Emilia Studzińska1, Maria Anna Staniszewska1.
Abstract
BACKGROUND: To evaluate the impact of the Clarity IQ technology on reducing radiation risk in patients undergoing cardiac interventional radiology (IR) procedures. MATERIAL/Entities:
Keywords: Interventional; Ionizing; Radiation; Radiology; Thermoluminescent Dosimetry
Year: 2017 PMID: 29662591 PMCID: PMC5894036 DOI: 10.12659/PJR.902350
Source DB: PubMed Journal: Pol J Radiol ISSN: 1733-134X
Duration and C-arm angulation for the particular steps of left atrial appendage closure using the investigated angiographic units (FD10 Allura Xper and FD10 Allura Clarity).
| Order of steps | Fluoroscopy/DSA | C-arm angle | Duration (sec) | Number of obtained images |
|---|---|---|---|---|
| 1 | Fluoroscopy | Rao 0° | 321 | – |
| 2 | DSA | Rao 30° | – | 30 |
| 3 | Fluoroscopy | Rao 30° | 79 | – |
| 4 | DSA | Rao 30° | – | 6 |
| 5 | Fluoroscopy | Rao 30° | 144 | – |
Mean values of doses absorbed by particular areas of the CIRS phantom during left atrial appendage closure using the investigated angiographic units (FD10 Allura Xper and FD10 Allura Clarity).
| Organ | Allura xper [mGy | Allura clarity [mGy] |
|---|---|---|
| Red bone marrow | 4.25 | 1.89 |
| Colon | 0.11 | 0.02 |
| Lungs | 8.44 | 4.33 |
| Stomach | 1.81 | 0.40 |
| Gonads | 0.00 | 0.00 |
| Bladder | 0.00 | 0.00 |
| Liver | 0.90 | 0.25 |
| Thyroid | 1.54 | 0.71 |
| Compact bone | 2.92 | 1.30 |
| Brain | 0.22 | 0.09 |
| Eye lens | 0.06 | 0.02 |
| Heart | 8.29 | 3.98 |
| Intestine | 0.12 | 0.02 |
| Kidneys | 1.29 | 0.18 |
| Pancreas | 0.52 | 0.12 |
| Spleen | 4.35 | 0.83 |
Figure 1Percentage dose reduction of organ-absorbed doses using the FD10 Allura Clarity system in comparison to the FD 10 Allura Xper, for organs exposed to radiation doses above 1 mGy.