High accuracy dosimetry with small pieces of Gafchromic films.

Ryszard Dąbrowski1, Izabela Drozdyk1, Paweł Kukołowicz1.   

Abstract

AIM: To investigate the influence of several factors on the accuracy of dose measurements and feasibility of application of small Gafchromic detectors for postal audit.
BACKGROUND: Our experience showed that precision of dose measurements with small pieces of Gafchromic films may be significantly improved.
MATERIALS AND METHODS: Gafchromic films with dimensions of 1 × 1, 2 × 2 and 3 × 3 cm2 were exposed to 6 MV X-rays at dose levels of 50 cGy-210 cGy. The single- and multichannel methods (MM) were used for dose measurements. Detectors were scanned with an Epson V750PRO flatbed colour scanner. For 1 × 1 and larger detector sizes, separate calibration curves were established. The influence of the following factors was investigated: the heterogeneity of Gafchromic detectors group for single- and MM, ambient thermal detector conditions, the dose delivered on the measurement accuracy, application of two separate calibration curves for the smallest and larger pieces of films.
RESULTS: The MM improves significantly the precision of dose measurement. The uncertainty attributed to detector active layer differences and scanner instabilities was about 1 cGy (1 StDev) regardless of dose and detector size. The ambient temperature of the environment in which films were stored after irradiation influenced the dose reading. Significant difference of transmission for detectors sized 1 × 1 and 2 × 2cm2 was observed. The maximal difference between applied dose and dose reading performed was 1.1%.
CONCLUSIONS: The MM with a scaling protocol leads to a very high precision of dose measurements. The ambient thermal detector environment causes significant changes of measured signal. The detector size has relevant impact on dose reading.

Entities:  

Keywords:  Gafchromic films; High precision measurements; Multichannel method; Postal dosimetry audit

Year:  2018        PMID: 29681774      PMCID: PMC5908271          DOI: 10.1016/j.rpor.2018.01.001

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  10 in total

1.  Tolerances for the accuracy of photon beam dose calculations of treatment planning systems.

Authors:  J Venselaar; H Welleweerd; B Mijnheer
Journal:  Radiother Oncol       Date:  2001-08       Impact factor: 6.280

2.  The IAEA/WHO TLD postal programme for radiotherapy hospitals.

Authors:  J Izewska; P Andreo
Journal:  Radiother Oncol       Date:  2000-01       Impact factor: 6.280

3.  Enhanced dosimetry procedures and assessment for EBT2 radiochromic film.

Authors:  Rulon R Mayer; Fuhwa Ma; Yu Chen; Rachel I Miller; Arnaud Belard; James McDonough; John J O'Connell
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

4.  Precise radiochromic film dosimetry using a flat-bed document scanner.

Authors:  Slobodan Devic; Jan Seuntjens; Edwin Sham; Ervin B Podgorsak; C Ross Schmidtlein; Assen S Kirov; Christopher G Soares
Journal:  Med Phys       Date:  2005-07       Impact factor: 4.071

5.  High sensitivity radiochromic film dosimetry using an optical common-mode rejection and a reflective-mode flatbed color scanner.

Authors:  Hiroko Ohuchi
Journal:  Med Phys       Date:  2007-11       Impact factor: 4.071

6.  Radiochromic film dosimetry with flatbed scanners: a fast and accurate method for dose calibration and uniformity correction with single film exposure.

Authors:  L Menegotti; A Delana; A Martignano
Journal:  Med Phys       Date:  2008-07       Impact factor: 4.071

7.  Accurate dosimetry with GafChromic EBT film of a 6 MV photon beam in water: what level is achievable?

Authors:  L J van Battum; D Hoffmans; H Piersma; S Heukelom
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

8.  Multichannel film dosimetry with nonuniformity correction.

Authors:  Andre Micke; David F Lewis; Xiang Yu
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

9.  Response to "Comment on 'A method for multichannel dosimetry with EBT3 radiochromic films' " [Med. Phys. 41, 062101 (10pp.) (2014)].

Authors:  Jose Fernando Pérez Azorín; Luis Isaac Ramos García; Josep M Martí-Climent
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

10.  An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan.

Authors:  David Lewis; Andre Micke; Xiang Yu; Maria F Chan
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.506

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.