| Literature DB >> 29611386 |
Joonhong Park1,2, Seungok Lee1,2, Yeongsic Kim1,2, Aeran Choi3, Hyeyoung Lee4, Jihyang Lim1,2, Yonggoo Kim1,2, Kyungja Han1,2, Eun Jee Oh1,5.
Abstract
BACKGROUND: Carcinoembryonic antigen (CEA) is one of the tumor markers available for evaluating disease progression status after initial therapy and monitoring subsequent treatment modalities in colorectal, gastrointestinal, lung, and breast carcinoma. We evaluated the correlations and differences between widely used, automated CEA immunoassays at four different medical laboratories.Entities:
Keywords: ADVIA Centaur XP; ARCHITECT i2000sr; Between-assay comparison; Carcinoembryonic antigen; Elecsys E170; Electrochemiluminescence; Unicel DxI800
Mesh:
Substances:
Year: 2018 PMID: 29611386 PMCID: PMC5895865 DOI: 10.3343/alm.2018.38.4.355
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
Summary of performance characteristics of four carcinoembryonic antigen (CEA) immunoassays according to information provided by the manufacturers
| Characteristics | ADVIA Centaur XP | ARCHITECT i2000sr | Elecsys E170 | UniCel DxI 800 |
|---|---|---|---|---|
| Method principle | Chemiluminescence immunoassay (CLIA) | Chemiluminescent Microparticle Immunoassay (CMIA) | Electrochemiluminescence immunoassay (ECLIA) | Chemiluminescence immunoassay (CLIA) |
| Assay principle | Two-site sandwich immunoassay | Two-step immunoassay | Sandwich immunoassay | Two-site immunoenzymatic “sandwich” assay |
| Sample volume (μL) | 50 | 60 | 10 | 35 |
| Sample stability | < 8 hours at 24℃ | < 24 hours at 24℃ | < 2 hours at 24℃ | < 8 hours at 24℃ |
| Sample storage | < 48 hours at 2–8℃ | < 7 days at 2–8℃ | < 7 days at 2–8℃ | < 48 hours at 2–8℃ |
| > 48 hours at –20℃ | > 7 days at –20℃ | < 6 months at –20℃ | > 48 hours at –20℃ | |
| Reagent | Purified polyclonal rabbit anti-CEA antibody labeled with acridinium ester, monoclonal mouse anti-CEA antibody covalently coupled to paramagnetic particles | Anti-CEA (mouse, monoclonal) coated microparticles, anti-CEA (mouse, monoclonal) acridinium-labeled conjugate | Streptavidin-coated microparticles, Biotinylated monoclonal anti-CEA antibody (mouse/human), monoclonal anti-CEA antibody (mouse) labeled with ruthenium complex | Anti-CEA MAb-alkaline phosphatase conjugate, anti-CEA MAb bound to paramagnetic particles |
| Calibration | Traceable to an internal standard manufactured using highly purified material | Traceable to an internal standard manufactured using highly purified material | Standardized against the 1st IRP WHO Reference Standard 73/601 | Traceable to an internal standard manufactured using highly purified material |
| Measuring range (ng/mL) | 0.5–100 | 0.5–500 (can be extended to 1,500) | 0.2–1,000 | 0.1–1,000 |
| Limit of detection (ng/mL) | 0.5 | 0.5 | 0.2 | 0.1 |
| Precision (total % CV) | 3.4–5.5 | 2.7–4.0 | 4.6–5.1 | 3.8–4.5 |
| Reference value (ng/mL) | 95th percentile; 5.0 | 93.5th percentile; < 5.0 | 95th percentile; 4.7, 5.2 | 95th percentile; 3.1–5.0 |
Comparison data of four carcinoembryonic antigen (CEA) immunoassays using 393 serum samples with CEA<1,000 ng/mL
| Statistics | Centaur XP (y) vs i2000sr (x) | Centaur XP (y) vs E170 (x) | Centaur XP (y) vs DxI800 (x) | i2000sr (y) vs E170 (x) | i2000sr (y) vs DxI800 (x) | E170 (y) vs DxI800 (x) |
|---|---|---|---|---|---|---|
| Deming regression | ||||||
| Regression equation | y=−1.1579+0.6335x | y=−3.8739+0.8197x | y=−1.1501+0.7326x | y=−4.0789+1.2895x | y=0.0232+1.1562 x | y=3.1615+0.8970x |
| Intercept, 95% CI | −3.1532 to 0.8373 | −7.7759 to 0.02813 | −3.2721 to 0.9719 | −8.1134 to −0.04427 | −2.9394 to 2.9858 | 0.9961 to 5.3269 |
| Slope B, 95% CI | 0.5679 to 0.6991 | 0.6744 to 0.9650 | 0.6474 to 0.8177 | 1.1479 to 1.4310 | 1.0485 to 1.2639 | 0.8245 to 0.9696 |
| Bland-Altman analysis | ||||||
| Mean difference (ng/mL) (95% CI) | −21.5 (−27.1 to −15.8) | −12.2 (−16 to −8.5) | −14.0 (−18 to −9.9) | 9.3 (5.3 to 13.3) | 7.5 (4.4 to 10.6) | −1.8 (−4.02 to 0.46) |
| | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | 0.1196 |
Abbreviations: y, y axis; x, x axis; 95% CI, 95% confidence interval.
Fig. 1Method comparison of four carcinoembryonic antigen (CEA) immunoassays by Deming regression analysis. Results are in ng/mL. Black solid lines are from Deming regression. Black dashed lines are x=y. Comparison of CEA concentrations between ARCHITECT i2000sr and ADVIA Centaur XP (A), Elecsys E170 and ADVIA Centaur XP (B), UniCel DxI800 and ADVIA Centaur XP (C), Elecsys E170 and ARCHITECT i2000sr (D), UniCel DxI800 and ARCHITECT i2000sr (E), and UniCel DxI800 and Elecsys E170 (F).
Fig. 2Method comparison of four carcinoembryonic antigen (CEA) immunoassays using Bland-Altman plots in 393 serum samples. The solid line indicates the % difference from the averages of two assays. Dashed lines indicate the upper and lower 95% confidence limits of % difference between two analyzers. The red lines are the acceptable bias limits of 30.0%. Comparison of CEA concentrations between ADVIA Centaur XP and ARCHITECT i2000sr (A), ADVIA Centaur XP and Elecsys E170 (B), ADVIA Centaur XP and UniCel DxI800 (C), ARCHITECT i2000sr and Elecsys E170 (D), ARCHITECT i2000sr and UniCel DxI800 (E), and Elecsys E170 and UniCel DxI800 (F).
Comparison data of four carcinoembryonic antigen (CEA) immunoassays using 355 serum samples with CEA<100 ng/mL
| Statistics | Centaur XP (y) vs i2000sr (x) | Centaur XP (y) vs E170 (x) | Centaur XP (y) vs DxI800 (x) | i2000sr (y) vs E170 (x) | i2000sr (y) vs DxI800 (x) | E170 (y) vs DxI800 |
|---|---|---|---|---|---|---|
| Passing-Bablok regression | ||||||
| Regression equation | y=−0.629+0.628x | y=−0.377+0.671x | y=−0.038+0.709x | y=0.665+1.038x | y=1.214+1.114x | y=0.557+1.045x |
| Intercept (95% CI) | −0.910 to −0.386 | −0.595 to −0.181 | −0.199 to 0.104 | 0.269 to 1.048 | 0.640 to 1.727 | 0.240 to 0.904 |
| Slope B (95% CI) | 0.597 to 0.657 | 0.648 to 0.693 | 0.691 to 0.729 | 0.996 to 1.074 | 1.049 to 1.175 | 1.002 to 1.085 |
| Cusum test for linearity | ns ( | ns ( | ns ( | ns ( | ns ( | ns ( |
| Bland-Altman analysis | ||||||
| Mean difference (ng/mL) (95% CI) | −8.1 (−9.1 to −7.2) | −6.5 (−7.2 to −5.7) | −5.1 (−6.0 to −4.3) | 1.6 (1.1 to 2.2) | 3.0 (2.2 to 3.8) | 1.3 (0.6 to 2.1) |
| | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | 0.0003 |
Abbreviations: y, y axis; x, x axis; 95% CI, 95% confidence interval.
Agreement of positivity between CEA assays using a cutoff value of 5.0 ng/mL
| ADVIA Centaur XP | ARCHITECT i2000sr | Elecsys E170 | |
|---|---|---|---|
| ARCHITECT i2000sr | 70.7% | ||
| Elecsys E170 | 74.6% | 95.2% | |
| UniCel DxI800 | 81.9% | 87.3% | 92.1% |