Younhee Park1, Yongjung Park, Jungyong Park, Hyon-Suk Kim.
Abstract
PURPOSE: Tumor marker concentrations in a given specimen measured by different analyzers vary according to assay methods, epitopes for antibodies used, and reagent specificities. Although great effort in quality assessment has been instituted, discrepancies among results from different analyzers are still present. We evaluated the assay performance of the UniCel™ DxI 800 automated analyzer in measuring the alpha-fetoprotein (AFP), carcinoembryonic antigen (CEA), carbohydrate antigen (CA) 125, CA 15-3 and CA 19-9 tumor markers.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22477000 PMCID: PMC3343438 DOI: 10.3349/ymj.2012.53.3.557
Source DB: PubMed Journal: Yonsei Med J ISSN: 0513-5796 Impact factor: 2.759
Precision Performances of UniCel™ DxI 800 in Measuring Five Tumor Markers
CEA, carcinoembryonic antigen; AFP, alpha-fetoprotein; CA, carbohydrate antigen.
*Units are µg/L for CEA and AFP, and kU/L for CA 15-3, CA 125, and CA 19-9.
Fig. 1Linearity of the five tumor marker assays performed by UniCel™ DxI 800. The observed mean tumor marker levels showed excellent linear responses to the expected values (p<0.001). (A) The linearity range of the CEA assay was from 0.80 to 951.41 µg/L (R2=0.9997). (B) The AFP assay was linear in the range between 1.86 and 2092.91 µg/L (R2=0.9991). (C) The CA 15-3 assay showed linear responses to the expected concentrations in the range of 5.55 to 633.45 kU/L (R2=0.9983). (D) The CA 125 assay was verified to be linear at the concentrations between 7.68 and 1752.53 kU/L (R2=0.9994). (E) Linearity of the CA 19-9 assay was validated in the range of 0.80 to 1389.73 kU/L (R2=0.9995). CEA, carcinoembryonic antigen; AFP, alpha-fetoprotein; CA, carbohydrate antigen.
Parameters of Passing and Bablok Regression between UniCel™ DxI 800 and the Other Systems
CI, confidence interval; CEA, carcinoembryonic antigen; AFP, alpha-fetoprotein; CA, carbohydrate antigen.
*Units are µg/L for CEA and AFP, and kU/L for CA 15-3, CA 125, and CA 19-9.
Analytic Concordance between UniCel™ DxI 800 and the Other Analyzers
CEA, carcinoembryonic antigen; AFP, alpha-fetoprotein; CA, carbohydrate antigen.
*CEA, 5.0 µg/L; AFP, 7.4 µg/L; CA 15-3, 31.3 kU/L; CA 125, 35.0 kU/L; CA 19-9, 35.0 kU/L.
Fig. 2Comparison of the results of five tumor markers measured by the UniCel™ DxI 800 with those measured by other instruments using Bland and Altman difference plots. The solid line indicates the mean difference between the methods, and the dashed lines indicate the upper and lower 95% confidence limits of the difference between the methods. (A) The mean difference of CEA was 5.62 µg/L [95% confidence interval (CI), -6.53 to 17.77 µg/L]. (B) The mean difference of AFP was -145.39 µg/L (95% CI, -220.47 to -70.32 µg/L). (C) The mean difference of CA 15-3 was -27.59 kU/L (95% CI, -36.24 to -18.94 kU/L). (D) The mean difference of CA 125 was 26.73 kU/L (95% CI, 3.72 to 49.74 kU/L). (E) The mean difference of CA 19-9 was -209.59 kU/L (95% CI, -312.58 to -106.60 kU/L). CEA, carcinoembryonic antigen; AFP, alpha-fetoprotein; CA, carbohydrate antigen.