[Capillary HbA1c determination on type 2 diabetes patients in a primary health centre].

María Teresa Carrera Font1, María Claustre Solé Brichs, María Clara Sala Álvarez, Jose María Navarro Olivella, Josefina Servent Turó, María Pilar Felipe Fernández.   

Abstract

OBJECTIVE: To determine the reliability and practicability of the point-of-care- test (POCT) analyzer, Afinion, for capillary HbA1c testing. To assess the benefits of its implementation on the intra-annual follow up of type 2 diabetic patients.
DESIGN: Descriptive cross-sectional study. Analytical validation of the Afinion reader.
SETTING: Primary Health Care (CAP Carmel and Bon Pastor Clinic Laboratory). PARTICIPANTS: A total of 94 type 2 diabetic patients selected according to their previous HbA1c value.
METHODS: We performed one capillary puncture and one venous extraction on each visit. The capillary sample was assessed in real time on the Afinion in the Primary Health Care Centre and the venous sample was sent to Bon Pastor Clinic Laboratory for assessment on an Afinion analyzer and by a high performance liquid chromatrography (HPLC) reference method. Practicability was assesses by both by the operators of the Afinion and the patients using an 11 question questionnaire. The efficiency in terms of process timings was also evaluated.
RESULTS: Intra-serial coefficient of variation (CV) was lower than 1% and inter-serial lower than 3%. The regression analysis showed: Afinion capillary sample=0.95 Afinion venous+0.21. No systematic or proportional error was detected in the 95% confidence interval (95% CI). The comparison between venous HPLC and Afinion showed: Afinion capillary sample=0.80 HPLC+1.14. A statistically significant difference was shown for these values at the 95% CI. Practicability was valued by users from 7 to 9.2 (professionals) and from 7.7 to 9.2 (patients). Implementation of the Afinion capillary method for intra-annual testing in follow up of diabetic patients could result in the saving of 600-900 professional hours/year.
CONCLUSIONS: Afinion seems to be a good choice for the intra-annual determination of HbA1c when compared to the traditional process due to its accessibility, practicability and efficiency. Professionals should know the limitations of the POCT method in order to consider the validity of the results.
Copyright © 2010 Elsevier España, S.L. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21411189      PMCID: PMC7024957          DOI: 10.1016/j.aprim.2010.09.017

Source DB:  PubMed          Journal:  Aten Primaria        ISSN: 0212-6567            Impact factor:   1.137


  21 in total

1.  Impact of active versus usual algorithmic titration of basal insulin and point-of-care versus laboratory measurement of HbA1c on glycemic control in patients with type 2 diabetes: the Glycemic Optimization with Algorithms and Labs at Point of Care (GOAL A1C) trial.

Authors:  Laurence Kennedy; William H Herman; Poul Strange; Anthony Harris
Journal:  Diabetes Care       Date:  2006-01       Impact factor: 19.112

Review 2.  [Evolution of prevalence of type 2 diabetes in adult Spanish population].

Authors:  Sergio Valdés; Gemma Rojo-Martínez; Federico Soriguer
Journal:  Med Clin (Barc)       Date:  2007-09-15       Impact factor: 1.725

3.  Haemoglobin A1c: evaluation of a new HbA1c point-of-care analyser Bio-Rad in2it in comparison with the DCA 2000 and central laboratory analysers.

Authors:  Chin-Pin Yeo; Carol Hui-Chen Tan; Edward Jacob
Journal:  Ann Clin Biochem       Date:  2009-06-01       Impact factor: 2.057

4.  Relationship between glycated haemoglobin and microvascular complications: is there a natural cut-off point for the diagnosis of diabetes?

Authors:  C Sabanayagam; G Liew; E S Tai; A Shankar; S C Lim; T Subramaniam; T Y Wong
Journal:  Diabetologia       Date:  2009-04-22       Impact factor: 10.122

5.  HbA1c standardisation destination--global IFCC Standardisation. How, why, where and when--a tortuous pathway from kit manufacturers, via inter-laboratory lyophilized and whole blood comparisons to designated national comparison schemes.

Authors:  Ian Goodall
Journal:  Clin Biochem Rev       Date:  2005-02

6.  Rapid A1c availability improves clinical decision-making in an urban primary care clinic.

Authors:  Christopher D Miller; Catherine S Barnes; Lawrence S Phillips; David C Ziemer; Daniel L Gallina; Curtiss B Cook; Sandra D Maryman; Imad M El-Kebbi
Journal:  Diabetes Care       Date:  2003-04       Impact factor: 19.112

7.  Assessment of the practicability and analytical performance of a point-of-care affinity chromatography haemoglobin A1c analyser for use in the non-laboratory setting.

Authors:  Mark Shephard; Malcolm Whiting
Journal:  Ann Clin Biochem       Date:  2006-11       Impact factor: 2.057

Review 8.  [Out-patient determination of glycosylated haemoglobin in the monitoring and control of diabetes mellitus: systematic review of the literature].

Authors:  Jesús Ruiz-Aragón; Román Villegas Portero; Sandra Flores Moreno
Journal:  Aten Primaria       Date:  2008-02       Impact factor: 1.137

9.  Diagnosis and classification of diabetes mellitus.

Authors: 
Journal:  Diabetes Care       Date:  2010-01       Impact factor: 19.112

10.  Longitudinal association of glucose metabolism with retinopathy: results from the Australian Diabetes Obesity and Lifestyle (AusDiab) study.

Authors:  Robyn J Tapp; Gabriella Tikellis; Tien Y Wong; C Alex Harper; Paul Z Zimmet; Jonathan E Shaw
Journal:  Diabetes Care       Date:  2008-04-14       Impact factor: 19.112

View more
  1 in total

1.  Verification of a novel point-of-care HbA1c device in real world clinical practice by comparison to three high performance liquid chromatography instruments.

Authors:  Yufei Wang; Wei Peng; Junling Tang; Lu Dong; Chengchen Gu; Xin Zhang; Jian Zhou; Weiping Jia
Journal:  Biochem Med (Zagreb)       Date:  2018-04-15       Impact factor: 2.313

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.