INTRODUCTION: A real world clinical study was designed and conducted to evaluate the performance of a novel point-of-care device for determination of glycated haemoglobin A1c (HbA1c), A1C EZ 2.0, in daily clinical practice.Entities:
Keywords: diabetes mellitus; evaluation studies; glycated haemoglobin A1c; point-of-care testing
Mesh:
Substances:
Year: 2018 PMID: 29666558 PMCID: PMC5898960 DOI: 10.11613/BM.2018.020705
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Figure 1Frequency distribution of the HbA1c values measured by the A1C EZ 2.0 and three secondary reference methods based on HPLC methodology including Bio-Rad Variant II Turbo, Tosoh HLC-723 G8 and Premier Hb9210. Distribution of mean SRM is also included in the figure. In total, 635 samples were analyzed by each instrument. The population percentage (y-axis) is plotted against HbA1c values (x-axis).
HbA1c results for all the devices investigated
| HbA1c range (%) | 4.4 - 13.0 | 4.4 - 13.2 | 4.4 - 13.3 | 4.7 - 14.3 | 4.5 - 13.6 |
| Median | 6.9 | 6.7 | 6.8 | 6.8 | 6.7 |
| KS test | P < 0.001 | P < 0.001 | P < 0.001 | P < 0.001 | P < 0.001 |
| Skewness coefficients | 1.062 | 1.190 | 1.199 | 1.282 | 1.226 |
| Kurtosis coefficients | 0.678 | 1.312 | 1.348 | 1.539 | 1.388 |
| KS test - Kolmogorov-Smirnov test for normal distribution. Skewness coefficients and Kurtosis coefficients were used for normal distribution determination. SRM - secondary reference method. | |||||
Regression analysis of A1C EZ 2.0 results versus three secondary reference methods and the mean SRM
| All subjects | y = 0.03 (- 0.14 to 0.20) + 1.02 (1.00 to 1.05) x | 0.17 | 2.5 |
| SGI (venous group) | y = 0.30 (- 0.12 to 0.30) + 1.00 (1.00 to 1.06) x | 0.24 | 3.4 |
| SGII (capillary group) | y = - 0.36 (- 0.65 to 0.10) + 1.06 (1.00 to 1.10) x | 0.06 | 0.7 |
| SGIII (for venous) | y = - 0.10 (- 0.39 to 0.30) + 1.06 (1.00 to 1.10) x | 0.24 | 3.8 |
| SGIII (for capillary) | y = 0.20 (- 0.33 to 0.20) + 1.00 (1.00 to 1.09) x | 0.16 | 2.7 |
| All subjects | y = - 0.12 (- 0.27 to 0.00) + 1.02 (1.00 to 1.05) x | 0.04 | 0.6 |
| SGI (venous group) | y = 0.10 (0.04 to 0.10) + 1.00 (1.00 to 1.02) x | 0.09 | 1.3 |
| SGII (capillary group) | y = - 0.46 (- 0.76 to 0.00) + 1.06 (1.00 to 1.11) x | - 0.04 | - 0.6 |
| SGIII (for venous) | y = - 0.32 (0.59 to 0.10) + 1.06 (1.00 to 1.10) x | 0.08 | 1.1 |
| SGIII (for capillary) | y = - 0.20 (0.52 to 0.00) + 1.04 (1.00 to 1.09) x | 0.05 | 0.8 |
| All subjects | y = 0.00 (- 0.00 to 0.00) + 1.00 (1.00 to 1.00) x | 0.02 | 0.4 |
| SGI (venous group) | y = 0.22 (0.00 to 0.43) + 0.97 (0.94 to 1.00) x | - 0.05 | - 0.4 |
| SGII (capillary group) | y = 0.00 (- 0.25 to 0.00) + 1.00 (1.00 to 1.04)x | 0.00 | 0.2 |
| SGIII (for venous) | y = - 0.29 (- 0.54 to - 0.10) + 1.06 (1.00 to 1.10) x | 0.11 | 1.6 |
| SGIII (for capillary) | y = - 0.37 (- 0.64 to - 0.11) + 1.07 (1.02 to 1.11) x | 0.07 | 1.0 |
| All subjects | y = 0.10 (- 0.17 to 0.10) + 1.00 (1.00 to 1.04) x | 0.08 | 1.1 |
| SGI (venous group) | y = 0.10 (0.10 to 0.10) + 1.00 (1.00 to 1.00) x | 0.09 | 1.4 |
| SGII (capillary group) | y = - 0.30 (- 0.56 to 0.00) + 1.05 (1.00 to 1.08) x | 0.01 | 0.1 |
| SGIII (for venous) | y = - 0.22 (- 0.49 to 0.10)+ 1.06 (1.00 to 1.10) x | 0.14 | 2.1 |
| SGIII (for capillary) | y = - 0.23 (- 0.54 to 0.10) + 1.05 (1.00 to 1.10) x | 0.09 | 1.4 |
Figure 2Scatter diagram and linear regression analysis of A1C EZ 2.0 results versus mean SRM. The regression line (solid line), 95% CIs (dashed lines) and the identity line (y = x, dotted line) are displayed. SRM- secondary reference method.
Figure 3Bland-Altman relative difference plot shows the differences between the paired A1C EZ 2.0 results and the mean SRM. Horizontal lines are drawn at the mean difference% (solid line) and at the limits of agreement (dashed lines). Mean relative difference of A1C EZ 2.0 results versus the mean SRM is 1.1%, with the limits of agreement from -6.9% to 9.2%. SRM- secondary reference method.
Coefficients of variation calculated according to CLSI EP5-A2
| Within-run | 1.9% | 1.8% |
| Between-run | 0.0% | 0.6% |
| Between-day | 1.4% | 0.0% |
| Total | 2.3% | 1.9% |
| CV – coefficient of variation. | ||
Diagnostic sensitivity and specificity of A1C EZ 2.0 for diagnosing of diabetes
| 6.5% | Sensitivity (%) | 96 | 95 | 96 | 96 |
| Specificity (%) | 90 | 96 | 96 | 93 | |
| 6.3% | Sensitivity (%) | 97 | 95 | 96 | 97 |
| Specificity (%) | 87 | 92 | 89 | 89 | |
| SRM- secondary reference method. | |||||