Nick Polychronakis1, Aspasia Sarafianou2, Alcibiadis Zissis1, Triantafillos Papadopoulos3. 1. Removable Prosthodontics, Dental School, National and Kapodistrian University of Athens, Greece. 2. Department of Prosthodontics, Dental School, National and Kapodistrian University of Athens, Greece. 3. Department of Biomaterials, Dental School, National and Kapodistrian University of Athens, Greece.
Abstract
OBJECTIVE: The aim of the present study was to evaluate the influence of thermocycling on the flexural strength of a polyamide base denture material. MATERIALS AND METHODS: A polyamide denture base material (Valplast) was tested, whereas a PMMA material (Vertex) was used as a control. Thirty specimens of each material were fabricated for flexural strength testing according to ISO 1567. They were prepared and stored in water at 37°C for 48 hours. The specimens of each material were divided into three equal groups (n=10). Flexural strength testing was performed immediately after water storage and after thermocycling (5°C / 55°C, 2 c/min) for 3000 and 5000 cycles. A three point flexural test was performed on a universal testing machine at a crosshead speed of 5 mm/min. The final flexural strength was calculated using the formula: FS = 3 PL/2 bd2. A two-way ANOVA with post-hoc analysis using Tukey's procedure was applied at .05 level of statistical significance. RESULTS: A statistically significant reduction in flexural strength was recorded after thermocycling at 3000 cycles for PMMA and at 5000 cycles for both materials. The flexural strength of PMMA was significantly higher compared to polyamide for all the conditions tested (p<0.05). CONCLUSION: Thermocycling had a significant adverse effect on the flexural strength of polyamide and PMMA denture base materials.
OBJECTIVE: The aim of the present study was to evaluate the influence of thermocycling on the flexural strength of a polyamide base denture material. MATERIALS AND METHODS: A polyamide denture base material (Valplast) was tested, whereas a PMMA material (Vertex) was used as a control. Thirty specimens of each material were fabricated for flexural strength testing according to ISO 1567. They were prepared and stored in water at 37°C for 48 hours. The specimens of each material were divided into three equal groups (n=10). Flexural strength testing was performed immediately after water storage and after thermocycling (5°C / 55°C, 2 c/min) for 3000 and 5000 cycles. A three point flexural test was performed on a universal testing machine at a crosshead speed of 5 mm/min. The final flexural strength was calculated using the formula: FS = 3 PL/2 bd2. A two-way ANOVA with post-hoc analysis using Tukey's procedure was applied at .05 level of statistical significance. RESULTS: A statistically significant reduction in flexural strength was recorded after thermocycling at 3000 cycles for PMMA and at 5000 cycles for both materials. The flexural strength of PMMA was significantly higher compared to polyamide for all the conditions tested (p<0.05). CONCLUSION: Thermocycling had a significant adverse effect on the flexural strength of polyamide and PMMA denture base materials.
Despite the increasing popularity of dental implants for full mouth rehabilitation, complete and partial dentures remain a useful treatment modality for the restoration of edentulous patients. The proper selection of the base resin denture polymer is a significant factor for the long term success of removable restorations. Polymethyl methacrylate (PMMA) has been the material of choice for denture base constructions for decades. Alternatives to PMMA include materials which have been introduced in dental market in the past decades, such as high impact polymers, poly-carbonates and polyamides (-).Polyamides (NH (CH2) mCOn) are known in everyday life as nylon. In their molecule bonds produced by an amine group (NH) and a terminal carbonyl component of a functional group (COOH) react to produce a carbon-nitrogen bond (amide). The amine group and the carboxylic acid group may either be on the same monomer, or the polymer may be constituted of two different bifunctional monomers, one with two amine groups and the other with two carboxylic acid or acid chloride groups (-). Nylon is synthetically produced and polymerized by condensation reaction and must specifically include a straight chain (aliphatic) monomer (, ).The first attempts to use polyamides as denture base materials were made in the 1950s, but they have been used extensively for this purpose after the introduction of new generations of nylon materials (-). Physical and mechanical properties favor the use of polyamides as denture base materials since they exhibit high flexibility, low density, high impact resistance and low water sorption and solubility (). They are non-toxic materials with a low possibility of allergic reactions. They offer relatively good color stability (, , ). One of the disadvantages of polyamides as denture base materials is their low modulus of elasticity, flexural and tensile strength (, ), their low adherence to denture liners (), the absence of chemical bond with acrylic teeth and the inability of repair (). Although a large number of studies (, , , ) have been published on the flexural strength of PMMA under different experimental conditions, there is a lack of evidence concerning the flexural behavior of polyamides (, , ).The ultimate flexure strength of a material reflects its potential to resist catastrophic failure under a flexural load. The high flexural strength of the denture base material is essential for the clinical success of removable partial restorations, given the fact that the alveolar resorption is a gradual, irregular process that may result in uneven support of tissue-borne prostheses. A high proportional limit is necessary in order for the base material to resist plastic deformation and enhanced fatigue resistance is essential to withstand repeated masticatory loading (-). The high flexibility of polyamides is necessary for the function of metal-free removable partial denture (RPD) clasps (). The effect of the presence of water, and consequently saliva, and the effect of temperature changes when hot and cold foods are consumed on the flexural properties of the polyamides have not been thoroughly investigated.The aim of the present study was to investigate the flexural strength of a polyamide denture base material in comparison to a conventional PMMA denture base material after they had been submitted to a large number of thermal cycles. The null hypothesis was that there would be no significant difference in the flexural strength before and after thermocycling and that the number of cycles would not significantly affect the flexural strength of the materials.
Materials and Method
Two denture base resin material, a PMMA (Vertex Rapid Simplified, Vertex-Dental B.V., The Netherlands) and a polyamide (Valplast, Valplast Int. Corp, Long Island City, NY, USA) were tested. Stainless steel patterns were used to fabricate a total of 60 bars (30 of each material), measuring 64 mm×10 mm×2.5 mm, according to the ISO 1567 specification ().
Specimen preparation
The specimen number was estimated before testing. For this purpose, G*Power software (G*Power v.3.1.5, Franz Faul, University of Kiel, Germany) were used.Three metal patterns were invested in a flask with ISO type III dental stone (Microstone, Whip-Mix, USA). Before investing, a wax sprue (3 mm in diameter) was positioned on every pattern. The patterns were removed after boil-out and the mold cavities were filled with the respective material for specimen fabrication.Before injection into the mold cavities, the polyamide material was plasticized in a digital melting Valplast furnace at 280°C for 11 minutes. The flask was pressed for 3min in a Valplast injection press and then allowed to bench cool before opening.The Vertex acrylic resin was fabricated according to the manufacturer's recommendation by mixing of 1ml of liquid (monomer) to 2.3 g of powder (polymer). When it reached the dough stage, the acrylic mass was inserted in the mold cavities and polymerized at 100°C for 20 minutes. After curing, the flasks were bench-cooled at room temperature for 30 minutes.All specimens were removed from the molds and the excess margins were trimmed with tungsten carbide burs. Subsequently, the specimens were polished up to 600 grits in a polishing unit (Ecomet III Buehler Ltd, Evanston, Ill, USA).The specimens were then stored in water at 37°C for 48 hours, according to ISO 1567.According to the material used, the specimens were divided in two groups of 30 (Group 1/Vertex, Group 2/Valplast). The specimens of each material were further divided into three equal subgroups (n=10) (Table 1). Subgroups A1 and B1 were submitted to flexural test immediately after storage in distilled water for 48 hours at 37°C. Subgroups A2 and B2 were further submitted to thermocycling for 3000 cycles, while groups A3 and B3 were submitted to thermocycling for 5000 cycles (5°C and 55°C, 2 cycles/min) (Table 1). The three point flexural testing was accomplished in a universal testing machine (Tensometer 10; Monsanto, Akron, Ohio) at a force applied with a crosshead speed of 5mm/min. Specimens were placed in a rig with incorporated supports distanced at 50 mm.
Table 1
Groups and subgroups of tested materials
Subgroups
A
B
C
48h (water 37°C)
3000 cycles (TC*)
5000 cycles (TC*)
Groups1 (Vertex)2 (Valplast)
A1
Β1
C1
A2
Β2
C2
*TC=Thermocycling
*TC=ThermocyclingThe fracture force (F) was recorded in Newtons (N) and the flexural strength (FS) was calculated in MPa following the formula FS = 3 PL/2 bd2 (P=maximum load, L= specimen length, b=specimen width, and d=specimen thickness). The Valplast failure load values (N) were derived from the stress-strain curve at the point of proportional limit.A two-way ANOVA to the ranked data was done to test the effect of material, number of thermal cycles and their interactions on the flexural strength using the post hoc Tukey’s test. All tests used a .05 level of statistical significance. Statistical software (Sigma Plot, Version 12.0, SSI, Jandel CA) was used for data analysis.
Results
The results of the flexural strength test revealed a statistically significant difference between the tested materials under the same experimental conditions (p<0.05). There was a statistically significant interaction between materials and thermocycling (p<0.05). Generally, both materials showed a decrease in flexural strength values when the number of cycles was increased. The statistical estimation among all hydro-thermocycling conditions was significant for all subgroups for both materials except for the subgroup of 48 hours to 3000 cycles for Valplast. Vertex flexural strength values after 48 hours storage in water were almost double compared to the ones of Valplast (119.13 to 60.31 MPa).The results of the flexural strength testing are presented in Table 2. A statistically significant decrease in the flexural strength of the PMMA material (Vertex) was recorded both after 3000 and 5000 thermal cycles (119.13 MPa compared to 99.82 MPa and 94.80 MPa respectively) (p<0.05). No statistically significant difference in flexural strength was recorded for 3000 thermal cycles when the polyamide material (Valplast) was examined (60.31 MPa compared to 56.34 MPa). However, thermocycling for 5000 cycles resulted in a statistically significant decrease in flexural strength values for Valplast (60.31 MPa compared to 35.39 MPa).
Table 2
Mean values and standard deviations of all subgroups in MPa
A1
A2
B1
B2
C1
C2
119.13(±7,28)a
60.31(±2,34)b
99.82(±14,02)c
56.34(±3,1)b
94.80(±13,06)d
35.39 (±2,67)e
Same superscript letters mean no statistical differences.
Same superscript letters mean no statistical differences.Flexural strength values of the two materials tested revealed a statistically significant difference for all the experimental conditions tested: after 48 hours of water storage the Vertex values were almost double than the ones of Valplast (119.13 to 60.31 MPa). A statistically significant difference between the two materials was found after submission to 3000 thermal cycles. After 5000 thermal cycles a statistically significant difference in flexural strength values between the two materials was also recorded.
Discussion
The null hypothesis was not verified because significant differences in flexural strength were recorded among the tested groups.From the results of the present study it may be concluded that temperature fluctuations and immersion in water during thermocycling decrease the flexural strength of both PMMA and polyamide denture base materials.In the present study, PMMA presented significantly higher flexural strength in relation to polyamide. Determining the failure point of the flexible polyamide materials during flexural testing has been a substantial problem. In the present study, the value corresponding to the proportional limit of the material in the stress–strain curve was presumed to be the failure point. The clinical importance of the flexural strength at the proportional limit is that it reflects the resistance of a material to plastic deformation.In a number of studies (, , , -), the flexural strength of polyamide materials in comparison to PMMA under a variety of experimental conditions has been examined. The values of the flexural strength of polyamide reported in the present study are in agreement with the results of other authors, who tested similar materials following the same calculation of the failure point (, ). The differences in the polyamide flexural strength values reported by other authors may be attributed to different microstructures of specific trademark materials tested or to the differences in the calculation of the failure point (, ). The high flexural strength values (163.62 MPa) recorded by Abhay et al. () may probably be due to different ways of calculating.Due to their high flexibility, polyamide denture base materials do not fracture in clinical situations even under extreme bite forces. They cannot be considered clinically efficient if plastic deformation exceeds the proportional limit, due to the resulting dimensional inaccuracies in the restoration (). In this case, denture base materials may lead to inferior masticatory ability and alveolar resorption. Masticatory bite forces which are exerted by adult men with full dentition range between 60 N and 305 N, with a mean value of 137 N and 150 N (, ). The mean value of the proportional point of Valplast in the present study was about 220 N after 3000 hydro-thermocycling as it can be seen in Figure 1, which means that a full denture constructed by Valplast polyamide material can withstand the usual masticatory forces successfully. Nevertheless, there will be a permanent deformation in extreme loads. In all cases, the recorded values of tested materials exceeded the 1567 minimum accepted force values (55 N) proposed by ISO.
Figure 1
Diagrammatical illustration of bending curve of Valplast after 3000 hydro- thermocycling.
Diagrammatical illustration of bending curve of Valplast after 3000 hydro- thermocycling.Many researchers (-) reported that water sorption and thermal changes (, -) decrease the flexural strength of denture base materials. In most of the studies, PMMA materials showed higher water sorption values than polyamide denture base materials. When PMMA is immersed in water solutions plasticizers and other soluble components may leach out over extended periods, while water or saliva is being absorbed. Absorbed water has a detrimental effect on the physical and mechanical properties of the resin denture base material (). Over a period of time, primarily because of the polar properties of the resin molecules, water can soften an acrylic resin by acting as a plasticizer of acrylates and reducing the strength of the material ().On the other hand, the low water sorption and solubility of polyamides is attributed to the amide groups (): the higher the amide group concentration, the greater the water sorption. It has been suggested that the amide group concentration of the polyamide denture base materials should be adjusted to a level as low as that of commonly used industrial materials such as nylon 6 or 66 ().The results of the present study are in agreement with the findings of Machado et al. (), who reported a significant decrease in the flexural strength of PMMA thermocycled for 5000 cycles between 5 and 55°C. The authors attributed this to the increase of temperature, which caused water molecules to diffuse more rapidly between the polymer chains, acting as plasticizers and allowing the chains to slip over each other more easily under load. In another study, Takahashi et al. () attributed the decrease of polyamide flexural strength to the ongoing thermal changes that cause continuous expansions and contractions and lead to static fatigue of the material. It may be concluded that in the present study the prolonged immersion in water during thermocycling led to significant reduction of the flexural strength of PMMA material, while the decrease of flexural strength of polyamide was mostly due to temperature changes.
Conclusions
Within the limitations of this in vitro study, the following conclusions may be drawn:Thermocycling adversely affected the flexural strength of polyamide and PMMA denture base materials. Thermocycling at 3000 cycles significantly reduced the flexural strength of PMMA denture base material. Thermocycling at 5000 cycles significantly reduced the flexural strength of both PMMA and polyamide denture base material.The polyamide denture base material exhibited significantly lower flexural strength values than PMMA for all the experimental conditions tested.Both materials presented flexural strength values within the ISO specification.