Measuring PET Spatial Resolution Using a Cylinder Phantom Positioned at an Oblique Angle.

Martin A Lodge1, Jeffrey P Leal2, Arman Rahmim2, John J Sunderland3, Eric C Frey2.   

Abstract

A cylinder phantom positioned at a slightly oblique angle with respect to the z-axis of a PET scanner allows for fine sampling of the edge-spread function. We show how this technique can be used to measure the spatial resolution that can be expected with clinical PET protocols, potentially providing more relevant estimates than are typically obtained with established experimental procedures.
Methods: A 20-cm-diameter water-filled cylinder phantom containing a uniform 18F solution was centrally positioned at a small angle with respect to the z-axis of a clinical PET/CT system. The oblique angle ensures that the phantom edge intersects the image matrix differently in different slices. Combining line profiles from multiple slices results in a composite profile with fine sampling. Spatial resolution was measured as the full width at half maximum (FWHM) by fitting a model to the finely sampled edge-spread functions in both radial and axial directions. The technique was validated by controlled modulation of image reconstruction parameters and by comparison with extended phantoms with fillable inserts. Separate experiments with uniform cylinders containing 18F, 11C, 13N, 68Ga, and 124I were used to further assess the proposed method.
Results: Controlled adjustment of a gaussian postreconstruction filter was accurately reflected in the measured FWHM values. Recovery coefficients derived using the cylinder FWHM values agreed closely with recovery coefficients derived from physical phantoms over a range of insert-to-background ratios, phantom geometries, and reconstruction protocols. The effect of increasing positron energy was clearly reflected in the FWHM values measured with different isotopes.
Conclusion: A method has been developed for measuring the spatial resolution that is achieved with clinical PET protocols, providing more relevant estimates than are typically obtained with established procedures. The proposed method requires no special equipment and is versatile, being capable of measuring resolution for different isotopes as well as for different reconstruction protocols. The new technique promises to aid standardization of PET data acquisition by allowing a more informed selection of reconstruction parameters.
© 2018 by the Society of Nuclear Medicine and Molecular Imaging.

Entities:  

Keywords:  PET; cylinder phantom; quality assurance; spatial resolution; standardization

Mesh:

Substances:

Year:  2018        PMID: 29903932      PMCID: PMC6225537          DOI: 10.2967/jnumed.118.209593

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  18 in total

1.  On the assessment of spatial resolution of PET systems with iterative image reconstruction.

Authors:  Kuang Gong; Simon R Cherry; Jinyi Qi
Journal:  Phys Med Biol       Date:  2016-02-11       Impact factor: 3.609

2.  Imaging as a quantitative science.

Authors:  Daniel C Sullivan
Journal:  Radiology       Date:  2008-08       Impact factor: 11.105

3.  A practical, automated quality assurance method for measuring spatial resolution in PET.

Authors:  Martin A Lodge; Arman Rahmim; Richard L Wahl
Journal:  J Nucl Med       Date:  2009-07-17       Impact factor: 10.057

4.  Quantitative PET/CT scanner performance characterization based upon the society of nuclear medicine and molecular imaging clinical trials network oncology clinical simulator phantom.

Authors:  John J Sunderland; Paul E Christian
Journal:  J Nucl Med       Date:  2014-12-18       Impact factor: 10.057

5.  The line spread function and modulation transfer function of a computed tomographic scanner.

Authors:  P F Judy
Journal:  Med Phys       Date:  1976 Jul-Aug       Impact factor: 4.071

6.  Validation of phantom-based harmonization for patient harmonization.

Authors:  Joseph V Panetta; Margaret E Daube-Witherspoon; Joel S Karp
Journal:  Med Phys       Date:  2017-06-09       Impact factor: 4.071

7.  Predicting the Future - Big Data, Machine Learning, and Clinical Medicine.

Authors:  Ziad Obermeyer; Ezekiel J Emanuel
Journal:  N Engl J Med       Date:  2016-09-29       Impact factor: 91.245

8.  Evaluation of strategies towards harmonization of FDG PET/CT studies in multicentre trials: comparison of scanner validation phantoms and data analysis procedures.

Authors:  Nikolaos E Makris; Marc C Huisman; Paul E Kinahan; Adriaan A Lammertsma; Ronald Boellaard
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-06-11       Impact factor: 9.236

9.  Simultaneous measurement of noise and spatial resolution in PET phantom images.

Authors:  Martin A Lodge; Arman Rahmim; Richard L Wahl
Journal:  Phys Med Biol       Date:  2010-01-28       Impact factor: 3.609

Review 10.  EANM/EARL harmonization strategies in PET quantification: from daily practice to multicentre oncological studies.

Authors:  Nicolas Aide; Charline Lasnon; Patrick Veit-Haibach; Terez Sera; Bernhard Sattler; Ronald Boellaard
Journal:  Eur J Nucl Med Mol Imaging       Date:  2017-06-16       Impact factor: 9.236

View more
  5 in total

1.  The impact of iterative reconstruction protocol, signal-to-background ratio and background activity on measurement of PET spatial resolution.

Authors:  Sahar Rezaei; Pardis Ghafarian; Mehrdad Bakhshayesh-Karam; Carlos F Uribe; Arman Rahmim; Saeed Sarkar; Mohammad Reza Ay
Journal:  Jpn J Radiol       Date:  2020-01-01       Impact factor: 2.374

2.  Measurement of PET Quantitative Bias In Vivo.

Authors:  Martin A Lodge; Wojciech Lesniak; Michael A Gorin; Kenneth J Pienta; Steven P Rowe; Martin G Pomper
Journal:  J Nucl Med       Date:  2020-10-09       Impact factor: 10.057

Review 3.  Current Status of Radiopharmaceutical Therapy.

Authors:  Sara St James; Bryan Bednarz; Stanley Benedict; Jeffrey C Buchsbaum; Yuni Dewaraja; Eric Frey; Robert Hobbs; Joseph Grudzinski; Emilie Roncali; George Sgouros; Jacek Capala; Ying Xiao
Journal:  Int J Radiat Oncol Biol Phys       Date:  2020-08-14       Impact factor: 7.038

4.  Comments on the NEMA NU 4-2008 Standard on Performance Measurement of Small Animal Positron Emission Tomographs.

Authors:  Patrick Hallen; David Schug; Volkmar Schulz
Journal:  EJNMMI Phys       Date:  2020-02-24

5.  Advanced quantitative evaluation of PET systems using the ACR phantom and NiftyPET software.

Authors:  Pawel J Markiewicz; Casper da Costa-Luis; J Dickson; A Barnes; G Krokos; J MacKewn; T Clark; C Wimberley; G MacNaught; M M Yaqub; J D Gispert; B F Hutton; P Marsden; A Hammers; A J Reader; S Ourselin; K Herholz; J C Matthews; F Barkhof
Journal:  Med Phys       Date:  2022-03-31       Impact factor: 4.506

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.