| Literature DB >> 35071206 |
Boo Young Hwang1,2, Eunsoo Kim1,2, Seung-Ha Kim1, Hyundoo Hwang3.
Abstract
Carcinoembryonic antigen (CEA) is a biomarker indicated in different cancers, targeted for quantitative analysis via immunoassay. Here we introduce a new technique called magnetic force-assisted electrochemical sandwich immunoassay (MESIA) for determination of CEA level in a drop of human serum using a fully automated point-of-care testing (POCT) device. The analytical performances of the assay are assessed based on precision, accuracy, limit of blank (LoB), limit of detection (LoD) and limit of quantitation (LoQ), linearity, Hook effect, interference, cross-reactivity, and method comparison following the guidelines of the Clinical Laboratory Standards Institute (CLSI). The LoD is 0.50 ng/ml. A linear relationship is shown in the range of 0.5-200 ng/ml. A high dose effect is not seen up to approximately 500,000 ng/ml. The recovery range is from 94.7 to 108.9%. The %CV of run-to-run and within-lab variations are less than 2.04 and 4.41% across the CEA concentrations, respectively, whereas reproducibility is 4.45-6.24%. Method comparison shows that the assay correlates well with the reference device (R 2 = 0.9884). The assay demonstrates acceptable precision, accuracy, LoB, LoD and LoQ, hook effect, linearity, interference, cross-reactivity, and high correlation with its reference device. Thus, the system is suitable for the quantification of CEA in clinical practices with a POCT manner.Entities:
Keywords: CEA; MESIA; electrochemistry; immunoassay; performance evaluation
Year: 2022 PMID: 35071206 PMCID: PMC8767762 DOI: 10.3389/fbioe.2021.798079
Source DB: PubMed Journal: Front Bioeng Biotechnol ISSN: 2296-4185
FIGURE 1Schematic diagram of magnetic force-assisted electrochemical sandwich immunoassays (MESIA).
FIGURE 2Comparison of measured CEA values against known concentrations of nine different CEA-spiked samples.
FIGURE 3Representation of peak current over a concentration range from 10 to 500,000 ng/ml CEA.
Mean % recoveries evaluated from four replicates for each level of concentration.
| Sample | Expected concentration (ng/ml) | Measured concentration (ng/ml) | %Recovery | Mean %Recovery |
|---|---|---|---|---|
| Level 1 | 3.42 | 3.30 | 96.5 | 99.9 |
| 3.61 | 105.6 | |||
| 3.24 | 94.7 | |||
| 3.51 | 102.6 | |||
| Level 2 | 15.50 | 14.90 | 96.1 | 99.7 |
| 16.20 | 104.5 | |||
| 15.60 | 100.6 | |||
| 15.10 | 97.4 | |||
| Level 3 | 75.20 | 72.30 | 96.1 | 102.0 |
| 78.50 | 104.4 | |||
| 81.90 | 108.9 | |||
| 74.20 | 98.7 | |||
| Level 4 | 151.40 | 151.20 | 99.9 | 99.6 |
| 146.20 | 96.6 | |||
| 149.90 | 99.0 | |||
| 155.70 | 102.8 |
Statistical analysis for single-site precision.
| Sample | Mean (ng/ml) | N | Repeatability | Between-run | Between-day | Within-lab | ||||
|---|---|---|---|---|---|---|---|---|---|---|
| SD | %CV | SD | %CV | SD | %CV | SD | %CV | |||
| 1 | 1.30 | 80 | 0.058 | 4.45 | 0.000 | 0.00 | 0.000 | 0.00 | 0.053 | 4.06 |
| 2 | 3.35 | 80 | 0.138 | 4.11 | 0.000 | 0.00 | 0.038 | 1.13 | 0.137 | 4.10 |
| 3 | 15.3 | 80 | 0.539 | 3.52 | 0.312 | 2.04 | 0.263 | 1.71 | 0.676 | 4.41 |
| 4 | 74.7 | 80 | 2.478 | 3.32 | 1.228 | 1.64 | 0.750 | 1.00 | 2.866 | 3.84 |
| 5 | 154.9 | 80 | 5.372 | 3.47 | 2.698 | 1.74 | 0.604 | 0.39 | 4.606 | 2.97 |
Statistical analysis for multi-site precision.
| Sample | Mean (ng/ml) | N | Repeatability | Between-day | Between-lot | Between-site | Reproducibility | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | |||
| 1 | 1.29 | 150 | 0.096 | 6.63 | 0.016 | 1.22 | 0.000 | 0.00 | 0.004 | 0.32 | 0.081 | 6.24 |
| 2 | 3.34 | 150 | 0.141 | 4.21 | 0.028 | 0.84 | 0.093 | 2.77 | 0.000 | 0.00 | 0.149 | 4.45 |
| 3 | 15.50 | 150 | 0.872 | 5.62 | 0.000 | 0.00 | 0.201 | 1.30 | 0.000 | 0.00 | 0.844 | 5.44 |
| 4 | 75.90 | 150 | 3.473 | 4.58 | 0.991 | 1.31 | 1.596 | 2.10 | 0.906 | 1.19 | 3.813 | 5.02 |
| 5 | 158.20 | 150 | 6.971 | 4.41 | 5.186 | 3.28 | 3.330 | 2.10 | 1.441 | 0.91 | 7.838 | 4.95 |
FIGURE 4Comparison of the CEA values of 131 samples derived from patients measured by the proposed method with those measured by a reference device, UniCel DxI 800 Access Immunoassay System (Beckman Coulter) (r = 0.9942, y = 0.9985x - 0.1924).