OBJECTIVES: To systematically investigate and directly compare, for the first time, the sample stability of a large number of quantitative fecal immunochemical tests (FITs) at different storage conditions.Entities:
Mesh:
Substances:
Year: 2018 PMID: 29976921 PMCID: PMC6033915 DOI: 10.1038/s41424-018-0035-2
Source DB: PubMed Journal: Clin Transl Gastroenterol ISSN: 2155-384X Impact factor: 4.488
Overview of ten quantitative FITs
| FIT brand | Manufacturer | Fecal sampling device (fecal mass/buffer volume) | Analytical instrument | Analytical range (µg Hb/g feces) | Preset threshold (µg Hb/g feces) |
|---|---|---|---|---|---|
|
| |||||
| CAREprime Hb | Alfresa Pharma (Osaka, Japan) | Specimen collection container A (10 mg/1.9 ml) | CAREprime | 0.76–228 | 6.30 |
| ELISA Test Hb | ImmuChrom (Heppenheim, Germany) | Frost Diagnostika stool collecting tube (40 mg/2 ml) | ELx 800 ELISA Reader | 0.44–21 | 2.0 |
| IDK Hb ELISA | Immundiagnostik (Bensheim, Germany) | IDK extract (15 mg/1.5 ml) | Dynex DSX | 0.086–50 | 2.0 |
| OC Sensor | Eiken Chemical (Tokyo, Japan) | OC auto-sampling bottle 3 (10 mg/2.0 ml) | OC Sensor io | 10.0–200 | 10.0 |
| RIDASCREEN Hb | R-Biopharm (Darmstadt, Germany) | RIDA TUBE Hb (10 mg/2.5 ml) | Dynex DSX | 0.65–50 | 8.0 |
| SENTiFIT-FOB Gold | Sentinel Diagnostics (Milan, Italy) | SENTiFIT pierceTube (10 mg/1.7 ml) | SENTiFIT 270 analyzer | 1.70–129.88 | 17.0 |
|
| |||||
| Eurolyser FOB test | Eurolyser Diagnostica (Salzburg, Austria) | Eurolyser FOB sample Collector (19.9 mg/1.6 ml) | Eurolyser CUBE | 2.01–80.4 | 8.04 |
| immoCARE-C | CARE diagnostica a (Voerde, Germany) | Sample collection tube (20 mg/2.5 ml) | immoCARE Test cassette + CAREcube | 3.75–250 | 6.25 |
| QuantOn Hem | Immundiagnostik (Bensheim, Germany) | QuantOn Hem TUBE (15 mg/1.5 ml) | QuantOn Hem Test cassette + Smartphone App/iOSa | 0.30–100 | 3.70 |
| QuikRead go iFOBT | Orion Diagnostica (Espoo, Finland) | QuikRead go iFOBT Sampling Set (10 mg/2.0 ml) | QuikRead go | 15–200 | 15.0 |
FIT fecal immunochemical test, Hb Hemoglobin, App mobile application software, iOS iPhone operating system
aiPhone 6 s with special software for test analysis (designed by the manufacturer) was used for this study
Fig. 1Flow diagram of the workflow in this study.
FSD fecal sampling device, FIT fecal immunochemical test
Fig. 2Hemoglobin concentration according to storage conditions.
Hb hemoglobin. Black horizontal dashed lines = upper and lower analytical limit. *The analysis of CAREprime is based on 19 samples on the seventh day at 20 °C storage, whereas all other analyses are based on all 20 samples
Fig. 3Positivity rate at preset manufacturers' thresholds at a 5, b 20, and c 35 °C, respectively. Legend in (a) is also applicable to b and c. *The analysis of CAREprime is based on 19 samples on the seventh day at 20 °C storage, whereas all other analyses are based on all 20 samples
Fig. 4Positivity rate at adjusted thresholds yielding an equal positivity rate of 85% on the first day at a 5, b 20, and c 35 °C, respectively. Legend in (a) is also applicable to b and c. *The analysis of CAREprime is based on 19 samples on the seventh day at 20 °C storage, whereas all other analyses are based on all 20 samples.