post analytical QIs | 12.8(4827/37,797) | |
Abbreviations: TAT, turnaround time; QIs, quality indicators.
Distribution of Errors Frequency in the Total Testing Process
| Phase | Error | % Total Errors | % Errors | Chi Square Test | |||
|---|---|---|---|---|---|---|---|
| Yes | No | ||||||
| Pre-analytical | 17,017 | 77,363 | 11.7 | 76.3 | p-value | ||
| Analytical | 457 | 12,749 | 0.3 | 2.1 | 2153 | 2 | <0.0001 |
| Post-analytical | 4827 | 32,970 | 3.3 | 21.6 | |||
| Total | 22,301 | 123,082 | 15.3 | 100 | |||
Association of Patient Location with Pre-Analytical and Post-Analytical Errors
| Variables | Location of Patients (Clinic/Ward) | Test of Significance | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Emergency | OPD | IPD | Unknown | ||||||
| Yes | No | Yes | No | Yes | No | Yes | No | ||
| Patient age missed | 9 | 144 | 116 | 763 | 21 | 524 | 1125 | 2017 | 392*(p<0.0001) |
| Gender of patient missed | 5 | 148 | 72 | 807 | 17 | 528 | 1044 | 2098 | 432*(p<0.0001) |
| Signature of the physicians missed | 69 | 84 | 599 | 280 | 318 | 227 | 2672 | 470 | 353.5*(p<0.0001) |
| Location of the patient missed | 5 | 148 | 6 | 873 | 8 | 537 | 3103 | 39 | 4463.5*(p<0.0001) |
| Date of ordered missed | 62 | 91 | 335 | 544 | 161 | 384 | 2157 | 985 | 487.5*(p<0.0001) |
| Unauthorized requests | 0 | 153 | 9 | 870 | 1 | 544 | 70 | 3072 | 17.8*(p<0.0001) |
| Hemolyzed specimen | 6 | 147 | 17 | 862 | 21 | 524 | 67 | 3075 | 8.3*(p=0.04) |
| Icteric specimen | 4 | 149 | 9 | 870 | 23 | 522 | 61 | 3081 | 17.77*(p<0.0001) |
| Mislabeled specimen | 1 | 152 | 4 | 875 | 3 | 542 | 37 | 3105 | 5.06(p=0.167) |
| Specimen rejection | 9 | 144 | 31 | 848 | 43 | 502 | 140 | 3002 | 15.9*(p=0.001) |
| Unrecorded results | 1 | 152 | 21 | 858 | 9 | 536 | 31 | 3111 | 11.4*(p=0.010) |
| Results released out of TAT | 3 | 150 | 14 | 865 | 10 | 535 | 26 | 3116 | 7.67(p=0.053) |
Note: *Statistically significant association (p<0.05).
Abbreviations: IPD, impatient department; OPD, outpatient department; TAT, turnaround time.
Factors Associated with Prolonged TAT
| Variable | Category | Prolonged TAT | COR(95% CI) | AOR(95% CI) | p-Value | |
|---|---|---|---|---|---|---|
| Yes | No | |||||
| Clinic/ward | Emergency | 3 | 150 | 1.66(0.49–5.63) | 1.66(0.49–5.63) | 0.412 |
| OPD | 14 | 865 | 1.596(0.82–3.09) | 1.596(0.82–3.09) | 0.165 | |
| IPD | 10 | 535 | 1.58(0.75–3.34) | 1.59(0.75–3.34) | 0.224 | |
| Unknown | 26 | 3116 | 1 | 1 | ||
| Shift | 1st shift | 13 | 3410 | 1 | 1 | |
| 2nd shift | 40 | 1256 | 7.83(4.15–14.77) | 8.35(4.45–15.67) | <0.001 | |
Note: p < 0.05= significant association.
Abbreviations: AOR, adjusted odds ratio; CI, confidence interval; COR, crude odds ratio; IPD, impatient department; OPD, outpatient department; TAT, turnaround time.
Total Allowable Error (TEa), CV, Bias, Sigma Value and QGI of the Routine Clinical Chemistry Tests
| Analyte | IQC Level | TEa% | Average Bias (%) | CV (%) | Sigma | QGI | Problem |
|---|---|---|---|---|---|---|---|
| AST | I | 20 | 7.4 | 10.6 | 1.2 | 0.46 | Imprecision |
| II | 20 | 7.4 | 6 | 2.1 | 0.82 | Imprecision and inaccuracy | |
| ALT | I | 20 | 4.2 | 9.6 | 1.6 | 0.3 | Imprecision |
| II | 20 | 4.2 | 7 | 2.3 | 0.4 | Imprecision | |
| ALP | I | 30 | 2.7 | 7 | 3.9 | 0.26 | Imprecision |
| II | 30 | 2.7 | 7.2 | 3.8 | 0.3 | Imprecision | |
| UR | I | 19 | 7 | 8.6 | 1.4 | 0.54 | Imprecision |
| II | 19 | 7 | 6.8 | 1.8 | 0.7 | Imprecision | |
| CRE | I | 15 | 5 | 7.2 | 1.4 | 0.46 | Imprecision |
| II | 15 | 5 | 6.6 | 1.5 | 0.5 | Imprecision | |
| GLU | I | 10 | 5.4 | 7.6 | 0.6 | 0.47 | Imprecision |
| II | 10 | 5.4 | 5.6 | 0.8 | 0.6 | Imprecision | |
| TC | I | 10 | 4 | 7.6 | 0.8 | 0.35 | Imprecision |
| II | 10 | 4 | 7.2 | 0.8 | 0.4 | Imprecision | |
| TG | I | 15 | 13.7 | 11.8 | 0.1 | 0.77 | Imprecision |
| II | 15 | 13.7 | 16 | 0.1 | 0.6 | Imprecision | |
| HDL | I | 20 | 18.7 | 18.9 | 0.1 | 0.7 | Imprecision |
| II | 20 | 18.7 | 10.8 | 0.1 | 1.1 | Imprecision and Inaccuracy | |
| LDL | I | 20 | 19.8 | 6 | 0.03 | 2.2 | Inaccuracy |
| II | 20 | 2.6 | 18.4 | 0.9 | 0.1 | Imprecision | |
| TP | I | 10 | 6.3 | 5.9 | 0.6 | 0.7 | Imprecision |
| II | 10 | 6.3 | 4.2 | 0.9 | 1 | Imprecision and inaccuracy | |
| ALB | I | 10 | 9.9 | 7.2 | 0.01 | 0.9 | Imprecision and inaccuracy |
| II | 10 | 9.9 | 6.2 | 0.02 | 1 | Imprecision and inaccuracy | |
| TBIL | I | 20 | 9.9 | 7.4 | 1.4 | 0.9 | Imprecision and inaccuracy |
| II | 20 | 9.9 | 6 | 1.7 | 1.1 | Imprecision and inaccuracy | |
| UA | I | 17 | 6.3 | 8.3 | 1.3 | 0.5 | Imprecision |
| II | 17 | 6.3 | 5.7 | 1.9 | 0.7 | Imprecision |
Abbreviations: CV, coefficient of variation; EQC, external quality control; IQC, internal quality control; QGI, quality goal index ratio; TE, total observed error; TEa, total allowable error.
Quality Control Strategy Based on the Sigma Value of the Test Parameters
| Parameters | Sigma Metrics | Levels of Control | Run | Westgard Rules | Status of the Method | Decision |
|---|---|---|---|---|---|---|
| AST, ALT, UR, CRE, GLU, TC, TG, HDL. LDL, TP, ALB, TBIL, UA | <3 | – | – | – | Poor | Needs rechecking |
| ALP | 3.8 | 2 | 4 | 13S/22S/R4S/41S/10x | Good | Acceptable |
Note: No Westgard rule, IQC level and runs selected rather immediate trouble shooting requires.