Evelinda Baerends1, Luuk J Oostveen2, Casper T Smit3, Marco Das4, Ioannis Sechopoulos1, Monique Brink1, Frank de Lange1, Mathias Prokop1.
Abstract
OBJECTIVES: To compare contrast-to-noise ratios (CNRs) and iodine discrimination thresholds on iodine maps derived from dual energy CT (DECT) and subtraction CT (SCT).Entities:
Keywords: Contrast Media; Iodine; Phantoms, Imaging; Subtraction technique; Tomography, X-Ray Computed
Mesh:
Substances:
Year: 2018 PMID: 29808430 PMCID: PMC6223839 DOI: 10.1007/s00330-018-5496-x
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1a Oval-shaped abdominal phantom with six tubular inserts used for this study b The contrast-detail inserts of the phantom. The inner diameter of the tubes from top to bottom is 15, 10, 6, 4 and 2 mm
Scan and reconstruction parameters for SCT and DECT
| SCT | DECT | |
|---|---|---|
| Scanner | Aquilion ONE ViSION, Canon Medical Systems | Somatom Flash, Siemens Healthineers |
| Tube voltage (kVp) | 100 | 100 / Sn-140 |
| Computed Tomography Dose Index volume (mGy) (three dose levels) | 7.80; 11.8; 14.8 | 8.02; 11.5; 15.0 |
| Effective Tube current time (mAs) (tube current x rotation time / pitch) (three dose levels) | 64; 96; 120 | [94 / 73]; [135 / 104]; |
| Rotation time (s) | 0.5 | 0.5 |
| Scan mode, collimation (mm) | Helical, 0.5 x 64 | Helical, 0.6 x 40 |
| Pitch | 0.625 | 0.6 |
| Field of view (mm) | 400 | 400 |
| Automatic tube current modulation | Off | Off |
| Reconstruction method | Iterative reconstruction (AIDR3d enhanced) | Iterative reconstruction (SAFIRE, setting 3) |
| Slice thickness, increment (mm) | 1 mm, increment 1 mm | 1 mm, increment 1 mm |
| Reconstruction kernela | FC08 | Q30F |
| Voxel size | 0.78x0.78x1 mm3 | 0.78x0.78x1 mm3 |
aAs used in clinical abdomen protocols as recommended by manufacturer
Fig. 2Creating four-panel composition images for the observer study by merging three square ROIs from three independently acquired iodine maps with tubes containing only water (blue) and one square ROI from an iodine map with a tube containing iodine contrast (green). This example shows a set of DECT data
Fig. 3Examples of the final image compositions for the observer study showing an example of (a) SCT and (b) DECT. The 4 mm tube with 10 mg/mL iodine concentration is positioned in the lower left corner in the SCT image (a) and in the lower right corner in the DECT image (b). Above the image compositions a marker for the size of the tube diameter was included
Fig. 4Bar graph shows contrast-to-noise (CNR) values of the 4 mm tube diameter at the 12 mGy dose level for all iodine concentrations (green = SCT, purple = DECT). Error bars represent one standard deviation
Fig. 5Graph shows adjusted means of CNR resulting from the univariate analysis, evaluated at the mean iodine concentration for (A) technique (p < .001), (B) diameter (p = .001) and (C) exposure level (p < .49). Error bars represent 95% confidence intervals
Results from the forced choice observer study: Minimal required iodine concentrations (Cr) for discriminating iodine from water on iodine maps derived from SCT and DECT
| CTDIvol (mGy) | Diameter (mm) | Minimal required iodine concentration (mg/mL) | Adjusted | |
|---|---|---|---|---|
| SCT | DECT | |||
| 8.0 | 2 | 3.0 ± 0.71 | 6.9 ± 0.72 | <.001* |
| 4 | 1.2 ± 0.084 | 4.0 ± 1.2 | <.001* | |
| 6 | 0.60 ± 0.036 | 1.8 ± 0.30 | <.001* | |
| 10 | 0.58 ± 0.023 | 0.78 ± 0.085 | <.001* | |
| 15 | <0.5 † | 0.54 ± 0.12 | n.a. | |
| 12 | 2 | 4.4 ± 0.53 | 5.5 ± 1.0 | .020* |
| 4 | 0.95 ± 0.039 | 3.1 ± 0.80 | <.001* | |
| 6 | 0.50 ± 0.089 | 1.1 ± 0.12 | <.001* | |
| 10 | <0.5a | <0.5a | n.a. | |
| 15 | <0.5a | <0.5a | n.a. | |
| 15 | 2 | 3.8 ± 0.89 | 7.4 ± 1.9 | <.001* |
| 4 | 1.3 ± 0.13 | 1.8 ± 0.55 | .07 | |
| 6 | <0.5 † | 0.64 ± 0.040 | n.a. | |
| 10 | <0.5 † | <0.5 † | n.a. | |
| 15 | <0.5 † | <0.5 † | n.a. | |
Concentrations are presented as mean ± standard deviation
* Significant difference (adjusted p < .05) a Calculated Cr is lower than the lowest presented contrast concentration of 0.5 mg/mL.
Fig. 6Bar graph shows minimal iodine concentrations required (Cr) to discriminate iodine from water by human observers for the 12 mGy exposure level (green = SCT, purple = DECT). The horizontal axis shows the tube diameter (mm). Error bars represent one standard deviation, * indicates statistical significance p <.05