Literature DB >> 29619962

A Prospective 4π Radiation Therapy Clinical Study in Recurrent High-Grade Glioma Patients.

Victoria Y Yu1, Angelia Landers1, Kaley Woods1, Dan Nguyen1, Minsong Cao1, Dongsu Du1, Robert K Chin1, Ke Sheng1, Tania B Kaprealian2.   

Abstract

PURPOSE: To evaluate the feasibility, safety, dosimetric benefits, delivery efficiency, and patient comfort in the clinical implementation of 4π radiation therapy. METHODS AND MATERIALS: Eleven patients with recurrent high-grade glioma were recruited for the trial. 4π plans integrating beam orientation and fluence-map optimization were created using an in-house column-generation algorithm. The collision-free beam solution space throughout the 4π steradian was determined using a computer-aided-design model of the Varian TrueBeam system and a human subject. Twenty beams were optimized for each case and imported into Eclipse for intensity modulated radiation therapy planning. Beam orientations with neighboring couch kicks were merged for increased delivery efficiency, generating plans with an average of 16 beam orientations. Volumetric modulated arc therapy (VMAT) plans with 3-4 arcs were also generated for each case, and the plan achieving superior dosimetric quality was selected for treatment. Patient comfort was surveyed after every fraction. Multiple 2-dimensional X-ray images were obtained to measure intrafractional motion.
RESULTS: Of 11 patients, 9 were treated with 4π. Mean and maximum organ at risk doses were equal or significantly lower (P < .05) with 4π than with VMAT. Particularly substantial dose reduction of 2.92 Gy in the average accumulated brainstem maximum dose enabled treatments that would otherwise not satisfy safe dose constraints with VMAT. One patient was not treated because neither plan met the dosimetric criteria. The other was treated with VMAT owing to comparable dosimetry resulting from a planning target volume located in a separate co-plane superior to organs at risk. Treatments were well tolerated, with an average patient comfort score of 8.6/10. Intrafractional motion was <1.5 mm for all delivered fractions, and the average delivery time was 34.1 minutes.
CONCLUSIONS: The feasibility, safety, dosimetric benefits, delivery efficiency, and patient comfort of 4π radiation therapy have been clinically demonstrated with a prospective clinical trial. The results elucidate the potential and challenges of wider clinical implementations.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2018        PMID: 29619962     DOI: 10.1016/j.ijrobp.2018.01.048

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  15 in total

1.  Many-isocenter optimization for robotic radiotherapy.

Authors:  Qihui Lyu; Ryan Neph; Victoria Y Yu; Dan Ruan; Salime Boucher; Ke Sheng
Journal:  Phys Med Biol       Date:  2020-02-10       Impact factor: 3.609

2.  Comparison of conventional and advanced radiotherapy techniques for left-sided breast cancer after breast conserving surgery.

Authors:  Yibo Xie; Daniel Bourgeois; Beibei Guo; Rui Zhang
Journal:  Med Dosim       Date:  2020-07-07       Impact factor: 1.482

3.  Parallel beamlet dose calculation via beamlet contexts in a distributed multi-GPU framework.

Authors:  Ryan Neph; Cheng Ouyang; John Neylon; Youming Yang; Ke Sheng
Journal:  Med Phys       Date:  2019-06-30       Impact factor: 4.071

4.  A novel optimization framework for VMAT with dynamic gantry couch rotation.

Authors:  Qihui Lyu; Victoria Y Yu; Dan Ruan; Ryan Neph; Daniel O'Connor; Ke Sheng
Journal:  Phys Med Biol       Date:  2018-06-13       Impact factor: 3.609

5.  A reinforcement learning application of a guided Monte Carlo Tree Search algorithm for beam orientation selection in radiation therapy.

Authors:  Azar Sadeghnejad-Barkousaraie; Gyanendra Bohara; Steve Jiang; Dan Nguyen
Journal:  Mach Learn Sci Technol       Date:  2021-05-13

6.  A sparse orthogonal collimator for small animal intensity-modulated radiation therapy. Part II: hardware development and commissioning.

Authors:  Kaley Woods; Ryan Neph; Dan Nguyen; Ke Sheng
Journal:  Med Phys       Date:  2019-11-04       Impact factor: 4.071

7.  A sparse orthogonal collimator for small animal intensity-modulated radiation therapy part I: Planning system development and commissioning.

Authors:  Kaley Woods; Dan Nguyen; Ryan Neph; Dan Ruan; Daniel O'Connor; Ke Sheng
Journal:  Med Phys       Date:  2019-11-04       Impact factor: 4.071

Review 8.  Recent developments in non-coplanar radiotherapy.

Authors:  Gregory Smyth; Philip M Evans; Jeffrey C Bamber; James L Bedford
Journal:  Br J Radiol       Date:  2019-02-01       Impact factor: 3.039

9.  ROAD: ROtational direct Aperture optimization with a Decoupled ring-collimator for FLASH radiotherapy.

Authors:  Qihui Lyu; Ryan Neph; Daniel O'Connor; Dan Ruan; Salime Boucher; Ke Sheng
Journal:  Phys Med Biol       Date:  2021-01-29       Impact factor: 3.609

10.  Cochlea-sparing acoustic neuroma treatment with 4π radiation therapy.

Authors:  Kaley Woods; Percy Lee; Tania Kaprealian; Isaac Yang; Ke Sheng
Journal:  Adv Radiat Oncol       Date:  2018-02-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.