Difference in intraocular pressure measurements between non-contact tonometry and Goldmann applanation tonometry and the role of central corneal thickness in affecting glaucoma referrals.

Khalid Kamel1, Edward Dervan2, Kevin Falzon3, Colm O'Brien2.   

Abstract

BACKGROUND: Patients at glaucoma risk are commonly identified by optometrists and subsequently referred to glaucoma specialists. Optometrists mainly use non-contact tonometry (NCT) for intraocular pressure (IOP) measurement. AIMS: To investigate the role of differences in IOP measurement between NCT and Goldmann applanation tonometry (GAT) and the effect of central corneal thickness (CCT) on these differences in optometrist referrals
METHODS: Details of the initial clinical visit of patients referred with IOP > 21 mmHg in either eye as measured by NCT to a consultant glaucoma specialist were retrospectively reviewed. Demographic and referral data, IOP, CCT, and glaucoma diagnosis were obtained. The main outcome measure was the IOP measurement differences between NCT and GAT.
RESULTS: Of the 98 patients referred, only 23% had IOP > 21 mmHg when measured by GAT. NCT (Nidek NT400, Reichert Puff, Pulsair Easy Eye) measured the IOP greater than GAT by a mean of 5.8 mmHg (NCT 24.1 ± 3.5, GAT 18.3 ± 3.0). The effect of CCT on IOP measurement was less for GAT (R2 0.034, p = 0.067) than for NCT (R2 0.088, p = 0.003). The NCT/GAT IOP differences increased with increasing CCT (R2 0.166, p < 0.0001). The NCT/GAT differences decreased with patient age (R2 0.048, p = 0.03). Patients were classified as normal 67% (66/98), ocular hypertension 11% (11/98), glaucoma suspect 14% (14/98), and glaucoma 7% (7/98).
CONCLUSIONS: The difference in IOP measurement between NCT and GAT leads to a possible increase in glaucoma referrals, particularly in patients with thicker corneas. Repeat IOP using GAT and CCT measurement would help in triaging referrals.

Entities:  

Keywords:  Central corneal thickness; Goldmann tonometry; Intraocular pressure; Non-contact tonometry; Optometry

Mesh:

Year:  2018        PMID: 29616408     DOI: 10.1007/s11845-018-1795-0

Source DB:  PubMed          Journal:  Ir J Med Sci        ISSN: 0021-1265            Impact factor:   1.568


  21 in total

1.  The changing pattern of glaucoma referrals by optometrists.

Authors:  S A Vernon
Journal:  Eye (Lond)       Date:  1998       Impact factor: 3.775

2.  Effect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanation tonometer, and Tono-Pen.

Authors:  Archana Bhan; Andrew C Browning; Sunil Shah; Robin Hamilton; Dinesh Dave; Harminder S Dua
Journal:  Invest Ophthalmol Vis Sci       Date:  2002-05       Impact factor: 4.799

3.  Automated perimetry by optometrists in patients at low risk of glaucoma.

Authors:  R Stevenson
Journal:  Br J Ophthalmol       Date:  1999-06       Impact factor: 4.638

4.  Varying effects of corneal thickness on intraocular pressure measurements with different tonometers.

Authors:  Y-C Ko; C J-L Liu; W-M Hsu
Journal:  Eye (Lond)       Date:  2005-03       Impact factor: 3.775

5.  Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis.

Authors:  Jun Liu; Cynthia J Roberts
Journal:  J Cataract Refract Surg       Date:  2005-01       Impact factor: 3.351

6.  The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry.

Authors:  P-A Tonnu; T Ho; T Newson; A El Sheikh; K Sharma; E White; C Bunce; D Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-07       Impact factor: 4.638

7.  Outcomes of referrals by community optometrists to a hospital glaucoma service.

Authors:  B Bowling; S D M Chen; J F Salmon
Journal:  Br J Ophthalmol       Date:  2005-09       Impact factor: 4.638

8.  Positive predictive value of optometrist-initiated referrals for glaucoma.

Authors:  J Theodossiades; I Murdoch
Journal:  Ophthalmic Physiol Opt       Date:  1999-01       Impact factor: 3.117

9.  Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer.

Authors:  Aachal Kotecha; Ahmed Elsheikh; Cynthia R Roberts; Haogang Zhu; David F Garway-Heath
Journal:  Invest Ophthalmol Vis Sci       Date:  2006-12       Impact factor: 4.799

10.  Community refinement of glaucoma referrals.

Authors:  D B Henson; A F Spencer; R Harper; E J Cadman
Journal:  Eye (Lond)       Date:  2003-01       Impact factor: 3.775

View more
  2 in total

1.  Effects of baffle and intraocular pressure on aerosols generated in the noncontact tonometer measurement during COVID-19.

Authors:  Yuan Tang; Yan-Yan Chen; Chun-Chun Li; Zhang-Yan Chen; Chen Chen; Si-Qi Wen; Xiao-Qiong Huang; Jia Qu; Yan-Miao Chen; Ai-Ai Chen
Journal:  Int J Ophthalmol       Date:  2022-04-18       Impact factor: 1.779

2.  Corrected Intraocular Pressure Variability with Central Corneal Thickness Measurement.

Authors:  Mohamed Attia Ali Ahmed; Ahmed Shawkat Abdelhalim
Journal:  Clin Ophthalmol       Date:  2020-12-24
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.