Literature DB >> 29962919

Supply-Doubled Pulse-Shaping High Voltage Pulser for CMUT Arrays.

Gwangrok Jung1, Coskun Tekes2, Amirabbas Pirouz2, F Levent Degertekin2, Maysam Ghovanloo1.   

Abstract

A supply-doubled pulse-shaping high voltage (HV) pulser is presented for medical ultrasound imaging applications, particularly those that use capacitive micromachined ultrasonic transducers (CMUT). The pulser employs a bootstrap circuit combined with dynamically-biased stacked transistors, which allow HV operation above process limit without lowering device reliability. The new pulser overcomes supply voltage limitation of conventional unipolar pulsers by generating output signals that are almost twice the supply level. It also can generate three-level pulses to further optimize the transmit pressure signals. A proof-of-concept prototype has been implemented in 0.18-μm HV CMOS/DMOS technology with 60 V devices. Measurement results show that the HV pulser can safely generate controllable three-level pulses with up to 85 Vpp from 45 V supply. Acoustic measurements are conducted connecting the pulser to a CMUT with 2 pF capacitance and 8.3 MHz center frequency. The pulse shape has been adjusted for the CMUT under test to generate maximum pressure output and the results are in good agreement with a large signal CMUT model.

Entities:  

Keywords:  High voltage pulser; capacitive micromachined ultrasound transducer (CMUT); medical ultrasound imaging

Year:  2017        PMID: 29962919      PMCID: PMC6022846          DOI: 10.1109/TCSII.2017.2691676

Source DB:  PubMed          Journal:  IEEE Trans Circuits Syst II Express Briefs        ISSN: 1549-7747            Impact factor:   3.292


  9 in total

1.  Harmonic reduction in capacitive micromachined ultrasonic transducers by gap feedback linearization.

Authors:  Sarp Satir; F Levent Degertekin
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2012-01       Impact factor: 2.725

2.  An integrated circuit with transmit beamforming flip-chip bonded to a 2-D CMUT array for 3-D ultrasound imaging.

Authors:  Ira O Wygant; Nafis S Jamal; Hyunjoo J Lee; Amin Nikoozadeh; Omer Oralkan; Mustafa Karaman; Butrus T Khuri-Yakub
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2009-10       Impact factor: 2.725

3.  Transceiver Design for CMUT-Based Super-Resolution Ultrasound Imaging.

Authors:  Parisa Behnamfar; Reza Molavi; Shahriar Mirabbasi
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2015-05-13       Impact factor: 3.833

4.  Capacitive micromachined ultrasonic transducers for medical imaging and therapy.

Authors:  Butrus T Khuri-Yakub; Omer Oralkan
Journal:  J Micromech Microeng       Date:  2011-05       Impact factor: 1.881

5.  A large-signal model for CMUT arrays with arbitrary membrane geometry operating in non-collapsed mode.

Authors:  Sarp Satir; Jaime Zahorian; F Levent Degertekin
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2013-11       Impact factor: 2.725

6.  Front-end receiver electronics for high-frequency monolithic CMUT-on-CMOS imaging arrays.

Authors:  Gokce Gurun; Paul Hasler; F Degertekin
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2011-08       Impact factor: 2.725

7.  CMUTs with high-K atomic layer deposition dielectric material insulation layer.

Authors:  Toby Xu; Coskun Tekes; F Degertekin
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2014-12       Impact factor: 2.725

8.  Monolithic CMUT-on-CMOS integration for intravascular ultrasound applications.

Authors:  Jaime Zahorian; Michael Hochman; Toby Xu; Sarp Satir; Gokce Gurun; Mustafa Karaman; F Levent Degertekin
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2011-12       Impact factor: 2.725

9.  Single-chip CMUT-on-CMOS front-end system for real-time volumetric IVUS and ICE imaging.

Authors:  Gokce Gurun; Coskun Tekes; Jaime Zahorian; Toby Xu; Sarp Satir; Mustafa Karaman; Jennifer Hasler; F Levent Degertekin
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2014-02       Impact factor: 2.725

  9 in total
  3 in total

Review 1.  Circuits on miniaturized ultrasound imaging system-on-a-chip: a review.

Authors:  Jaemyung Lim
Journal:  Biomed Eng Lett       Date:  2022-05-12

2.  A Reduced-Wire ICE Catheter ASIC With Tx Beamforming and Rx Time-Division Multiplexing.

Authors:  Gwangrok Jung; Coskun Tekes; M Wasequr Rashid; Thomas M Carpenter; David Cowell; Steven Freear; F Levent Degertekin; Maysam Ghovanloo
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2018-11-16       Impact factor: 3.833

3.  Supply-Inverted Bipolar Pulser and Tx/Rx Switch for CMUTs Above the Process Limit for High Pressure Pulse Generation.

Authors:  Gwangrok Jung; Amirabbas Pirouz; Coskun Tekes; Thomas M Carpenter; David Cowell; Steven Freear; Maysam Ghovanloo; F Levent Degertekin
Journal:  IEEE Sens J       Date:  2019-08-28       Impact factor: 3.301

  3 in total