| Literature DB >> 29542269 |
Luc Beaulieu1,2, Dee-Ann Radford3,4, J Eduardo Villarreal-Barajas3,4.
Abstract
The Canadian Organization of Medical Physicists (COMP), in close partnership with the Canadian Partnership for Quality Radiotherapy (CPQR) has developed a series of Technical Quality Control (TQC) guidelines for radiation treatment equipment. These guidelines outline the performance objectives that equipment should meet in order to ensure an acceptable level of radiation treatment quality. The TQC guidelines have been rigorously reviewed and field tested in a variety of Canadian radiation treatment facilities. The development process enables rapid review and update to keep the guidelines current with changes in technology. This article contains detailed performance objectives and safety criteria for low-dose-rate (LDR) permanent seed brachytherapy.Entities:
Keywords: brachytherapy; low-dose-rate; quality control; seed implants
Mesh:
Year: 2018 PMID: 29542269 PMCID: PMC5978974 DOI: 10.1002/acm2.12307
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Daily quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Daily | |||
| DPB1 | Radiation survey meter | Functional | |
| DPB2 | Source strength verification (well chamber) | 3% | 5% |
| DPB3 | Ultrasound system/probe | Functional | |
| DPB4 | Source inventory | Complete | |
| DPB5 | Records | Complete | |
| DPB6 | Room survey (drape, needle, template, etc.) | Complete | |
| DPB7 | Console displays (treatment status indicator, date, time) | Functional | |
| DPB8 | Printer operation, paper supply | Functional | |
| DPB9 | System self‐test | Functional | |
| DPB10 | Delivery interrupt | Functional | |
| DPB11 | Power failure recovery | Functional | |
| DPB12 | Data transfer from planning computer | Functional | |
| DPB13 | Seed loading devices and disposable elements | Functional | |
| DPB14 | Communication between all systems | Functional | |
| DPB15 | Emergency seed loading kit | Functional/Sterilized | |
| DPB16 | Online source strength verification | 8% | 15% |
| DPB17 | Needle loading sequence as per treatment plan | Complete | |
Annual and bi‐annual quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Annually | |||
| APB1 | Ultrasound positional accuracy | 1 mm | 2 mm |
| APB2 | Ultrasound volumetric accuracy | 3% | 5% |
| APB3 | Stepper positional accuracy | 1 mm | 2 mm |
| APB4 | Template positional accuracy | 1 mm | 3 mm |
| APB5 | Source parameters and TPS dose calculation verification | 2% | 3% |
| APB6 | Emergency seed handling procedures review | Complete | |
| APB7 | Independent quality control review | Complete | |
| APB8 |
| Functional | |
| APB9 | Online source strength measurements device calibration/verification | 3% | 5% |
| APB10 | Source positional accuracy (loading devices) | 2 mm | 3 mm |
| APB11 | Survey meter calibration | Complete | |
| Bi‐annually | |||
| BPB1 | Well‐chamber calibration | 1% | 2% |
| DPB1 | Verify that the handheld radiation survey meter (e.g., |
| DPB2 |
The AAPM Low Energy Brachytherapy Source Calibration Working Group has outlined specific criteria.39 In general 10% of the seeds or 10 seeds, whichever number is larger, should be tested. For a sterile assembly, such as a sterile seed cartridge or pre‐loaded needle, the recommendation is the lowest of 5% of the seeds or 5 seeds. Complete descriptions of the scenarios between these two extremes are given in “table |
| DPB3 | In addition, visually inspect images for any artifacts, such as black lines or bands. Ensure they are not due to poor contact between the probe and tissue. If present, such bands may indicate non‐functioning ultrasound detector elements within the probe. Persistence of these artifacts may warrant image quality tests using a dedicated ultrasound phantom to characterize the location of the signal dropout and identify non‐functioning elements within the probe, which may have to be sent for repairs. |
| DPB4 | Could be performed in conjunction with DPB2 above if done on the same day as the procedure. Otherwise, inventory should be validated before moving the sources to the procedure room. |
| DPB5 | Documentation relating to the daily quality control checks, preventive maintenance, service calls, and subsequent checks must be complete and legible. The operator(s) must be identified. |
| DPB6 | The workspace (including the floor), needles, template, probes, etc., must be surveyed using a calibrated survey meter (see DPB1). Reading should be consistent with no radioactive materials outside the seeds implanted in the patient. This task must be performed after each implant. |
| DPB7–15 | The configuration of these tests will depend on the equipment selected and the clinical workflow (pre‐planning/live planning with or without a seed loading device). Safety is the concern and tests should be designed accordingly. As a minimum, manufacturer's recommendations and applicable regulations must be followed. |
| DPB16 | See DPB2 above regarding detector such as the SeedSelectron. |
| DPB17 | It is crucial that the needle loading sequence of each needle composing a given plan be validated and correspond to the treatment plan. For pre‐loaded needles, auto‐radiograph or x‐ray imaging will confirm the seed‐spacer sequence (or seed sequence for stranded seeds). For intra‐operative loading, a second person could visually confirm the loading as it is being done and most brachytherapy needle have graduation that can confirm the overall sequence length. In any case, this length should be confirmed (pre‐loaded or intra‐operative loading) before every needle insertion. Some devices, such as the SeedSelectron, have an array of radiation detectors that is used to confirm the seed (radiation present) and spacer (no radiation) sequence before loading. In such a case, the device itself should be regularly tested (see APB9 below). |
| APB1–4 |
AAPM Task Group 128 constitutes the reference document with regard to ultrasound system performance and related quality assurance tasks; a detail description of each test is given.9
|
| APB5 |
Peer reviewed articles giving dosimetric parameters of each approved seed model can be found in the registry ( |
| APB6 | The configuration of these tests will depend on the design of the facility and equipment used. Review the emergency procedures for seed/needle loading if a seed loading device is normally used and fails. Emergency procedures (e.g., if a seed should drop on the floor, is stuck in a needle, or is found in the urine bag) should be reviewed. |
| APB7 | To ensure redundancy and adequate monitoring, a second qualified medical physicist must independently verify the implementation, analysis, and interpretation of the quality control tests at least annually. |
| APB8 | It is recommended that a complete system validation be conducted once a year. In the present document this would include all the necessary validation for full system recovery from power outage (planning system recovery, seed delivery system, etc.,), delivery interruption, and other potentially deleterious events, as indicated in DBP6 to DPB13. These tests should be performed away from the daily clinical pressure and busy operating room environment. |
| APB9, 10 | These measurements have been discussed in various publications.40,41 |
| APB11 | Survey meter should be calibrated once every 12 months as per CNSC requirements (Nuclear Substances and Radiation Devices Regulations [SOR/2000‐207]44). |
| BPB1 | The well chamber should be sent to an accredited dosimetry calibration laboratory once every 2 years. A calibrated source, of each seed model used, could also be acquired from the manufacturer each year for verification purposes. |