| Literature DB >> 29975714 |
Simon M F Triphan1,2,3, Jürgen Biederer1,2,3,4, Kerstin Burmester1,2,3, Iven Fellhauer1,2,3, Claus F Vogelmeier3,5, Rudolf A Jörres3,6,7, Hans-Ulrich Kauczor1,2,3, Claus P Heußel1,2,3,8, Mark O Wielpütz1,2,3,8, Bertram J Jobst1,2,3,8.
Abstract
INTRODUCTION: As there is an increasing number of multicentre lung imaging studies with MRI in patients, dedicated reference phantoms are required to allow for the assessment and comparison of image quality in multi-vendor and multi-centre environments. However, appropriate phantoms for this purpose are so far not available commercially. It was therefore the purpose of this project to design and apply a cost-effective and simple to use reference phantom which addresses the specific requirements for imaging the lungs with MRI.Entities:
Mesh:
Year: 2018 PMID: 29975714 PMCID: PMC6033396 DOI: 10.1371/journal.pone.0199148
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Sequence parameters of the study protocol used on the Siemens Aera Scanners for in vivo measurements and phantom measurements.
| Sequence unit | md | or | TR | TE | FoV | d | voxel size | matrix | pf | TA |
|---|---|---|---|---|---|---|---|---|---|---|
| VIBE | 3D | cor | 3.61 | 1.63 | 400×400 | 4.0 | 1.39×1.39 | 288×288 | 2 | 0:16 |
| VIBE | 3D | tra | 3.29 | 1.61 | 400×300 | 4.0 | 1.25×1.25 | 320×240 | 2 | 0:16 |
| HASTE | 2D | cor | 314.0 | 20.0 | 400×400 | 6.0 | 0.78×0.78 | 512×512 | 3 | 0:13 |
| HASTE | 2D | tra | 500.0 | 27.0 | 450×366 | 8.0 | 1.41×1.41 | 320×260 | 2 | 0:35 |
| TrueFISP | 2D | cor | 448.9 | 1.17 | 400×400 | 4.5 | 0.78×0.78 | 512×512 | 3 | 2:20 |
| BLADE | 2D | cor | 905.0 | 73.0 | 400×400 | 6.0 | 1.25×1.25 | 320×320 | 2 | 2:13 |
| HASTE IRM | 2D | tra | 502.0 | 72.0 | 400×400 | 6.0 | 1.56×1.56 | 256×256 | 2 | 0:38 |
| Angio | 3D | cor | 2.80 | 1.04 | 350×400 | 1.8 | 1.04×1.04 | 336×384 | 3 | 0:16 |
| TWIST | 3D | cor | 1.73 | 0.76 | 366×450 | 5.0 | 1.76×1.76 | 208×256 | 2 | 0:37 |
| VIBE FS | 3D | tra | 3.29 | 1.61 | 400×300 | 4.0 | 1.25×1.25 | 320×240 | 2 | 0:17 |
Shown are excitation mode, slice/slab orientation, repetition time, echo time, field of view, voxel size, image matrix, parallel imaging factor and total acquisition time. Note that the in vivo study contains additional repetitions of identical sequences in different respiratory states as well as after contrast agent injection.
Fig 1Photo of a phantom instance.
Fig 2Gradient echo and turbo spin echo MR images of the phantom.
a: A single, central slice of a 3D VIBE acquisition showing all phantom compartments. b: A slice acquired using HASTE. Slices are equivalent to coronal orientation.
Mean and standard deviation of SNR over all 15 scanners measured at baseline of the in vivo study using 12 phantom instances, ordered by sequence and phantom compartment.
| Sequence | Lung | Blood | Muscle | Fat |
|---|---|---|---|---|
| VIBE cor | 6.1±3.5 | 30.3±13.2 | 39.6±18.4 | 49.9±24.3 |
| VIBE tra | 6.5±3.3 | 33.3±12.3 | 42.1±16.7 | 52.8±20.3 |
| HASTE cor | 18.7±9.4 | 215.4±85.3 | 81.4±28.2 | 166.2±60.7 |
| HASTE tra | 26.5±13.7 | 234.6±99.1 | 99.4±41.4 | 176.6±74.5 |
| TrueFISP | 4.4±2.7 | 112.3±55.6 | 32.3±20.6 | 58.8±24.9 |
| BLADE | 3.0±1.7 | 118.2±81.0 | 12.3±3.9 | 75.7±24.2 |
| HASTE IRM | 4.8±2.5 | 125.5±30.5 | 15.7±6.8 | 4.4±2.3 |
| Angio | 1.2±0.3 | 2.8±1.6 | 4.9±2.7 | 15.4±3.6 |
| TWIST | 3.0±2.2 | 8.5±5.2 | 13.6±8.0 | 51.7±32.7 |
| VIBE FS | 3.9±2.0 | 29.9±8.7 | 39.4±15.2 | 12.6±7.9 |
Fig 3SNR obtained from all 15 scanner units (using 12 phantom instances) for the coronal VIBE sequence, sorted by MR scanner type and phantom compartment (a-d), as well as SNR for the HASTE, TWIST, BLADE and TrueFISP sequences, given for the lung compartment only (e-h).
Scanners using 33mT/m gradient systems are marked using stars. Please note that only the initial phantom measurements acquired at baseline of the multicentre in vivo study are shown. Data from follow-up measurements performed at 3 month intervals were left out to maintain clarity.
Fig 4In-plane (a) and through-plane (b) profile widths detected in three selected sequences, sorted by site and scanner type. While the actual diameter of the rod is 6mm, these values are derived from the FWHM of a fitted function and thus should be smaller. (c) Shows the edge widths detected at the muscle-equivalent compartment in phase encoding direction. Hatched bars represent MR scanners with 33mT/m gradient systems. Please note that only the phantom measurements acquired at baseline of the in vivo study are displayed here.