Biobeam-Multiplexed wave-optical simulations of light-sheet microscopy.

Martin Weigert1,2, Kaushikaram Subramanian1,2, Sebastian T Bundschuh1, Eugene W Myers1,2,3, Moritz Kreysing1,2.   

Abstract

Sample-induced image-degradation remains an intricate wave-optical problem in light-sheet microscopy. Here we present biobeam, an open-source software package that enables simulation of operational light-sheet microscopes by combining data from 105-106 multiplexed and GPU-accelerated point-spread-function calculations. The wave-optical nature of these simulations leads to the faithful reproduction of spatially varying aberrations, diffraction artifacts, geometric image distortions, adaptive optics, and emergent wave-optical phenomena, and renders image-formation in light-sheet microscopy computationally tractable.

Entities:  

Mesh:

Year:  2018        PMID: 29652879      PMCID: PMC5898703          DOI: 10.1371/journal.pcbi.1006079

Source DB:  PubMed          Journal:  PLoS Comput Biol        ISSN: 1553-734X            Impact factor:   4.475


This is a PLOS Computational Biology Software paper.

Introduction

Light-sheet fluorescence microscopy is a popular tool for the volumetric imaging of developing organisms [1-5]. As light-sheet microscopes continue to be developed for progressively bigger biological samples [6, 7], there is an increasing need for computers to process gigantic sets of imaging data and extract biologically relevant information [8]. With sample size, however, also light-scattering induced imaging artifacts become increasingly prevalent. During data acquisition as well as post-processing, these imaging artifacts are mostly dealt with on a purely phenomenological basis. A faithful forward model of the wave-optical imaging process, however, would i) enable rigorous benchmarks of deconvolution and segmentation strategies against ground-truth data [9], ii) serve as training platforms for machine learning approaches for image restoration and information extraction and iii) leverage the efficient use of adaptive optics [10, 11] to prevent sample-induced image degradation during the acquisition process. Predicting light-tissue interactions is particularly demanding when leaving the single scattering regime [12] or strictly diffusive transport [13]. And despite significant computational advances [14], generally applicable solutions [15] remain computationally costly, effectively prohibiting the simulation of image formation in microscopy. Even when constraining simulations to the biological relevant case of predominantly forward scattering tissue [10, 16], individual point spread function (PSF) calculations still require multiple seconds [17]. As aberrations deep inside tissues are unique for virtually each point in the sample [10, 12] a realistically large biological specimen, i.e. an embryo with a volume ∼ 100 μm3 would require 105–106 volumetric PSF calculations in order to faithfully mimic the wave-optical imaging process, which with current methods would take several weeks. As a consequence, attempts to simulate microscopic imaging have been limited to ray optics [18] or convolution with a constant PSF [19], approaches that do not reflect the wave-optical nature of light interaction with optically heterogeneous biological samples. Here, we report on biobeam, a software package that enables the first rigorous simulations of wave-optical image formation in light-sheet microscopes by i) a novel multiplexing scheme for PSF calculations and ii) efficient GPU parallelization.

Design and implementation

The pipeline underlying our software is based on the observation that the beam-propagation model (BPM) for fiber optics [20, 21] can also be used to mimic scattering biological cells [22]. To guarantee good accuracy of BPM beyond strictly paraxial wave propagation, we use the exact propagator together with a locally adapted expansion of refractive indices (see S1 Text, Notes 1 and 4 for details and validation against analytically tractable scattering models). To massively reduce the computational cost for ∼106 wave-optical PSF calculations, we introduce a novel multiplexing scheme. In this we exploit the fact that for typical imaging scenarios, a single camera image can be constructed from sets of 100s to 1000s of mutually independent, spatially varying, non-overlapping PSFs. Such sets can be calculated within single, highly multiplexed simulations (see S3 Fig and S1 Text, Note 8), which is similar to the operational principle of highly multiplexed confocal measurements of spinning disc microscopes [23, 24]. Our software is further accelerated ∼20 fold by the efficient use of GPU implementations for all low-level calculations. In this way, the propagation of an arbitrary light field such as the multiplexed set of ∼1000 PSFs through a typical volume of 10243 voxels takes less than 500ms on a single graphics card (cf. S1 Text, Note 5, S1 Table and S4 Video), corresponding to a 20.000 fold acceleration compared to the sequential calculation of PSFs. Implementation was done as an open source Python software package. All computationally heavy parts are lifted to the GPU via OpenCL [25], thereby keeping all the advantages of Python as a dynamically typed high level language that is vastly used in the scientific community without compromising on performance. Apart from its technical focus on speed, biobeam is specifically designed to make wave-optical experiments in-silico as easy as possible, e.g. by providing a simple API that makes it easy to apply different input fields as well as PSF/aberration calculations via diffraction-limited point source propagation through tissue (see S1 Text, Note 3 for example listings).

Results

Light-sheet microscopy

We first demonstrate the power of our software package by presenting the first wave-optical simulation of the volumetric image-formation process in a light-sheet microscope [5, 9, 18]. To start, a cylindrical sheet of light is propagated at a specific axial position through a fluorescently labelled embryo model with refractive indices of cytoplasm, nuclei and organelles as previously reported [12] (Fig 1a and S1 & S7 Videos). From this, one obtains the fluorescence excitation at every point in the volume for every axial position of the incident light field. Next, for each focal plane a full set of detection PSFs is obtained by propagating light from multiplexed, diffraction-limited point-sources orthogonally through an idealized, refocusing lens towards the camera (Fig 1c, upper left, S2 Video). This way, we obtain a quasi-continuum of spatially-dependent, volumetric PSFs with position-dependent aberrations that stem from distortions and scattering in both the illumination and the detection paths (Fig 1c left, see also S4 Fig for calculated aberration maps). Convolving the exhaustively sampled sets of spatially-varying PSFs with the fluorescent object finally yields the wave-optical image as seen by the camera (Fig 1c, right). These simulations are particularly demanding due to the large grid size (i.e. 1024 × 2048 × 1024 voxels for Fig 1) and sampling density requirements (typically 0.5–5μm), here resulting in a total of 106 PSFs that were calculated in well under 10 minutes on a single graphics card (see also S1 and S2 Videos). In contrast, the non-multiplexed calculations would have required more than two months on a single CPU. As a result of this computational pipeline, biobeam generates faithfully calculated 3D microscopy data sets that account for both refraction and diffraction based imaging artifacts. These include image blur and contrast loss, spatially varying PSFs, speckle artifacts, image granularity, as well as sample-induced geometric distortions such as lensing, image displacements and split-screen type double images (cf. Fig 2a and 2b, S3 Video and S1 Fig for benchmarks against analytically tractable models).
Fig 1

Rigorous wave-optical simulation of image formation process in light-sheet microscopy.

(a) Synthetic tissue phantom of a multicellular organism (100 × 200 × 100μm) comprising a complex refractive index distribution (left, n = 1.33–1.38) and a fluorophore distribution of interest (right). (b) Wave optical simulation of the illuminating light sheet and resulting excitation distribution within the sample at a given z position. (c) Partially coherent simulation of the detection path by multiplexed calculation of all independent point spread functions (left) and the resulting simulated camera image combining illumination and fluorescence path of light through the scattering sample. (d) Alternative light-sheet modalities (see also S1 and S5 Videos and main text for details).

Fig 2

Optical capabilities of the biobeam image-formation pipeline.

(a) A test chart at the mid-section of an optically heterogeneous embryo-model (n = 1.35–1.39 diameter 140μm) is illuminated by cylindrical light sheet (NA = 0.15), and imaged from an orthogonal position (NA = 0.6). (b) Details of these wave-optically calculated images reveal i) spatially varying image blur, contrast loss and absorption induced by the heterogeneity of the sample, ii) diffraction artifacts from the light-sheet-typical coherent illumination, and iii) geometric image distortions such as lensing, split-screen type image distortions, and object displacements lensing. (c) Biobeam is further capable of adaptive optics simulations by which reversal of guide star emitted light fields yields perfect foci in scattering tissues. (d) Adaptive optics simulations faithfully reproduces the shift-shift memory effect, an emergent wave-optical phenomenon, here at 4 mean-free-paths deep inside the tissue.

Rigorous wave-optical simulation of image formation process in light-sheet microscopy.

(a) Synthetic tissue phantom of a multicellular organism (100 × 200 × 100μm) comprising a complex refractive index distribution (left, n = 1.33–1.38) and a fluorophore distribution of interest (right). (b) Wave optical simulation of the illuminating light sheet and resulting excitation distribution within the sample at a given z position. (c) Partially coherent simulation of the detection path by multiplexed calculation of all independent point spread functions (left) and the resulting simulated camera image combining illumination and fluorescence path of light through the scattering sample. (d) Alternative light-sheet modalities (see also S1 and S5 Videos and main text for details).

Optical capabilities of the biobeam image-formation pipeline.

(a) A test chart at the mid-section of an optically heterogeneous embryo-model (n = 1.35–1.39 diameter 140μm) is illuminated by cylindrical light sheet (NA = 0.15), and imaged from an orthogonal position (NA = 0.6). (b) Details of these wave-optically calculated images reveal i) spatially varying image blur, contrast loss and absorption induced by the heterogeneity of the sample, ii) diffraction artifacts from the light-sheet-typical coherent illumination, and iii) geometric image distortions such as lensing, split-screen type image distortions, and object displacements lensing. (c) Biobeam is further capable of adaptive optics simulations by which reversal of guide star emitted light fields yields perfect foci in scattering tissues. (d) Adaptive optics simulations faithfully reproduces the shift-shift memory effect, an emergent wave-optical phenomenon, here at 4 mean-free-paths deep inside the tissue. We additionally validated our approach by comparing simulation results for several experimental situations, such as the diffraction pattern behind an edge knife (S5 Fig), the scattering of a light-sheet at agarose embedded micro-spheres (S6 Fig) and the image distortions obtained when projecting test images through a refracting sphere (S7 and S8 Figs). Beyond the theoretical validation of the core propagation model, these successful comparisons of simulations against experiments further demonstrate that our high-level implementation works correctly in the biological relevant case of low refractive index contrasts. Being a particularly flexible software package, biobeam provides further pre-implemented illumination modalities, including Bessel lattices (Fig 1d, right, see S2 Fig, S5 Video), and allows for practical extraction of sample-induced aberrations as spatially resolved Zernike maps (see S4 Fig and S1 Text, Note 9).

Adaptive optics simulations

Next, we demonstrate biobeam’s capability to accurately simulate wave-optical effects relevant to adaptive optics (AO) imaging [10]. It is well-established and exploited in imaging that perfect imaging foci can be created behind strongly scattering screens [26] and within biological samples by appropriately shaping the wavefront before entry into the scattering medium. We recapitulated this finding by explicit simulations of light propagating through a spatially extended synthetic tissue sample represented by a Perlin-noise refractive index distribution (n = 1.36 ± 0.03). As expected, we find also in simulations that conjugation of the wavefront at the surface of the sample allows one to generate diffraction-limited foci inside these scattering tissues (Fig 2c, S6 Video and S9 Fig). Furthermore, we show that biobeam is capable of reproducing the shift-shift memory effect, an emergent wave-optical phenomenon responsible for the significant robustness of adaptive imaging against lateral focus displacements. According to this wave-optical phenomenon, the radius of the iso-planatic patch (distance over which an AO correction pattern works) is determined by the statistics of a tissue generated speckle field inside this medium. Based on a total of 22500 PSF calculations, we show that our simulations faithfully capture this emergent wave-optical effect. In agreement with recent analytical arguments and their empirical confirmation [10], we show by our computational microscopy experiments that the average persistence of the laterally shifted focus is precisely limited by the autocorrelation length of a speckle pattern that would result from an incident plane wave. This phenomenon is accurately reproduced even four mean-free-path lengths inside the tissue (Fig 2d). While currently using a variant of the beam propagation method, our PSF multiplexing scheme is in principle also compatible with other low-level field stepping algorithms, including those that iteratively account for multiple back-reflections [14, 15], when even higher precision or penetration depths are required.

Discussion

We summarize that biobeam enables faithful whole-tissue wave-optical simulations of light-sheet microscopes due to i) a novel multiplexing scheme of PSF calculations and ii) efficient GPU parallelization. Biobeam renders the biological imaging process computationally tractable, thus providing the link between wave-optically recorded image material, and the ground truth object. Given the modular nature of our software package, these simulations are easy to implement and can be flexibly adapted to custom imaging scenarios and microscopes. Beyond the reproduction and identification of commonly occurring imaging artifacts (see Supplement for comparison with experimental data), we demonstrated that biobeam is compatible with imaging scenarios in which sample-induced PSFs degradation is overcome by the use of adaptive optics. Furthermore, we showcase the optical capabilities and accuracy of our software by explicitly demonstrating that emergent wave-optical phenomena such as the shift-shift memory effect are quantitatively reproduced deep inside tissues. While we chose a variant of BPM as a low-level field stepping routine, the here presented strategy of multiplexing PSF calculation is more general in nature, and may also be used in combination with other light propagation algorithms, e.g. in scenarios where higher accuracy at large angles and/or more isotropic scattering is of relevance [14]. We conclude that biobeam is a flexible and particularly powerful platform to systematically study wave-optical image-formation by microscopes in scattering biological tissues.

Availability and future directions

Prospectively, we see biobeam helping to improve microscope design, enhancing deconvolution and segmentation strategies by providing realistic imaging data-sets along with ground truth data, and paving the way for a new generation of smart, adaptive microscopes that learn to treat the sample as a part of the optical path. Of great help in this will be the rapidly increasing knowledge of stereotypical refractive index distributions in embryos, tissues, cells, subcellular compartments and their constituents as derived from tomographic phase microscopy [12, 27–29] and complemented by electron microscopy morphological data [30]. Beyond the simulation of light-sheet microscopes, our software can also be used to simulate other imaging modalities such as wide-field, laser-scanning confocal, and light-field [31] microscopes, as well as for novel micro-lens concepts [32] and the emerging field of soft photonics [33]. Especially, we emphasize the potential to improve the understanding of physiological image-formation inside the eye [34], taking into account the optics of the lens [35] as well as the retina [36-39]. Biobeam is available as open-source (BSD-3 license) python package at https://maweigert.github.io/biobeam. Datasets can be found at https://publications.mpi-cbg.de/6874-data/.

25 pages description of the numerical method, validation, benchmarks, details of the numerical experiments, and example listings.

(PDF) Click here for additional data file.

Wave-optical simulation of the image-formation process in light-sheet microscopy.

The tissue model represents a multicellular (760 nuclei) organism of size (100μm, 200μm, 100μm) in an aqueous medium with n = 1.33. The refractive index distribution is in the range n ∈ (1.35, 1.42) comprising reference values for cell nuclei, eggshell and the cytoplasm [12]. Weak absorption is homogenously present, but could also be localized (e.g. a spherical absorbing compartment in the center). The simulations of both the illumination and detection processes were carried out on a computational grid of (1024, 2048, 1024) voxels with a spacing of 100nm along each dimension. The illumination field is a cylindrical light sheet with NA = 0.1 focused laterally at the center and the detection system was assumed to have NA = 0.6. For generating the final stack both illumination and detection fields were simulated at 200 different axial positions. The deterioration of both resolution and intensity at regions where photons along either the illumination or detection path had to travel through large inhomogeneities can clearly be seen. (MP4) Click here for additional data file.

Illustration of a single PSF calculation inside the tissue via the propagation of analytically defined diffraction-limited input fields.

Due to the linearity of wave-optics, these PSF calculations can be highly multiplexed, as firstly exploited by biobeam. (MP4) Click here for additional data file.

A biobeam generated video illustrating rigorous wave-optical mimicry of a wide-field microscope.

The imaging of a 100 μm2 test chart is simulated while a refractive sphere is continuously introduced into the microscope’s optical path. biobeam generated the underlying wave-optical simulations in 30 seconds. (MP4) Click here for additional data file.

Screencast of an interactive command line session demonstrating biobeam’s capabilities and speed.

All calculations happen in real time. (MP4) Click here for additional data file.

Video showing the predefined illumination modes and simulated light sheets being scanned through a biologically plausible tissue model.

Both coherent (cylindrical lens SPIM) illumination and partially-incoherent illumination modes (time scanned Gaussian/Bessel beams) are simulated. (MP4) Click here for additional data file.

Showing the simulation of a aberration pre-compensated wavefront focusing deep into tissue and the shift-shift memory effect.

(MP4) Click here for additional data file.

Example of a plane-by-plane illumination of a tissue model mimicking an embryo.

(MP4) Click here for additional data file.

Validation of biobeam with analytical solutions.

a) Plane wave scattered by three solid spheres (λ = 500nm, r = 2-2.5μm, refractive index contrast m = 1.05), b) Comparison of analytical solution (Mie calculus) versus biobeam simulation. c) Error percentage of near field distribution as a function of single sphere radius r (Δn = 0.05) and refractive index contrast Δn (r = 2.5μm). d) Left: Phase function of analytically tractable coated spheres as cell models (m = 1.02/1.04, r = 5μm/4μm) shows high accuracy up to approximately 0.5 radians. Right: size dependent scattering efficiency of the same sphere architecture and its inverse. (PDF) Click here for additional data file.

Propagation of different predefined input fields through a tissue model of size (100μm, 100μm, 100μm) and grid dimension (10243).

The respective pupil function is shown in the upper row. (PDF) Click here for additional data file.

Detection aberration and PSF calculation.

Propagating a diffraction limited input field through parts of the sample and refocusing by an idealized optical system gives the focus field as seen by the detector. If the refocus spots are separated for different starting points, the propagation of a complete grid can be carried out in a highly multiplexed manner, accelerating the process for typical microscopy simulations by a factor 100–1000. (PDF) Click here for additional data file.

Calculating the aberrations of the detection PSF for a given z plane within a synthetic tissue model.

The model’s physical size is (200μm, 100μm, 100μm) and the dimension of the computational grid is (1024, 512, 512). The detection wavelength is λ = 522nm, the numerical aperture is NA = 0.5 and the aqueous immersion medium has refractive index n0 = 1.33. The refractive index distribution of the tissue model mimics an eggshell, cell nuclei and granular random fluctuations within the biological plausible range of n ∈ (1.35, 1.43). (PDF) Click here for additional data file.

Diffraction around a knife edge.

a) Experimental setup: Light is focused with an incoherent light source (M470L3 Thorlabs, λ = 470nm) such that an almost plane wave (NA = 0.001) illuminated the knife edge. The diffracting light was imaged below at different depths from the edge. b) The simulation was done on a computational cell of size (1024 × 256 × 1830) with voxel size Δx = 0.29μm. We simulated the diffraction in the case of a single plane wave (coherent, top) and the incoherent superposition of 100 random incident plane waves of different small incident angle (corresponding to NA = 0.001, bottom). c) The experimental acquired intensity. Scalebar is 12μm in both axial and lateral direction (depicted with axial/lateral aspect ratio of 8, due to space constraints). d) Intensity plot at a given axial position (dashed line) for simulation, experiment and the intensity calculated via Fresnel-integral (Theory). (PDF) Click here for additional data file.

Experimental validation on a commercial light-sheet microscope.

a) Poly-methylmethacrylate (PMMA) micro-particles with a diameter of 20μm and refractive index of n = 1.495 were embedded in an OptiPrep/agarose block (n ≈ 1.43) labelled with Alexa Fluor 488. A stationary illuminating light-sheet with a waist of 1.7μm and a lateral extension of ≈100μm was generated with a LZ1 (Zeiss) light-sheet microscope, incident on the agarose embedded spheres. Stacks were acquired at a step size of 0.414μm. b) Simulation results of the intensity distribution behind the sphere at a plane incident to the sphere center. c) Experimental intensity image. Scalebar is 20μm in both cases. Dashed lines indicate regions with specific diffraction patterns that the simulation correctly reconstitutes. (PDF) Click here for additional data file.

Custom micro-projection setup built in our lab, controlled with custom LabView programs.

The setup allows for patterns to be micro projected onto a sample with predefined illumination-source, size, magnification and NA of influx optics. The efflux optics allows for the collection of the light and recording on the camera. (PDF) Click here for additional data file.

Experimental micro-projection of a test-chart through a glass sphere and comparison with simulation.

Experimental setup as in S7 Fig. The negative USAF (R1DS1N, Thorlabs) test-chart was illuminated incoherently (M470L3 Thorlabs) and projected behind a glass sphere (Borosilicate material, n = 1.48, 110μm diameter, Cospheric LLC, USA). The images were captured using an Andor Zyla 5.5 sCMOS camera, while focusing through the sphere (see S7 Fig). Depicted are images from the experiment (Real) and the simulation. The difference images are calculated wrt. to the undistorted test-chart image, showing that the real sphere-induced image distortions are qualitatively reproduced by the simulation. (PDF) Click here for additional data file.

Simulation of shift-shift memory effect.

a) Guide-star assisted diffraction-limited focusing in scattering tissue model. b) Lateral translation of this aberration compensated beam leads to gradual degradation. c) Quantification of focus degradation via correlation C(Δx) vs. distance (dotted line) at different penetration depth (in mean free path). Agreement with the correlation function of a scattered plane wave (solid line) is most pronounced at higher penetration depth. Deviation at low penetration depth (1 mfp) agree with experimental observations [10]. (PDF) Click here for additional data file.

Aberration correction.

a) A rectilinear ground truth stripe pattern occurs distorted when imaged through a sphere. b) Fitting a radial distortion map to simulation results of this scenario, allows one to fix the aberrations at the region of interest (green) as seen through the spherical cell phantom. (PDF) Click here for additional data file.

Runtimes of different methods for the simulation of a plane wave propagating through a given refractive index distribution.

(PDF) Click here for additional data file.
  30 in total

1.  Optical sectioning deep inside live embryos by selective plane illumination microscopy.

Authors:  Jan Huisken; Jim Swoger; Filippo Del Bene; Joachim Wittbrodt; Ernst H K Stelzer
Journal:  Science       Date:  2004-08-13       Impact factor: 47.728

2.  Quantitative high-speed imaging of entire developing embryos with simultaneous multiview light-sheet microscopy.

Authors:  Raju Tomer; Khaled Khairy; Fernando Amat; Philipp J Keller
Journal:  Nat Methods       Date:  2012-06-03       Impact factor: 28.547

3.  Physical insight into light scattering by photoreceptor cell nuclei.

Authors:  Moritz Kreysing; Lars Boyde; Jochen Guck; Kevin J Chalut
Journal:  Opt Lett       Date:  2010-08-01       Impact factor: 3.776

4.  Tomographic phase microscopy.

Authors:  Wonshik Choi; Christopher Fang-Yen; Kamran Badizadegan; Seungeun Oh; Niyom Lue; Ramachandra R Dasari; Michael S Feld
Journal:  Nat Methods       Date:  2007-08-12       Impact factor: 28.547

Review 5.  Selective plane illumination microscopy techniques in developmental biology.

Authors:  Jan Huisken; Didier Y R Stainier
Journal:  Development       Date:  2009-06       Impact factor: 6.868

6.  Nuclear architecture of rod photoreceptor cells adapts to vision in mammalian evolution.

Authors:  Irina Solovei; Moritz Kreysing; Christian Lanctôt; Süleyman Kösem; Leo Peichl; Thomas Cremer; Jochen Guck; Boris Joffe
Journal:  Cell       Date:  2009-04-17       Impact factor: 41.582

7.  Light-sheet fluorescence microscopy for quantitative biology.

Authors:  Ernst H K Stelzer
Journal:  Nat Methods       Date:  2015-01       Impact factor: 28.547

Review 8.  Optical properties of biological tissues: a review.

Authors:  Steven L Jacques
Journal:  Phys Med Biol       Date:  2013-05-10       Impact factor: 3.609

9.  Fractal propagation method enables realistic optical microscopy simulations in biological tissues.

Authors:  Adam K Glaser; Ye Chen; Jonathan T C Liu
Journal:  Optica       Date:  2016       Impact factor: 11.104

10.  Rapid adaptive optical recovery of optimal resolution over large volumes.

Authors:  Kai Wang; Daniel E Milkie; Ankur Saxena; Peter Engerer; Thomas Misgeld; Marianne E Bronner; Jeff Mumm; Eric Betzig
Journal:  Nat Methods       Date:  2014-04-13       Impact factor: 28.547

View more
  5 in total

1.  Development of a beam propagation method to simulate the point spread function degradation in scattering media.

Authors:  Xiaojun Cheng; Yunzhe Li; Jerome Mertz; Sava Sakadžić; Anna Devor; David A Boas; Lei Tian
Journal:  Opt Lett       Date:  2019-10-15       Impact factor: 3.776

2.  How to define and optimize axial resolution in light-sheet microscopy: a simulation-based approach.

Authors:  Elena Remacha; Lars Friedrich; Julien Vermot; Florian O Fahrbach
Journal:  Biomed Opt Express       Date:  2019-12-02       Impact factor: 3.732

3.  Rod nuclear architecture determines contrast transmission of the retina and behavioral sensitivity in mice.

Authors:  Kaushikaram Subramanian; Martin Weigert; Oliver Borsch; Heike Petzold; Alfonso Garcia-Ulloa; Eugene W Myers; Marius Ader; Irina Solovei; Moritz Kreysing
Journal:  Elife       Date:  2019-12-11       Impact factor: 8.140

4.  3D fluorescence microscopy data synthesis for segmentation and benchmarking.

Authors:  Dennis Eschweiler; Malte Rethwisch; Mareike Jarchow; Simon Koppers; Johannes Stegmaier
Journal:  PLoS One       Date:  2021-12-02       Impact factor: 3.240

5.  Challenges and advances in optical 3D mesoscale imaging.

Authors:  Sebastian Munck; Christopher Cawthorne; Abril Escamilla-Ayala; Axelle Kerstens; Sergio Gabarre; Katrina Wesencraft; Eliana Battistella; Rebecca Craig; Emmanuel G Reynaud; Jim Swoger; Gail McConnell
Journal:  J Microsc       Date:  2022-05-05       Impact factor: 1.952

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.