| Literature DB >> 29662301 |
Naoki Tojo1, Mitsuya Otsuka1, Atsushi Hayashi1.
Abstract
PURPOSE: To compare the postoperative interventions and outcomes between conventional trabeculectomy and trabeculectomy with the Ex-PRESS® mini-glaucoma shunt device (Ex-Press).Entities:
Keywords: Ex-Press; glaucoma; laser suture lysis; postoperative management; trabeculectomy
Year: 2018 PMID: 29662301 PMCID: PMC5892968 DOI: 10.2147/OPTH.S160342
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
The characteristics of the 108 glaucoma patients who underwent a conventional trabeculectomy or trabeculectomy with an Ex-Press
| Characteristics | Conventional | Ex-Press | |
|---|---|---|---|
| No of patients | 39 | 69 | |
| Males/females | 24/15 | 31/38 | 0.28 |
| Age, years | 70.0±13.7 | 70.3±8.6 | 0.877 |
| Pre IOP, mmHg | 27.8±7.9 | 27.7±9.2 | 0.919 |
| No of medications | 3.8±0.6 | 4.1 ±1.1 | 0.102 |
| Follow-up, months | 37.8± 14.2 | 31.1 ± 13.1 | 0.0093 |
| CCT, μm | 528±45 | 535±38 | 0.352 |
| Scleral flap sutures | 4.3± 1.0 | 2.7± 1.0 | <0.001 |
| Triple surgery | 20.5% (8/39) | 26.1% (18/69) | 0.761 |
| Subtypes of glaucoma | |||
| POAG | 35.9% (14/39) | 37.7% (26/69) | 1.00 |
| PEX | 41.0% (16/39) | 60.8% (42/69) | 0.07 |
| SG | 20.5% (8/39) | 1.4% (1/69) | 0.0011 |
| PACG | 2.6% (1/39) | 0% (0/84) | 0.361 |
Note: Triple surgery: trabeculectomy and phacoemulsification + intraocular lens implantation simultaneously.
Abbreviations: CCT, central corneal thickness; Ex-Press, ex-press® mini-glaucoma shunt device; IOP, intraocular pressure; PACG, primary angle closure glaucoma; PEX, pseudo-exfoliation; POAG, primary open angle glaucoma; sg, secondary glaucoma.
The postoperative data (IOP and no. of medications) of the conventional and Ex-Press groupsa
| IOP, mmHg
| No of medications
| |||||
|---|---|---|---|---|---|---|
| Conventional | Ex-Press | Conventional | Ex-Press | |||
| Pre-IOP (no of eyes) | 27.8±7.9 (39) | 27.7±9.2 (69) | 0.919 | 3.8±0.6 (39) | 4.1±1.1 (69) | 0.102 |
| 1 year | 11.1±3.9 (34) | 11.5±3.7 (60) | 0.612 | 1.4±1.6 (34) | 1.6±1.7 (60) | 0.560 |
| 2 years | 11.4±3.1 (30) | 11.9±3.5 (34) | 0.519 | 1.5±1.6 (30) | 2.2±1.7 (34) | 0.129 |
| 3 years | 12.5±3.0 (18) | 11.5±3.3 (21) | 0.319 | 2.1±1.5 (18) | 2.3±1.7 (21) | 0.648 |
| 4 years | 13.0±3.4 (12) | 10.9±3.4 (8) | 0.178 | 2.3±1.5 (12) | 2.4±2.0 (8) | 0.876 |
| 5 years | 11.0±3.5 (5) | 8.5±0.7 (2) | 0.390 | 2.4±1.3 (5) | 2.0±2.8 (2) | 0.795 |
Note:
The results of the eyes that underwent additional glaucoma surgeries were excluded.
Abbreviations: Ex-Press, Ex-PRESS® mini-glaucoma shunt device; IOP, intraocular pressure.
Comparison of the results between Single and Triple surgeries
| Single surgery | Triple surgery | ||
|---|---|---|---|
| Preoperation | |||
| IOP | 27.6±8.4 (82) | 27.9±9.7 (26) | 0.886 |
| Medications | 4.0±1.0 | 4.0±0.8 | 0.871 |
| 1 year | |||
| IOP | 11.1±4.1 (72) | 12.3±2.4 (22) | 0.188 |
| Medications | 1.4±1.6 | 1.5±1.7 | 0.841 |
| 2 years | |||
| IOP | 11.4±3.4 (50) | 12.8±2.9 (14) | 0.163 |
| Medications | 1.9±1.7 | 1.6±1.7 | 0.564 |
| 3 years | |||
| IOP | 11.9±3.3 (29) | 12.2±2.9 (11) | 0.74 |
| Medications | 2.1±1.6 | 2.3±1.7 | 0.814 |
| 4 years | |||
| IOP | 12.7±5.5 (16) | 12.6±3.2 (5) | 0.974 |
| Medications | 2.1±1.7 | 3.0±1.7 | 0.297 |
| 5 years | |||
| IOP | 10.0±3.1 (7) | None | |
| Medications | 2.3±1.6 |
Abbreviation: IOP, intraocular pressure.
Figure 1(A) Kaplan–Meier survival curves for Criterion A (postoperative IOP ≤21 mmHg or ≥20% reduction from the baseline). (B) Kaplan–Meier survival curves for Criterion B (postoperative IOP ≤15 mmHg or $20% reduction from the baseline).
Note: There was no significant difference in the success rate between the conventional and Ex-Press groups in Criterion A (p=0.570) or Criterion B (p=0.991).
Abbreviations: Ex-Press, Ex-PRESS® mini-glaucoma shunt device; IOP, intraocular pressure.
Comparison of postoperative interventions between conventional trabeculectomy and trabeculectomy with Ex-Press
| Conventional | Ex-Press | ||
|---|---|---|---|
| Laser suture lysis executed (eyes) | 66.6% (26/39) | 84.1% (58/69) | 0.0368 |
| No of sutured threads (eyes) | 4.4±1.1 (26) | 2.7±1.1 (58) | <0.0001 |
| No of cut threads (eyes) | 2.7±1.4 (26) | 2.4±1.0 (58) | 0.297 |
| No of cut threads/No of sutured threads (eyes) | 61.2±26.3 (26) | 89.2±20.0 (58) | <0.0001 |
| No of days until cutting the first thread (eyes) | 8.8±10.6 (26) | 3.4±4.3 (58) | 0.0012 |
| No of days until cutting the second thread (eyes) | 11.9±10.9 (20) | 12.2±11.4 (53) | 0.904 |
| Needling executed (eyes) | 23.0% (9) | 15.9% (11) | 0.359 |
| No of days until needling | 32.5±15.8 | 29.2±18.9 | 0.691 |
| IOP before needling | 23.6±3.1 mmHg | 22.0±5.1 mmHg | 0.434 |
| IOP 1 month after needling (eyes) (rate of decrease) | 17.2±6.9 mmHg | 17.6±4.6 mmHg | 0.874 (0.292) |
| IOP 3 months after needling (eyes) (rate of decrease) | 13.1±3.0 mmHg | 12.0±2.9 mmHg | 0.467 (0.917) |
| IOP 6 months after needling (eyes) (rate of decrease) | 13.1±3.0 mmHg | 11.5±2.3 mmHg | 0.251 (0.872) |
Abbreviations: Ex-Press, Ex-PRESS® mini-glaucoma shunt device; IOP, intraocular pressure.
Comparison of complications between conventional trabeculectomy and trabeculectomy with Ex-Press
| Conventional | Ex-Press | ||
|---|---|---|---|
| Shallow anterior chamber | 7.7% (3/39) | 2.9% (2/69) | 0.255 |
| Hypotonic maculopathy | 5.1% (2/39) | 0% (0/69) | 0.0576 |
| Choroidal detachment | 33.3% (13/39) | 15.9% (11/69) | 0.0368 |
| Hyphema | 7.7% (3/39) | 1.4% (1/69) | 0.0989 |
| Corneal edema | 2.6% (1/39) | 1.4% (1/69) | 0.680 |
| Suprachoroidal hemorrhage | 2.6% (1/39) | 0% (0/69) | 0.181 |
| Endophthalmitis | 0% (0/39) | 0% (0/69) | – |
Note: Data presented as percentage (eyes).
Abbreviation: Ex-Press, Ex-PRESS® mini-glaucoma shunt device;